An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. DEAD WEIGHT TESTER A dead weight tester apparatus uses known traceable weights to apply pressure to a fluid for checking the accuracy of readings from a pressure gauge. A dead weight tester (DWT) is a calibration standard method that uses a piston cylinder on which a load is placed to make an equilibrium with an applied pressure underneath the piston. Deadweight testers are so called primary standards which means that the pressure measured by a deadweight tester is defined through other quantities: length, mass and time. Typically deadweight testers are used in calibration laboratories to calibrate pressure transfer standards like electronic pressure measuring devices. In general there are three different kind of DWT's divided by the medium which is measured and the lubricant which is used for its measuring element: gas operated gas lubricated PCU's gas operated oil lubricated PCU's oil operated oil lubricated PCU's All three systems have their own specific operational demands. Some points of attention: Gas-gas Make sure that the PCU is clean. This is a very important issue as the PCU's operation is sensitive to contamination. Also when connecting a DUT, make sure that the DUT does not introduce contamination in the measuring system. Gas-oil Lubricant of the PCU 'leaks' in the gas-circuit of the DWT. For this reason there is a small reservoir incorporated in the system. Before commencing a calibration it is a good practice to purge this reservoir. If the reservoir is full, oil will be introduced in critical tubing and will cause an uncontrollable oil-head. Address: 27th KM Stone, Bangalore National Highway NH4, Chembarambakkam, Chennai, Tamil Nadu 600123.

  2. DEAD WEIGHT TESTER Oil-oil When connecting an oil filled DUT on an oil DWT make sure that the DUT oil will not contaminate the DWT oil. If in doubt. Incorporate a small volume between DUT and DWT and manipulate pressure in such a matter that the oil flow is directed to the DUT. For high accuracy measurement, friction can be lowered by rotation of the piston. Address: 27th KM Stone, Bangalore National Highway NH4, Chembarambakkam, Chennai, Tamil Nadu 600123.

More Related