An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. PCR Team Thermocycler and Fluid Delivery System

  2. Outline • Background • Budget • Schedule • Thermocycler CFP • Fluid Delivery CFP • Conclusion • Contact Information

  3. Background • Purpose • Develop a marketable device to amplify DNA through a process called PCR • Benefits • Improving Personalized Medicine • Analyzing Potential Drug Interactions • Analyzing Genetic Characteristics • Requirements • Fast • Portable • Inexpensive

  4. Background: Current Problems • Fluid Delivery: • Number of Samples • Changing Samples • Thermocycler: • Time to Steady State • Uniform Heat Distribution

  5. Budget

  6. 9/01 10/01 11/15 12/01 1/15 3/01 3/15 4/15 Begin Design Prototypes Build Thermocycler Control System for Thermocycler Build Fluid Delivery System Combine Thermocycler and Fluid Delivery Customer Testing Finished Product Schedule

  7. Thermal CFP: Motivation • Slide needs to be secured against the heater • Slide must not move during testing • Slide needs to have a quick removal and installation mechanism

  8. Thermal CFP: Prototypes • Cam Clamping • Spring Loaded Clamping • Foam Clamping • Silicon Clamping

  9. Thermal CFP: Prototypes Cam Spring Aluminum Plate Adjustment nut Cam CFP

  10. Thermal CFP: Prototypes Aluminumplate Spring Glass Slide Spacer Chamfers Spring CFP

  11. Thermal CFP: Prototypes Silicon Foam Aluminum plates Glass Slide Silicon CFP

  12. Thermal CFP: Prototypes Foam Foam CFP

  13. Thermal CFP: Testing Three testing criteria: • Vibration test • Drop test • Force test

  14. Thermal CFP: Testing of Clamps • VibrationTest Foam CFP Vibration machine

  15. Thermal CFP: Testing of Clamps • Drop Test Spring CFP 6 inches

  16. Thermal CFP: Testing of Clamps • Force Test Custom slide Force gage

  17. Thermal CFP: Testing of Clamps • Based on testing criteria: 1st – Cam 2nd – Silicon 3rd – Foam 4th – Spring

  18. Thermal CFP: Results Chips from corners • Cam • Cracked Slides • Non-uniform clamping • Silicon and foam • Difficult to install slides • Spring loaded • Difficult to install slides • Did not hold slides well Crack Cracked slide from Cam CFP

  19. Thermal CFP: New Clamp Design convection shield • Incorporates both the silicon and the cam. • Cam clamps on base instead of slide • Only silicon is in contact with slide Silicon Modeled by Chad Meeks and Steve Rabe

  20. Thermal CFP: What’s next? • Finalize the solid model and drawings. • Build the device. • Begin controls portion of project.

  21. Fluid CFP: Motivation • Verify pressure drop in microfluidic channels where PCR is performed. • Determine the amount of tubing that will decrease the flow rate to 0.5 and 5 μl/min.

  22. Fluids CFP: Pressure Test What gage pressure is required to push the fluid through the microchannel? 1-1.5 bars Syringe Pressure Gage

  23. Fluid CFP: Decreasing Flow Rate Syringe Pump Coil of Tubing Short Tubing

  24. Fluid CFP: Test Results Third Order Poly-fit: dV/dt = -4.27e-5L3 + 0.97e-3L² - 5.53e-3L + 7.11

  25. Flow Rate (μl/min) 5 0.5 Tubing (in) 60 62.5 Fluid CFP: Test Results

  26. Fluid CFP: What’s Next? • Improve equation relating fluid flow rate to coiled tube length with further testing. • Analyze available pumps for application to the PCR device.

  27. Conclusion • Thermocycler clamping mechanism will combine the cam and silicon prototypes • Analyzing different pumps and the required tubing for flow reduction.

  28. Jenny Greer jennygreer.eng@gmail.com Seth Plaizier splaizier@gmail.com Chad Meeks chky071@yahoo.com Steven Rabe srabe@ybfa.com Dan Torgerson dantorgerson@hotmail.com Christian Sellers csellers82@hotmail.com Contact Information References: Thermal CFP pictures taken by Seth Plaizier Fluid CFP pictures taken by Dan Torgerson