1 / 14

Fountain Spectrometer

Fountain Spectrometer. Alexander Koukras Gyula Nagy Miguel Fernández. Topics. Geometry Calibration T ransmission Focusing properties Resolution Conclusions. Geometry. Just PPA with 2 π cylindrical symmetry ( rotating around x axis) Advantage compared to PPA :

brooke-levy
Télécharger la présentation

Fountain Spectrometer

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. FountainSpectrometer Alexander Koukras Gyula Nagy Miguel Fernández

  2. Topics • Geometry • Calibration • Transmission • Focusing properties • Resolution • Conclusions

  3. Geometry • JustPPAwith 2πcylindricalsymmetry (rotatingaround x axis) • AdvantagecomparedtoPPA: • Greateracceptance: more solidangle (2πsymmetry)

  4. Calibration • Linear relationbetweenE0 and Vd • Here, q=-1 (e) so • For 45º RelationVd, E0: UsingLUA

  5. Calibration (45º & 30º) Slope Formula  Slope

  6. Transmission • T(%) = • For a fixedKE of particles and itscorrespondingVd (calibration) • Flyparticlesslightlydifferentangles measure #particlesenter, #particlespassthrough • 45º  T= 98.5% • 30º  T= 20% !!! • Widerexitslit + 3rdplate (onlyfor 30º)

  7. Transmission (3 plates) • Now: T = 92% • f doesnotchange • 3rdplatenotto lose resolution

  8. Focusing properties • For 45º  1storderfocusing, focusinside • For 30º  2ndorderfocusing, focusoutside Trace width:

  9. Resolution • Luaprogramtoscanenergy,measuretheEmin and Emaxthatleavetheexitslit, and calculate R • Thelower R  thebetter

  10. Resolution • Compare 45º withtheory:

  11. Resolution • Independent of energy • 30º has betterresolutionforsamegeometry • For 30º, R doesnotchangewith3rdplate + widerexitslit

  12. Conclusions • Betterresolutionfor 30º • Betterfocusingfor 30º • Lowertransmissionfor 30º • Solution: 3rdelectrode + widerexitslit

  13. Allinformation fountainspectrometer.blogspot.com QRCode

  14. Thanksforyourattention Thankstoouradvisors: • ManolisBenis • Pilar Garcés And allotherCPOTSteachers

More Related