1 / 22

E.Gaubas, T. Č eponis, J.Kusakovskij, A.Uleckas, S.Sakalauskas, and J.Vaitkus

Evaluation of fluence dependent variations of capacitance and generation current parameters by transient technique. E.Gaubas, T. Č eponis, J.Kusakovskij, A.Uleckas, S.Sakalauskas, and J.Vaitkus. Vilnius U niversity, Institute of Applied Research, Vilnius, Lithuania (VU). Outline

chin
Télécharger la présentation

E.Gaubas, T. Č eponis, J.Kusakovskij, A.Uleckas, S.Sakalauskas, and J.Vaitkus

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Evaluation of fluence dependent variations of capacitance and generation current parameters by transient technique E.Gaubas, T.Čeponis, J.Kusakovskij, A.Uleckas, S.Sakalauskas, and J.Vaitkus Vilnius University, Institute of Applied Research, Vilnius, Lithuania (VU) • Outline • Motivation for alternative techniques vs. impedance based frequency domain one  Principles of Barrier Evaluation by Linearily Increasing Voltage (BELIV) technique • Fluence dependent BELIV characteristics • Temperature dependent BELIV characteristics for reverse-biased diodes • Detector- barrier evaluation summary • Photo-conductivity gain in heavily irradiated full-depleted detectors  Summary

  2. Motivation for alternative techniques vs. impedance based frequency domain one Limitations for impedance, frequency domain based C-V (LRC) techniques  Principle (LCR meter) “works” only if complexity (jXC) appears due to conductance/impedance Applicable only when diode can be emulated by the linear elements Uac<<kT/e CP CS (UR) Cb0 (1+U/Ubi)-1/2 Cb0=(eND0S2/2eUbi)1/2slope C-2 vs U ND; intersect. Ubi-UF=0 Ubi Ubi=(kT/e)ln(NAp+ND/ni2) at least Uac/URdc<<1` Whether standard paradigm of C-V technique is valid when diode current iCigen, icapt for UR -? LRC phasor

  3. Motivation for alternative techniques vs. impedance based frequency domain one Limitations for impedance, frequency domain based C-V (LRC) techniques  Principle (LCR meter) “works” only if complexity (jXC) appears due to conductance/impedance Applicable only when diode can be emulated by the linear elements Uac<<kT/e CP CS (UR) Cb0 (1+U/Ubi)-1/2 Cb0=(eND0S2/2eUbi)1/2 Ubi=(kT/e)ln(NAp+ND/ni2) at least Uac/URdc<<1` Whether standard paradigm of C-V technique is valid when diode current iCigen, icapt & how to correlate with I-V-? Why huge difference in CP and Cs appears for heavily irradiated diode -? Why C-V measurements are improved at high frequencies-? Why extracted depletion voltage depends on frequency -? LRC phasor 100 kHz

  4. 30 Hz Motivation for alternative techniques vs. impedance based frequency domain one Limitations for impedance, frequency domain based C-V (LRC) techniques  Principle (LCR meter) “works” only if complexity (jXC) appears due to conductance/impedance Applicable only when diode can be emulated by the linear elements Uac<<kT/e CP CS (UR) Cb0 (1+U/Ubi)-1/2 Cb0=(eND0S2/2eUbi)1/2 Ubi=(kT/e)ln(NAp+ND/ni2) Whether standard paradigm of C-V technique is valid when diode current iCigen, icapt for UR -? Why huge difference in CP and Cs appears for heavily irradiated diode -? Why C-V measurements are improved at high frequencies-? Why extracted depletion voltage depends on frequency -? LRC phasor

  5. i*gen Motivation for alternative techniques vs. impedance based frequency domain one Limitations for impedance, frequency domain based C-V (LRC) techniques  Principle (LRC meter) “works” only if complexity (jXC) appears due to impedance How are the generation and carrier capture currents included into standard C-V for (heavily) irradiated pin diodes-? with large igen=eniw(U)S/gen gen=2cosh(EL-Ei)/vTNL- short i=iC+(igen+idiff - complex quatn.) Igen,capt=eniw(U)S/gen,captbecomes a complex quantity for ac U~ Is it possible to separate an impact of igen, capt from C-V to correlate with I-V -? LRC phasor becomes multi-dimensional • Indications that LRC impedance principle and a paradigm of parameter extraction does not work anymore: • - appears huge difference between CP and CS - artefact of principle due to i*gen, • - impossible to separate iCand i*gen, • appears crucial reverse voltage drop: •  e.g. Ugen=igenRdepl=w2(U)/genand indefinite voltage drop (on junction and within bulk) Utechn =Ugen • UFDshifts crucially to the high voltage/high frequency rangedue to increase of Ugen with Uext • -depletion width becomes indefinite(un-known gen) • - estimation by iteration procedure-- simulations by TCAD/SPICE or approach of positive root • w(Uex)=[20 (Ubi+Uext )/{eND(1+20 /eNDgen)}]1/2 – valid for Uac/Udc<<1; necessary additional parameters) • intricate (w(Uext)) dependence on frequency, temperature, voltage - due togen • appears an artificial effective dopingNefart= ND(120 /eNDgen) – seeming variations • Technical limitations for LRC-meters based measurements - necessary to make measurements in the range of low UR voltages to avoid the impact of igen,~ U1/2 - LRC meters with small ac voltage U~<<kT/e - dc and ac voltage sources connected in series (dependence on Rindc), - noise (Uns) suppressed dc voltage sources due to small Uns <<U~, - no additionalloops (capacitors, resistors), limitations to ground for LRC-meters with wide range of external dc voltages

  6.  To include dominant physical processes, analysis of Cbj(U)=0S/w(U) is an alternative way with estimation of w(U) by iteration procedure- approach of positive w(U) root using Ugen=w2/gen,capt valid for Uac/Udc<<1 and leads to different Neff=ND (120 /eND+gen,-capt) for 1/gen (+) and for 1/capt (-)  Evaluation of w(Ujunct) by simulations by TCAD/SPICE etc. Motivation for alternative techniques vs. impedance based frequency domain one  Principles to include igen, capt • Alternative measurement techniques capable to separate components

  7. Cryo-chamber AT-DSO-6102A iC BELIV technique U(t)=UP/PLt =At LIV ramp A=UP/PL = U/t LIV always starts from U(t)=0 PL= 10 ns  500 s UP= 0.01  5 V Cb= 2 pF  40 µF with resolution  0.2 pF | (2-20 pF) UC=iC*50  =10 mV  4 pF|A=5*E8 V/s

  8. Charge extraction current Displacement current te Charge extraction Charge generation BELIV technique, model and simulations Reverse bias Short transient processes acting in series due to tD, DR, capt, gen etc (to complete a circuit) determine a delay,- reduction of the initial displacement current step. Similar effect perturbs the C-V characteristic at UR0 measured by impedance technique (LRC-meters).

  9. BELIV technique UF model U(t)=UP/PLt =At Forward bias Transcendental, iterative simulations

  10. Barrier (charge) capacitance Diffusion (storage charge) capacitance BELIV transients – measurement technique Reverse biasing Forward biasing a b a- Barrier evaluation by linearily increasing voltage (BELIV) technique based on charge extraction current transients measured in the non-irradiated and irradiated with small fluence pin diode at reverse (UR) biasing by LIV pulses. b- Charge injection BELIV transients for forward (UF) biased pin diode irradiated with small fluence varying ramp A of LIV pulses.

  11. BELIV transients – measurement technique UR pulse duration dependent BELIV transients at a constant LIV ramp is equivalent to Cb -V~t

  12. Cb dominates for 1E12 n/cm2 Igen & irec dominate for 1E16 n/cm2 Barrier capacitance prevails for short LIV pulses BELIV transients on WODEAN pad-detectors neutron-irradiated with fluences 1012 -1016 n/cm2 Reverse bias Reverse & Forward bias a b a- LIV pulse duration (LIV ramp) dependent BELIV transients at reverse (UR) bias. b- Comparison of charge extraction (UR) and injection (UF)BELIV transients measured on diodes irradiated by neutrons of different fluence. To separate the displacement, generation/capture and diffusion currents, a wide range of duration/voltage and perfect LIV pulses are necessary. For diodes irradiated with rather small fluence charge extraction current prevails for Rev biasing, while charge storage (diffusion) capacitance dominates for Frw biased diodes. In heavily irradiated material generation and recombination currents dominate. Voltage on junction is governed by Ujnc=At-iRL

  13. iC(t) te BELIV results –dependence on fluence a b a- Variations of BELIV transients with irradiation fluence for reverse biased Si pin pad-detector at the same LIV parameters (A= 3 MV/s, PL =1.5 s). b- Injection current BELIV transients for different fluence neutron irradiated Si pin diodes. The extreme points (te, tFk) are denoted on transients.

  14. BELIV results –dependence on temperature T=294 K Temperature dependent variations of BELIV transients in heavily (with fluence of 1016 n/cm2) irradiated Si pin diode. Ubi still exists (for 1E16 n/cm2) but diode at RT resembles a photo-resistor at large Ubias

  15. for tdr,e<tdr,h to keep el. neutrality additional carrier(s) is injected from electrode or thermally generated in vicinity of electrode - p+ n+ + tdr,e=le2/eU tdr,h=lh2/hU Rec gT tr iat FD n, H+ Photo-conductivity gain (PCG) • PCG=tdr,h/tdr,e  CCE>1 if tdr,e<tdr,h<Rec • CCE0 for Rec<tdr • radiation inducedgeneration (trapping) centers (increased density with fluence) • enhance probability for PCG processes; • reduction of tdrby enhancement of U>UFD and by reduction of d~lh; • resolvable PCG pulses if time gap tarbetween arrival of HE particles tar>tr

  16. Summary: detector- barrier evaluation by LIV For barrier evaluation better to control Ubi=(kT/e)ln(NAND/ni2) BELIV shows ND is invariable, barrier exists, detector functional, but necessary to shift the operational range towards high () frequency/short time domain range, to decrease bias voltage (to suppress igen) etc. • Optimization of the detector’ functionality is possible only by a trade-off among desirable parameters: • necessity to shift the time domain TD towards short shaping time TD1/ is compatible with LHC • operation regime, but, to reduce an impact of igen,rec , it is desirable to keep rec,gen >TD>DR • (reduction of DR is possible by enhancement of doping); • to suppress igen, rec - reduction of the operational voltages, temperature and base thickness d; • to reach full-depletion – enhancement of operational voltage, but increases a problem with igen~U1/2 ; • to reduce impact of g-r noises – enhancement of detector volume V=Sd • through base thickness or area S ( 3-D detectors); • - to approach a photo-conductivity gain regime (to increase CCE) enhancement of operational voltage • and proper reduction of base thickness d;

  17. Summary •  Validity of a simple paradigm (standard C-V method) of the parameter extraction from C-V characteristics in irradiated diodes should be verified for every experimental regime and adjusted by selecting , UR, Uac, T parameters • Developed BELIV transient technique enables one to control variations of Ubi and generation current dependent on fluence and temperature. Increase of generation current for reverse biased diode and reduction of diffusion capacitance/current due to recombination of injected carriers in forward biased diodes leads to the symmetric (Rev/Frw) I-V/C-V transient characteristics. Symmetry of these characteristics indicate the dominant mid-gap centers/clusters with pinned EF. • BELIV shows Ubiand ND is invariable and detector is functional, while in heavily irradiated detectors for 20 Hz< <10 MHz deterioration of characteristics is determined by capt,gen. • Optimization of the detector’ functionality is possible only by a trade-off among desirable parameters. CCE can be increased through proper design of detector (d) and applied detection regimes (U, response) to reach PCG.

  18. Thank You for attention!

  19. Trivial remarks concerning SRH relations between recombination – generation lifetime rec, gen rec, gen Generation lifetime Generation lifetime Without applied E field Recombination lifetime Recombination lifetime With applied dc E field ni/ND ni/ND EDL-Ei (EDL-Ei)/kT R>0 Eqv. G<0

  20. Calibration of LRC-meters

  21. BELIV results –dependence on fluence a b a- Variations of BELIV transients with irradiation fluence for reverse biased Si pin pad-detector at the same LIV parameters (A= 3 MV/s, PL =1.5 s). b- Injection current BELIV transients for different fluence neutron irradiated Si pin diodes. The extreme points (te, tFk) are denoted on transients.

  22. Relation among Cb,, Rdepl and LRC: Y,Rp ,Cp