1 / 8

Deux types de circuit à basse pression!!!

Deux types de circuit à basse pression!!!. Valve de surpression présente uniquement sur respirateur Ohmeda. Détails de la valve de surpression. Fermeture si hyperpression dans le circuit inspiratoire pour éviter effet de pompage dans les vaporisateurs.

cole-rose
Télécharger la présentation

Deux types de circuit à basse pression!!!

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Deux types de circuit à basse pression!!! Valve de surpression présente uniquement sur respirateur Ohmeda

  2. Détails de la valve de surpression Fermeture si hyperpression dans le circuit inspiratoire pour éviter effet de pompage dans les vaporisateurs

  3. Test de fuite sur appareil type Drager Si fuite chute de pression: maintenir 30 cm H2O pendant 30 sec

  4. Test fuite sur Ohmeda Excel Test pression négative Test pression positive: Inutile pour fuite en amont de la valve

  5. Rotamètres !! Sur Ohmeda debit minimal fixé à 200 ml/min O2 ! Risque de dysfonctionnement (humidité,…)

  6. Link 25: ohmeda Assure 25% de FiO2

  7. ORC: Oxygen ratio controller Système pneumatique assurant FiO2: 25%

  8. Support des vaporisateurs Risque de mauvais positionnement ou dysfonction du vapo