Summary of RMD 8x8 mm 2 APDs

Summary of RMD 8x8 mm 2 APDs. T. Tsang, BNL , Dec. 13, 2013. spatial uniformity. time walk. time jitter. Conclusion: metallic mesh with gold sintered device is the best of all.

daisey
Télécharger la présentation

Summary of RMD 8x8 mm 2 APDs

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Summary of RMD 8x8 mm2 APDs T. Tsang, BNL, Dec. 13, 2013 spatial uniformity time walk time jitter Conclusion: metallic mesh with gold sintered device is the best of all. We need only a simple parallel plate capacitor geometry to maintain the field lines throughout the entire ADP surface

  2. Dec. 13, 2013 APD 8x8 mm2 Al-mesh, Au sintered, s/n 432-6 APD bias at -1.85 kV 980 nm Vcsel, ~1 ns pulse, 1 kHz ~1 MIP equivalent photons on APD (accounted for the transmission of mesh + etc ~50%) Circuit diagram no amplifier direct APD output 1-MIP at -1.85 kV 256 ave. Mesh APD signal output ground -HV marking on the back

  3. Dec. 13, 2013 APD 8x8 mm2 Al-mesh, Au sintered, s/n 432-6 -1.85 kV ~1 MIP spatial response map ← + y-position ← + x-position

  4. Dec. 13, 2013 APD 8x8 mm2 Al-mesh, Au sintered, s/n 432-6 -1.85 kV ~1 MIP time walk map ← + y-position ← + x-position

  5. Dec. 13, 2013 APD 8x8 mm2 Al-mesh, Au sintered, s/n 432-6 -1.85 kV ~1 MIP time jitter map ← + y-position ← + x-position 1-MIP 8x8mm2σ(0,0)=36.5 ps ~1 MIP σIRF=7.3 ps