An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. PULSION Medical Systems SE PulsioFlex Software V4.0 – New Features

  2. Data Visualization – SpiderVision SpiderVision • SpiderVision screen needs to be activated in the layout settings • Shows dynamic changes of continuous parameters graphically • Screen display and settings similar to PiCCO2

  3. Data Visualization – SpiderVision • SpiderVision configuration • To be opened by touching the Spider diagram • Basic: predefined basic settings • Individual: individual configuration (number of arms and parameters)

  4. Data Visualization – Profile Profile • Profile screen needs to be activated in the layout settings • Static display of the values during the last thermodilution measurement • Screen display and configuration similar to PiCCO2

  5. Data Visualization – Profile • Profile configuration • To be opened by toucing the Profile diagram • Left hand side buttons enables switching between the predefined parameter configurations • Individual configuration is not possible

  6. PiCCO Thermodilution Measurement Values with insufficient pressure curve • In case a calibration cannot be performed due to an insufficient pressure signal, a related message appears on the screen • After confirmation the discontinuous thermodilution results are available

  7. DO2 Calculation • Calculation of oxygen delivery (DO2) is now also available without CeVOX module • Input screen is opened by touching the DO2 button in the main menu • After entering the SaO2 and Hb values DO2 is calculated and displayed continuously

  8. Data Acquisition – Start Trend Data Recording • Activation in the Main Menu • Pushing the Start button starts data recording • All parameter values are stored every 12 seconds for later download • File name is created automatically and indicated on the screen

  9. Data Acquisition – Event Input • Events can always be put in by touching the input field next to ‘Add Event’ (max. 25 characters per event) • All entered events will be stored in the event list • Event inputs while data recording is running will also be stored in the export data file

  10. Data Acquisition – Stop & Data Export • Stop button stops the recording and the save file is closed • Transferable files are listed • Plug in USB storage device to one of the USB ports and push Export Data • Observe screen messages on USB device detection and finishing of data export • File format is ‘.csv’ (comma separated values) which can be opened by spread sheet programs (e.g. MS Excel)