1 / 5

DDC112 Electrometer

DDC112 Electrometer. 13 Feb 2013 D. Peterson. DDC112 Internals. DDC112 Implementation. Pulse out every 1nC. Full scale internal reading is 350 pC in 20 bits. Cycle time is about 600 uS . Max input rate is 583 nA. Max output rate is 70 pulses/sec * 1nC/pulse = 70 nA.

hagen
Télécharger la présentation

DDC112 Electrometer

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. DDC112Electrometer 13 Feb 2013 D. Peterson

  2. DDC112 Internals

  3. DDC112 Implementation • Pulse out every 1nC. • Full scale internal reading is 350 pC in 20 bits. • Cycle time is about 600 uS. • Max input rate is 583 nA. • Max output rate is 70 pulses/sec * 1nC/pulse = 70 nA. • 32 bit accumulator of 20 bit values. Saturates at about 1.43 uC. Takes about 20 sec to dump that at 70 Hz.