1 / 20

Femtosecond-Laser Assisted Cataract Surgery: Is it living up to the hype?

Femtosecond-Laser Assisted Cataract Surgery: Is it living up to the hype?. 117 th Annual Meeting of the American Academy of Ophthalmology Press Briefing Nov. 18, 2013. Introduction of Speakers. Bret L Fisher, M.D., medical director of the Eye Center of North Florida

havard
Télécharger la présentation

Femtosecond-Laser Assisted Cataract Surgery: Is it living up to the hype?

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Femtosecond-Laser Assisted Cataract Surgery: Is it living up to the hype? 117th Annual Meeting of the American Academy of Ophthalmology Press Briefing Nov. 18, 2013

  2. Introduction of Speakers • Bret L Fisher, M.D., medical director of the Eye Center of North Florida • Visual Acuity and Predictability in Femtosecond Laser Cataract Surgery With Intraoperative Aberrometry • Mike P Holzer, M.D., vice chairman and director of refractive surgery, Department of Ophthalmology, University of Heidelberg, Germany • Single-center Contralateral Evaluation of Femtosecond Laser Cataract Surgery Compared with Manual Cataract Surgery • Tim Schultz, M.D., Department of Ophthalmology, University Hospital Bochum, Germany • Toward the Elimination of Ultrasound With Femtosecond Laser Cataract Surgery

  3. Visual Acuity and Predictability in Femtosecond Laser Cataract Surgery With Intraoperative Aberrometry Presented by: Bret Fisher, M.D. The Eye Center of North Florida Panama City, FL

  4. Financial Disclosures • Alcon - Consultant and Lecture Fees

  5. Purpose • To determine the impact of intraoperative aberrometry on uncorrected visual acuity and predictability (MAVPE) in Femtosecond laser cataract surgery

  6. Outcomes analysis in the literature • Narvaez / Stulting JCRS Dec. 2006 • 643 eyes • 46 % within 0.50 D of formula predicted target • MAVPE: 0.52 D ± 0.44 D • Gale et al, Eye Aug. 2007 • 4806 eyes (2688 in the PCI group) • PCI group results • 55 % of eyes within 0.50 D of formula predicted target • 85% of eyes within 1.00 of formula predicted target • MAVPE: 0.55D ± 0.55D • Behndig et al, JCRS July 2012 • 17,506 case (performed between 2008-2010) • 55% achieved refractive target (0.00 D sphere ,<1.00 D of astigmatism • MAVPE: 0.50 D ± 0.34 D Background

  7. Method • One surgeon • No postrefractive cases • Retrospective analysis of eyes within the following groups: • Group 1 – Consecutive cases that had cataract surgery with IOL implantation utilizing a femtosecond laser (LenSx®) without intraoperative wavefront aberrometry (ORA System®) • Group 2 – Consecutive cases that had cataract surgery with IOL implantation utilizing a femtosecond laser (LenSx) in combination with intraoperative wavefront aberrometry (ORA System)

  8. Baseline CharacteristicsGroup 1 – Femtosecond Laser Only • 55 eyes of 55 patients • 23 of 55 (42%) male • 32 of 55 (58%) female

  9. Baseline CharacteristicsGroup 2 – Femtosecond Laser and IntraoperativeAberrometry • 66 eyes of 50 patients • 22 of 50 (44%) male • 38 of 50 (56%) female

  10. Uncorrected Distance Visual Acuity (UCDVA) Results Group 1 - Femtosecond Laser Only n = 55 eyes

  11. Uncorrected Distance Visual Acuity (UCDVA) Results Group 1 - Femtosecond Laser Only Eyes Targeted for Distance n = 49 eyes *Eyes targeted for distance only

  12. Mean Absolute Value Prediction Error (MAVPE) ResultsGroup 1 – Femtosecond Laser Only n = 55 eyes

  13. Uncorrected Distance Visual Acuity (UCDVA) Results Group 2 - Femtosecond Laser and IntraoperativeAberrometry n = 66 eyes

  14. Uncorrected Distance Visual Acuity (UCDVA) Results Group 2 - Femtosecond Laser and IntraoperativeAberrometry n = 60 eyes *Eyes targeted for distance only

  15. Mean Absolute Value Prediction Error (MAVPE) ResultsGroup 2 – Femtosecond Laser and Intraoperative Aberrometry n = 66 eyes

  16. Advances in Intraoperative Aberrometry ORA System with VerifEye • Function: Provides streaming refractive information to the surgeon during preview (prior to measurement) • Provides: High quality measurements • Verifies that the eye is stable and ready for measurement • Result: Greater consistency and accuracy • IOL power recommendations • Guidance for astigmatic measurements • Shorter measurement time • Faster processor • 2 seconds for measurement • 3 seconds for processing

  17. VerifEye

  18. Mean Absolute Value Prediction Error (MAVPE) ResultsGroup 2 – Femtosecond Laser and ORA with VerifEye n=26 eyes

  19. Summary • Femtosecond laser (FSL) assisted cataract surgery provided refraction outcomes that are better that those reported in literature • FSL combined with Intraoperative aberrometry provided better UDVA than FSL alone • Difference in mean acuity is statistically significant (p<0.03) • FSL combined with Intraoperative aberrometry provided better refractive outcomes (as measured by MAVPE) than FSL alone • Difference in mean acuity is statistically significant (p<0.04)

  20. Conclusion • Preliminary results suggest that intraoperative aberrometry can improve results with Femtosecond laser cataract surgery • Proposed addition evaluation • Included a non FSL, non ORA group • Include an ORA only group