An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Setting prescriptive targets Greg Nassar Clinical Scientist MCHAS RECD Refresher Course 17th November 2004

  2. Prior to child arriving • Select a client • PTA (save and close as we don’t have PMS) • Select appropriate hearing aid • Open HA software (new battery, connect leads, switch on then….) • Open REM software • DSL [i/o] or NAL-NL1 set-up screens • Calibrate new probe tube • Prepare coupler response

  3. DSL [i/o] set up on Unity • Access via the screwdriver icon • Check DOB • Transducer- how did you test hearing? • HL to SPL- measured or manual • RE to 2cc- Measured or manaul • Max Output (ULLs) • Speech- UWO is specific child spectrum that takes child’s own voice into account (see Phonak Focus 20) • Circuit- majority of HA on contract have fixed CR. If planning a linear strategy set to LINEAR • Compr. Threshold- minimum or 40-45 if WDRC strategy

  4. DSL [i/o] set up on Aurical • First screen • Check DOB • Transducer- how did you test hearing? • Sound Field (uses REUR for HL- SPL transform) • Insert/custom (used EAR3 with earmould) • TDH- uses average REDD (discussed earlier) • ME Insert (more on this later) • HL to SPL- predicted or measured • HI type • Compression type- as before • Compression Threshold- as before

  5. NAL-NL1 set up on Unity • Similar to DSLi/o except: • Number of hearing aids • Limiting refers to how MPO on hearing aid is adjusted (single channel or multichannel) • Earmould acoustics • Number of compression channels ((new DSL m[i/o]) • Uses BC values (new DSL m[i/o]) • Can measure REIR & REAG

  6. NAL-NL1 set up on Aurical • Same as previous but with additional control to select REAG or REIG

  7. Hello young man! • Otoscopy • Tympanometry (if not performed during behavioural audiometry) • Measure RECD • Save session, then re-open from session list • Send him out/for a break/breather etc… • Attach aid to coupler • Close lid • 50, 65, 80 dB speech noise (modulated on Aurical) • 90 dB pure tone sweep • Program aid

  8. Demo then hands on practical session RECD Refresher Course 17th November 2004