1 / 38

P- and S-wave velocities in rock as a function of pressure and temperature

P- and S-wave velocities in rock as a function of pressure and temperature. I. Lassila 1 ,T. Elbra 2 , E. H æggström 1 and L. J. Pesonen 2 V. Kananen 1 and M. Perä J. Haapalainen 1 and R. Lehtiniemi 3 P. Heikkinen 4 and I. Kukkonen 5. 1 Electronics Research Unit

loring
Télécharger la présentation

P- and S-wave velocities in rock as a function of pressure and temperature

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. P- and S-wave velocities in rock as a function of pressure and temperature I. Lassila 1,T. Elbra 2, E. Hæggström1 and L. J. Pesonen 2 V. Kananen1 and M. Perä J. Haapalainen1 and R. Lehtiniemi 3 P. Heikkinen 4 and I. Kukkonen 5 1 Electronics Research Unit 2 Division of Geophysics 3 Nokia Research Center 4 Institute of Seismology 5 Geological Survey of Finland

  2. Motivation - Understanding the structure of the earth’s crust • FIRE (Finnish Reflection Experiment) - project • Seismic reflection and refraction measurements (longitudinal and shear wave modes) Photo: Seismic signal is produced by vibrators. Courtesy Jukka Yliniemi. Location of the FIRE reflection seismic lines.

  3. TOF and depth • Seismic measurements give TOF data • Need to know Vp and Vs to calibrate the depth Example of FIRE results from the end of line FIRE 3A in western Finland. The reflector amplitudes of a migrated section are presented as gray tone intensities.

  4. Samples • Outokumpu Deep Drilling Project (2516 m)

  5. Device: requirements • Vp and Vs measurements • preferably simultanously • 10 m/s accuracy • Controlled pressure • 0 - 300 MPa (15 ton for OKU samples) • Controlled temperature • 20-300ºC • Data acquisition • Preferably automatic 22 mm 25 mm

  6. Possible measurement setups • Uniaxial • Multianvil • Hydrostatic pressure

  7. Timetable Jan Feb Mar Apr May Jun Jul Material considerations Mechanical design Ultrasonic testing and designing Transducers, pulser / signal generator, amplifiers, switches, oscilloscope Pressure generating Pressure monitoring Heating Temperature monitoring Transducer cooling Ordering parts Planning the measurement procedure Assembling the setup Programming the DAQ software Validation

  8. Device: Vp and Vs • Pitch-catch method • Two similar transducers, both comprising shear (1,1 MHz) and longitudinal (1 MHz) piezo (Pz-27) ceramics • At first only the shear crystal was in use • Longitudinal mode well present • Caused by silver epoxy? • Removable delay lines • Fused quartz • Brass • Water cooling • No load over the piezo crystal

  9. Device: pressure simulation • Pressure simulations by Mr. Haapalainen • Device can withstand the required pressure • Fused quartz can be used as a delay line material in case of no roughness

  10. Device: pressure • Generating: 15 ton jack borrowed from Department of Chemistry • Measuring: Sensotec Model 53 (max 23 ton) + Lebow 7528 amplifier

  11. Device: pressure • Problem with sample durability • Solved with a brass jacket • Splitting sample holder allows sample removal after compression

  12. Device: Temperature, simulations • Thermal simulations by PhD Lehtiniemi and Mr. Haapalainen • 160 W heater is sufficient for 300ºC in case of fused quartz delay lines • Transducer temperature stays below solder melting / epoxy softening temperature

  13. Device: Temperature • Heating: Nozzle heater ACIM T197 (160 W / 240 Vac) • Max 400ºC • Covers the sample holder • Cooling: Water cooler (Lauda WK502) • Measuring: Custom AD595 based thermocouple amplifier • K-type Thermocouple inside the sample holder

  14. Device: Data acquisition • US signals: • 5072 PR, LeCroy 9410, GPIB, PC, LabVIEW, Matlab • Thermocouple and load cell: • AD-conversion and transfer to PC with NI PCI-6024E

  15. Device • Transducers • Delay lines • Heating element and sample • Thermocouple • Load cell • Water cooling tubes • Jack

  16. Preliminary results • 7 samples from Outokumpu Deep Drill Core • T: 300ºC20ºC, Load: 7000 kg  500 kg (resembling the conditions in the Earth’s crust) • Results comparable with literature values

  17. Pressure test • The error if we don’t measure the compression of the sample? • Compression = 0,1 mm (Δhsample- Δhno sample) • Error Vp = 24-33 m/s • Error Vs = 15-18 m/s

  18. TOF (time of flight) through the delay lines • Pulse-echo measurement of the delay line • Subtraction of the TOF through the delay lines from the total TOF • Pressure and temperature effects to the delay lines and transducers are cancelled

  19. Damping the transducers • Ringing of the piezo element makes pulse-echo (PE) measurements difficult. • Ringing can be reduced with applying attenuating, material with acoustic impedance close to the piezo to the back side of the transducer • PE responses to water load • a) zero backing, b) backing of crown glass, c) backing of tungsten-epoxy, d) backing of material with Z=Ztransducer Egypt. J. Sol., Vol. (23), No. (2), (2000)

  20. Damping test - ok • Reduced ringing time and increased bandwidth

  21. Transducers without backing

  22. Outcome of applying the backing • No signal • Resistance between transducer electrodes ca. 5 Ω  Short-circuit • Difference between test • Amount of tungsten in the mixture was higher • In the test the resistance between the electrodes was ca. 500 Ω • This type of backing method requires isolation of the electrodes • Instead of scraping out the backing it was decided to build new transducers

  23. New transducers • Increased sample size: • Height 20-70 mm • Diameter 25-62 mm • Better modal purity required • Mode conversion in the gap between transducer housing and delay line • Material: stainless steel • No separate delay lines

  24. New transducer drawings

  25. New transducer

  26. New thermal simulations • Stainless steel: thermal conductivity=20 W/(m K) Specific heat=500J/(kg K) • Sample (rock): thermal conductivity=2 W/(m K) Specific heat=790J/(kg K) T=? T=10ºC h = 20-70mm T(t=0)=350ºC D = 25-62 mm

  27. Temperature as a function of time in the middle of the sample and on the transducer inner surface where the piezos are fixed. Sample D = 25,5 mm, h = 24 mm Sample D = 62 mm, h = 70 mm

  28. Temperature distribution in the sample and the upper transducer Sample D = 25,5 mm, h = 24 mm t = 200s. Sample D = 62 mm, h = 70 mm, t = 400s.

  29. Other updates • PC controlled pressure generation • Separate heating of samples to increase the throughput rate

  30. New frame • Compressed air controlled one way hydraulic cylinder replaced with electric motor controlled two way hydraulic cylinder

  31. Modification for hydraulic control • Controls of the pump replaced with relay circuit that is controlled from PC DAQ-card • Two valves that are controlled • Valve 1 open increasing pressure • Valve 2 open decreasing pressure • Valves closed no change

  32. Testing new hydraulics • Pressure increase at 0,1 s intervals • OK for loads over 3000 kg

  33. Testing of new hydraulics • Pressure decrease at 0,1 s intervals • No control of outcome when decreasing pressure

  34. More control needed • Manual shut off valve, needle type control • Slows down the flow of the hydraulic oil

  35. Control achieved • Needle valve can be adjusted to allow precise control of the load

  36. Measurement diagram

  37. Conclusions • Device is used for measuring Vp and Vs values that are needed to interpret seismic data • Preliminary results ok • At the moment system is going through some changes

  38. Future tasks • Temperature inside the sample vs. on the sample surface • Validation tests • Implement a LVDT/gauge to measure the sample thickness and thickness change inline • Licentiate thesis

More Related