1 / 23

The AMO Instrument John Bozek, LCLS

The AMO Instrument John Bozek, LCLS. Location Change Scientific Motivation Design Status Next Steps: Procurement/fabrication Assembly Testing Controls Schedule. The AMO Instrument John Bozek, LCLS. Represents work of many people:

lucretiac
Télécharger la présentation

The AMO Instrument John Bozek, LCLS

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The AMO Instrument John Bozek, LCLS • Location Change • Scientific Motivation • Design Status • Next Steps: • Procurement/fabrication • Assembly • Testing • Controls • Schedule

  2. The AMO InstrumentJohn Bozek, LCLS • Represents work of many people: • Team leaders: Lou DiMauro, Nora Berrah, Linda Young, Phil Bucksbaum • Engineering: Jean Charles Castagna, Michael Holmes, Michael Kosovsky, Nadine Kurita • Laser: Greg Hayes, Bill White • LCLS Management: Stefan Moeller, John Arthur • With apologies to whomever I’ve forgotten

  3. LCLS Experimental Halls Near Experimental Hall X-ray Transport Tunnel AMO SXR XPP XCS CXI Far Experimental Hall

  4. Location of AMO experiment has changed… • AMO on soft x-ray branch 800-2000 eV

  5. Unique properties of LCLS radiation

  6. Scientific Goals of the AMO instrumentation • Investigate multiphoton and high-field x-ray processes in atoms, molecules and clusters • Multi-photon ionization/excitation in atoms/molecules/clusters • Accessible intensity on verge of high-field regime • Study time-resolved phenomena in atoms, molecules and clusters using ultrafast x-rays • Inner-shell side band experiments • Photoionization of aligned molecules • Temporal evolution of state-prepared systems

  7. AMO Instrument • Optics • High Field Physics Chamber • Gas jet • Electron TOFs • Ion spectrometers • X-ray spectrometers • Chamber, shielding, pumping, shutter, etc • Diagnostics Chamber • Magnetic bottle electron spectrometer • Beam screens • Pulse energy monitor • Pulse picker

  8. AMO Instrument

  9. Final Design Review of Stands & Chambers • July 18th

  10. Considered beam mis-steer Diagnostics HFP Beam stop Beam direction B4C aperture B4C aperture Differential pumping B4C tube Differential pumping B4C tube

  11. Final Design Review of HFP Instrumentation • September 3rd 5 x ETOF High field physics chamber instruments • 5 x ETOF adjustable aiming at IR • 1 ITOF with 3 options • Gas jet with adjustable positioning Interaction Region X-ray beam Gas jet Ion TOF (3 options)

  12. Final Design Review of Beam path components – Oct 1st Beam, fixed viewer 3 Beam, fixed viewer 2 Beam, fixed viewer 1 Differential pumping 3 Beam viewing screen 1 Beam viewing screen 2 Differential pumping 2 Differential pumping 1 Beam viewing paddle 1

  13. Remaining work to be Reviewed: • Focusing optics • Design/fab under contract with LBNL (Kelez) • Substrates ordered – 400mm long Si flats • Review expected in November • Magnetic Bottle Electron Spectrometer • Design/fab contract with Ohio State University • Already have a prototype – testing with laser • Review also expected in Nov.

  14. Remaining work to be Reviewed: • Fluorescence Spectrometers • Late deliverable (i.e. not for first science) • Difficulty determining appropriate solution…grating vs. crystal, etc. • Slits, pulse picker • Recently separated slits from mirror tank – allowing us to pursue commercial solution • Pulse picker commercially available – late deliverable

  15. Procurement • Utilize the LCLS procurement team • Hoping to use the Mechanical Fabrication Department (MFD) of SLAC to manage procurement/assembly • MFD tracks parts from order through assembly

  16. Assembling the AMO Instrument • Assembly in the MFD vacuum shops by MFD vacuum technicians • Clean rooms • Vacuum bake-out • Welding/machining

  17. Testing – first step is getting tables moving Moving frame Slave stages XY Inclinometer Spherical bearing joints Motorized stages XY Stiffening struts Motorized stages Z Anchored frame Wheeled Jacks (Alternatively, can mount on sides)

  18. Tables interlocked to two mechanisms: • Bellows protection mechanism • Tilt switch to prevent roll

  19. Controls & Data Acquisition • EPICS will be used to control instrumentation • All motions, voltages, etc. under remote control • Test-stands in G. Haller’s labs

  20. Controls & Data Acquisition • Acqiris DC282 high-speed 10-bit cPCI Digitizer (8) • 4 channels • 2-8 GS/s sampling rate • Acquisition memory from 1024 kpoints to 1024 Mpoints (optional)‏ • Low dead time (350 ns) sequential recording with time stamps • 6U PXI/CompactPCI standard, 64 bit, 66 MHz PCI bus • Sustained transfer rate up to 400MB/s to host SBC

  21. What’s there July 2009? • Focusing Optics • High Field Physics Chamber • Pulsed gas jet • Ion TOF (Wiley McLaren type) • Electron TOFs • Diagnostics • Effusive gas jet • Magnetic bottle TOF • Beam screens • Synchronized Laser

  22. And what’s commissioned later ? • Pulse picker(12/09) • High Field Physics Chamber • Velocity map imaging ion spectrometer(10/09) • Momentum resolving ion spectrometer (reaction microscope/ColTRIMS)(2/10) • X-ray spectrometers(1/10) • Diagnostics • Pulse energy monitor(11/09)

  23. Schedules • Ready to start ordering equipment – now! • Jan - May 2009 – Assembly • Feb - May 2009 – offline testing of controls • May 2009 – install in hutch • June 2009 – laser based testing • August 2009 – FEL testing • September 2009 – User Operations !