1 / 14

POST MARKET CLINICAL FOLLOW UP

14.05.2012. Abteilung:. 2. 2007/47/EG nderungen Klin. Bewertung. Historie:CETF Report (2000)Qualitt der klinischen Daten zu schlechtZu wenige klinische Prfungen Im Vorgriff auf die Revision der MDD publiziert:MEDDEV 2.7.1. Clinical EvaluationMEDDEV 2.12-2 Post Market Clinical Fo

mead
Télécharger la présentation

POST MARKET CLINICAL FOLLOW UP

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


    1. POST MARKET CLINICAL FOLLOW UP (MEDDEV 2.12-2 May 2004)

    2. 15.05.2012 Abteilung: 2 2007/47/EG nderungen Klin. Bewertung Historie: CETF Report (2000) Qualitt der klinischen Daten zu schlecht Zu wenige klinische Prfungen Im Vorgriff auf die Revision der MDD publiziert: MEDDEV 2.7.1. Clinical Evaluation MEDDEV 2.12-2 Post Market Clinical Follow Up

    3. Post market surveillance may include: active supervision by customer surveys inquiries of users and patient literature reviews post market clinical follow up PMCF MEDDEV 2.12-2

    4. Annex II: 5.2. The manufacturer must authorize the notified body to carry out all necessary inspections and supply it with all relevant information, in particular [...] the data stipulated in the part of the quality system relating to design, such as the results of analyses, calculation test, the solutions adopted in accordance with Annex I section(2), preclinical and clinical evaluation, post-market clinical follow-up and results of the post-market clinical follow up, if applicable [...] 2007/47/EG nderungen Klin. Bewertung

    5. 15.05.2012 Abteilung: 5 Klinische Daten Anhang X Die klinische Bewertung und ihre Dokumentation mssen aktiv anhand der aus der berwachung nach dem Inverkehrbringen erhaltenen Daten auf dem neuesten Stand gehalten werden Wird eine klinische berwachung PMCF Post Market Clinical Follow UP nach dem Inverkehrbringen [] nicht fr erforderlich gehalten, muss dies ordnungsgem begrndet und dokumentiert werden.

    6. PMCF should always considered for devices where: In the case the assessment of a product is performed through the concept of equivalence (literature route !) identification of possible emerging risks is critical the evaluation of long term safety and performance is critical (implants !) PMCF MEDDEV 2.12-2

    7. Criteria to identify emerging risk: innovation severity of the disease sensitive target population risky anatomical site well known risks from literature or marketed devices obvious discrepancy between the pre-market follow up timescale and the expected life of the product PMCF MEDDEV 2.12-2

    8. All PMCF should be planned: form of extended follow-up of patients enrolled in the pre-market trials and/or a prospective study of a representative subset of patients after the device is placed on the market PMCF MEDDEV 2.12-2

    9. This plan will need to take into account: Results of the clinical investigation including Adverse events identified Average life expectancy of the device The claims made by the manufacturer for the device Performances for which equivalence is claimed New information becoming available PMCF MEDDEV 2.12-2

    10. No PMCF: Products for which the medium/long term clinical performance and safety is already known from previous use of the device or from fully transferable experience with equivalent devices (Products quoted as "equivalent" devices where reference product is subjected to PMCF) PMCF MEDDEV 2.12-2

    11. Conclusion PMCF Plan Follow up of a pre-market trial Post-market prospective trial (cave: national requirements!) The notified body should review the appropriateness of the plan and the procedures The follow-up should take into account the average life expectancy In the case the assessment of a product is performed through the concept of equivalence, PMCF should always be considered. PMCF MEDDEV 2.12-2

    12. Mglichkeiten eines PMCF: Fortfhrung einer klinischen Prfung (z.B. Herzklappen) Planung einer neuen klinischen Prfung, klinischen Studie (Achtung bei zustzlichen invasiven oder belastenden Untersuchungen - MPG 23) Durchfhrung einer sogenannten Registry 15.05.2012 Abteilung: 12 Post Market Clinical Follow Up

    13. Requirements for documentation of Post Market Clinical Follow Up (PMCF) activities: NOTE: Results of PMCF and link to risk management are not addressed in MEDDEV 2.12-2 PMCF MEDDEV 2.12-2

    14. 15.05.2012 Abteilung: 14 PMCF MEDDEV 2.12-2

    15. Content of an integrated Risk Management File: NOTE: Timeframes for Risk Management File update (addressed in RM SOP / RM Plan) should be in compliance with PMCF plan and design of clinical follow up studies PMCF MEDDEV 2.12-2