An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. A (quick) overview of polysomnographic artefacts Jo Tiete Centre Hospitalier Luxembourg

  2. polysomnographic artefacts 1 Subject Induced Artefacts or Physiological Artefacts

  3. Skin artefact Symptoms • The skin is considered to contribute the greatest possible distortion of brain potentials. • The factors that effect skin impedance are its: • thickness, • the degree of cellular hydration, • the number of sweat glands • and hair follicles which provide low-resistance pathways into the skin. 7th ESST Meeting - Jo Tiete CHL

  4. Skin artefact Actions: • Proper preparation of the skin is the sleeptech most important concern. • Degreasing with aceton/ether. • Rubbing the skin (peeling), but with care… for the subject and for the sleeptech! (blood born viruses). 7th ESST Meeting - Jo Tiete CHL

  5. Sweat artefact Symptoms • Low frequency baseline swings that may affect one or more channels. • Is due to transpiration and is a combination of a salt-bridge between electrodes and the skin. 7th ESST Meeting - Jo Tiete CHL

  6. Sweat artefact Actions: • Identify the electrode and eliminate it by making appropriate changes to the channel’s input selector or replace electrode. • Cool the subject, by reducing the room T°, use a fan or remove blankets. • Place an absorbing towel under the subject’s head. • Use an anti-perspirant aerosol on the area around the electrode. • As last resort, you can reduce the low-frequency filter. 7th ESST Meeting - Jo Tiete CHL

  7. ECG artefact Symptoms • Is most prominent in subjects who are obese or have short, thick necks. • Often when using ear reference montages (A1-A2) and if the input impedance is unbalanced. • If EKG occurs on the EEG channels, you need toeliminate it to avoid misinterpretation. 7th ESST Meeting - Jo Tiete CHL

  8. ECG artefact Actions • Reposition or move the mastoid reference (A1 & A2) electrodes to the ear lob. • Link the 2 mastoids (A1<->A2) with a jumper cable, doing so will most likely alter the amplitude of the waveforms. • Lowering the high-freq filter is not a good practice! 7th ESST Meeting - Jo Tiete CHL

  9. Pulse artefact Symptoms • It is a mechanical artefact that is picked up because the recording electrode is positioned over an artery. • Monitoring EKG is useful. • R phase of EKG wave occurs at the same point in each EEG slow wave. 7th ESST Meeting - Jo Tiete CHL

  10. Pulse artefact Actions • Reposition the subject’s head. • Switch the amplifier’s input configuration. • Ad more electrolyte gel to the electrode. • Reposition the electrode. 7th ESST Meeting - Jo Tiete CHL

  11. Cardio ballistic artefact Symptoms • Cardiac pumping is detected on respiratory flow/effort channels during apneas. • Related to pulse artefact. • Also on oesophagial pressure channels but sometimes on any high impedance EEG channel. • Mostly seen in thin/skinny individuals. 7th ESST Meeting - Jo Tiete CHL

  12. Cardio ballistic artefact Actions • Repositioning the subject’s head. • Raising the subject’s head off the bed with a roll of towels may eliminate it. • Eliminate artefact by repositioning the oesophageal pressure catheter. 7th ESST Meeting - Jo Tiete CHL

  13. Pacemaker artefact Symptoms • This artefact is due to an electrical pulse of the pacemaker. • Can be periodic or intermittent and may appear in one or more channels. • May look like spikes or sharp waves in the EEG. 7th ESST Meeting - Jo Tiete CHL

  14. Pacemaker artefact Actions • Can not be eliminated without stopping the pacemaker. • Monitor the ECG channel to identify. 7th ESST Meeting - Jo Tiete CHL

  15. Eye movement artefact Symptoms • Eyes have a strong polar field (cornea-fundal & cornea-retinal potential). • Is of sufficient strength to introduce slow wave artefact into frontal & temporal EEG signals. • But useful for detecting stage 1 or REM. 7th ESST Meeting - Jo Tiete CHL

  16. Eye movement artefact Actions • There is no correction for this artefact. 7th ESST Meeting - Jo Tiete CHL

  17. Eye blink artefact Symptoms • Same process as eye movement artefact, but always in awake. • Blinking artefact can pollute the EEG tracings but are useful because the frequency of eye blinks decrease with drowsiness prior to sleep onset. 7th ESST Meeting - Jo Tiete CHL

  18. Eye blink artefact Actions • There is no correction for this artefact. 7th ESST Meeting - Jo Tiete CHL

  19. Unilateral REM Symptoms • Subject has lost functionality of one eye. • One EOG electrode is suspected of poor quality or high impedance. 7th ESST Meeting - Jo Tiete CHL

  20. Unilateral REM Actions • Ophtalmic exploring (glass eye ?). • Verify or change bad electrode, if electrode is the cause. 7th ESST Meeting - Jo Tiete CHL

  21. Movement artefact Symptoms • High amplitude slow waves sometimes with amplifier blocking in the EEG channels. • During a position change or limb movement or the characteristic head, neck and mandible movements seen in OSA. • Cause electrode popping and amplifier blocking. 7th ESST Meeting - Jo Tiete CHL

  22. Movement artefact Actions • Wrap electrodes and transducers together and form a neat bundle (poney tail) to reduce this type of artefact. • Avoid suspending the head box over the bed’s headboard with all wires in bed. • Better have the electrode wire bundle rest on the bed without the subject laying on the wires. 7th ESST Meeting - Jo Tiete CHL

  23. Tremor artefact Symptoms • 4-8 Hz that can be confused with theta. • In Parkinson’s disease. • Decrease with sleep onset but tremors may remain present during stage two or reappear during REM. • Plethysmography belts can also pick up this artefact. 7th ESST Meeting - Jo Tiete CHL

  24. Tremor artefact Actions • Repositioning the subject’s head to avoid contact with the bed will reduce but not eliminate this artefact. • Nothing that can be done to eliminate tremors detected by inductive plethysmography belts. 7th ESST Meeting - Jo Tiete CHL

  25. Body rocking & Head banging artefact Symptoms • Both disorders produce a periodic occurrence of movement artefacts. • Sometimes with amplifier blocking which in most cases affects all EEG channels. 7th ESST Meeting - Jo Tiete CHL

  26. Body rocking & Head banging artefact Actions • There is no correction for this artefact. • Bundle electrode wires. 7th ESST Meeting - Jo Tiete CHL

  27. Muscle artefact Symptoms • Can appear in all channel. • Can mimic cortical spikes If continuously present. • Can be used to identify bruxism, movement arousals, vocalization, motion, tension, etc. • Can render portions of the record unscorable. 7th ESST Meeting - Jo Tiete CHL

  28. Muscle artefact Actions • There is no correction for this artefact. 7th ESST Meeting - Jo Tiete CHL

  29. Respiration artefact Symptoms • Low frequency baseline sway in EEG. • Occurs in-phase with the respiration tracing. • Artefact is positional. • Caused by body movements associated with respiration. 7th ESST Meeting - Jo Tiete CHL

  30. Respiration artefact Actions • repositioning the subject’s head or the electrode wires. • Keep wires bundled and from under the subject’s head or body. 7th ESST Meeting - Jo Tiete CHL

  31. Fish-mouthing artefact Symptoms • False "breaths" recorded by thermal airflow probes. • Misclassify obstructive apnoea as an obstructive hypopnoea. • Produced by air being drawn into the mouth as the mandible retracts during a blocked inspiration. 7th ESST Meeting - Jo Tiete CHL

  32. Fish-mouthing artefact Actions • There is no correction for this artefact. 7th ESST Meeting - Jo Tiete CHL

  33. Snoring artefact Symptoms • Mostly, but not only, in EMG. • Indicate an increase in respiratory drive. • Sufficient to incorporate a secondary inspiratory musculature. • Is considered as a physiological response to increased pCO2 levels. 7th ESST Meeting - Jo Tiete CHL

  34. Snoring artefact Actions • There is no correction for this artefact. 7th ESST Meeting - Jo Tiete CHL

  35. Gurgle artefact Symptoms • Appears as spindle-like activity in the EEG channels on each inhale or exhale.  • Seen in comatose subjects.  • Can affect the estimated airflow signal.  • Can occur when CPAP machine is used with a humidifier in a cold room (condensation). 7th ESST Meeting - Jo Tiete CHL

  36. Gurgle artefact Actions • Swallowing clear the airway, but comatose subject may require suctioning of the airway. • When present in the estimated airflow channel of a CPAP you need to drain the condensed water from the hose.  • It may also be helpful to insulate a portion of the CPAP tubing (decrease condensation).  • Increase the room temperature in order to reduce condensation. 7th ESST Meeting - Jo Tiete CHL

  37. Genioglossal artefact Symptoms • The subject’s tongue is an electrically polarized mass that can introduce slow wave artefact into EEG channels.  • Raising the tip of the tongue to the roof of the mouth can generate a potential of 100 micro volts at the vertex.  • This artefact has been described as resembling intermittent, rhythmic delta. 7th ESST Meeting - Jo Tiete CHL

  38. Genioglossal artefact Actions • There is nothing you can do to prevent this artefact. 7th ESST Meeting - Jo Tiete CHL

  39. polysomnographic artefacts 2 Artefacts that arise from subject or the recording Equipment.

  40. Bi-metallic artefact Symptoms • Blocked Ionic flow between different metals that discharge suddenly. • Oral appliances, dental fillings can produce an intermittent artefact that presents as sharp waves or spikes. (=subject). • If different metals, scratched silverchloride or gold electrode are used. (= environment). 7th ESST Meeting - Jo Tiete CHL

  41. Bi-metallic artefact Actions • Record the discharges on a separate channel by positioning electrodes over the location of the dental fillings, or metal components of appliances to identify the artefact. • Chloride the scratched (silver) electrode. • Discard scratched gold electrodes. • Use of « disposable » electrodes. 7th ESST Meeting - Jo Tiete CHL

  42. polysomnographic artefacts 3 Recording Equipment or Environmental Induced Artefacts.

  43. 50 or 60 Hz Artefact Symptoms • 50 (60Hz) interference was very common on analog amplifiers (bad common mode rejection). • Source is environmental, but can also indicate malfunction or improper installedequipment.  • Electromagnetic radiation from the building’s electric wiring and from devices powered with these wires. • At the frequency of the alternating current. 7th ESST Meeting - Jo Tiete CHL

  44. 50 Hz Artefact Actions • Check impedance and balance impedance between electrodes, specially Ref. or ground. • Apply gel or place fresh electrode(s). • Verify if electrode cables are not lose and bundle them. • All conductive surfaces (antennas!) and electric appliances near subject should be grounded to the same ground. • Keep bio-signals and power leads far away. • Better a non-motorized wooden than a metal bed. • Don’t make loops in power cords. • Use a « Fahraday cage ». • As a last resort, you can use the 50/60Hz notch filter. 7th ESST Meeting - Jo Tiete CHL

  45. Salt-bridge Artefact Symptoms • When excessive amount of electrode gel spreads between two electrodes forming a conductive pathway. 7th ESST Meeting - Jo Tiete CHL

  46. Salt-bridge Artefact Actions • Switch to a backup electrode. • Cleaning the skin surface between the electrodes. • Remove the electrodes, clean the skin then, reapply the electrodes. 7th ESST Meeting - Jo Tiete CHL

  47. Amplifier-blocking artefact Symptoms • Excessif amplification leads to signal clipping or amplifier blocking. • Pen deflection is maximal for an amount of time. • No interpretation possible on one or more channels. 7th ESST Meeting - Jo Tiete CHL

  48. Amplifier-blocking artefact Actions • Reduce the channel’s sensitivity. • Wait till channels set. 7th ESST Meeting - Jo Tiete CHL

  49. Electrode-popping artefact Symptoms • Is usually intermittent. • Can be mistaken as spike trains, sharp waves and even K-Complexes. • The cause is most often a loose electrode or broken electrode wire (corrosion). 7th ESST Meeting - Jo Tiete CHL

  50. Electrode-popping artefact Actions • Switch the input to a backupelectrode.  • Repair or replace the electrode who suffer from high impedance. • Change the broken or corroded electrode lead. 7th ESST Meeting - Jo Tiete CHL