An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. TURKEY REPUBLIC MINISTRY OF ENVIRONMENT AND FORESTRY TURKISH STATE METEOROLOGICAL SERVICE DIRECTORY OF METEOROLOGICAL INSTRUMENTS AND CALIBRATION ECO CALIBRATION CENTER CalibrationCenter - Ankara 2010

  2. Calibration Center At the Calibration Center: Automated Weather Observing System's sensors that is installed to all around the Türkiye are calibrated by applying the 17025 standards. Calibration Center is accreditedfor the calibrations of tempetature, relative humidity, air pressure and wind speed sensors. Calibration Center - Ankara 2010

  3. TSMS’s Calibration Center acts as Economic Cooperation Organisation (ECO) calibration center. It would provide training facilities for the member states. Member States: 1- The Islamic Republic of Afghanistan . 2- The Republic of Azerbaijan. 3- The Islamic Republic of Iran . 4- The Republic of Kazakhstan . 5- The Kyrgyz Republic . 6- The Islamic Republic of Pakistan . 7- The Republic of Tajikistan . 8- The Republic of Turkey . 9- The Republic of Turkmenistan . 10-The Republic of Uzbekistan . Calibration Center - Ankara 2010

  4. CALIBRASTION CENTER ORGANIZATIONAL CHART Calibration Center - Ankara 2010

  5. DIRECTORY OF METEOROLOGICAL INSTRUMENTS and CALIBRATION PRECIPITATION WIND Calibration Center operates under Meteorological Instruments and Calibration Division. TEMP&HUM SOLAR RADIATION Mr. Nur DOĞAN Manager of Meteorological Instruments and Calibration Division PRESSURE Calibration Center - Ankara 2010

  6. QUALITY MANAGER Calibration Center of all activities;providing work in accordance with the rules of the17025 standard, providing continuous development of Laboratory and providing relations with the TURKAK duty and authorized to Institution to represent in TURKAK . Mr. Yusuf Özcan CANDEMİR QUALITY MANAGEMENT Calibration Center - Ankara 2010

  7. CHIEF OF CALIBRATION LABORATORIES • Temperature & Humidity Calibration Laboratory, • Pressure Calibration Laboratory, • Wind Speed Calibration Laboratory, • Radiation Calibration Laboratory, • Precipitation Calibration Laboratory. Mr. Tekin SOYLU Seniorchief of theLaboratories Calibration Center - Ankara 2010

  8. CALIBRATION CENTER PERSONNEL CHART Calibration Center - Ankara 2010

  9. INSTRUMENT ACCEPTANCE & REGISTRATION AND ARCHİVE Instruments accepted to the Calibration Center are registered and related documents are archivedin accordance with the rules of 17025 standards. Calibration Center - Ankara 2010

  10. CALIBRATION CENTER ENVIRONMENTAL CONDITIONS OF THE LABRATORIES Environmental conditions of the Laboratories are controlled by using precision air-conditioner. The measured data of temperature and humuidity are automatically displayed and stored. Calibration Center - Ankara 2010

  11. CALIBRATION CENTER ENVIRONMENTAL CONDITIONS OF THE LABRATORIES Devices to control the laboratory environmental conditions: Fluke Hart Scientific 1620A Measuring, displaying and recording. Liebert Hiross HPM Precision Air conditioner Calibration Center - Ankara 2010

  12. TEMPERATURE & HUMIDITY LABORATORY • The thermometer typesto be calibrated: • Liquid glass thermometers, • Resistance thermometers, • Analog output thermometers, • Thermometerswithdisplay. All types of relative humidity sensors are calibrated. Calibration Center - Ankara 2010

  13. TEMPERATURE & HUMIDITY LABORATORY Scope of Accreditation Calibration Center - Ankara 2010

  14. TEMPERATURE & HUMIDITY LABORATORY Reference devices for relative humidity calibration: Thunder Scientific 2500 Humidity Generator. Operation range: -10˚C / +70˚C 10%RH / 90%RH Vaisala HMP76 Calibration Center - Ankara 2010

  15. TEMPERATURE & HUMIDITY LABORATORY Climate provider for Relative Humidity Calibration: Weiss SB22 500 Climate Chamber. Operation Range: -50˚C / +50˚C 10%RH / 90%RH Calibration Center - Ankara 2010

  16. TEMPERATURE & HUMIDITY LABORATORY Reference devices for Temperature Calibration: Isotech 909Q Standard Platinum Resistance Thermometer Hart Scientific 5615 Platinum Resistance Thermometer Calibration Center - Ankara 2010

  17. TEMPERATURE & HUMIDITY LABORATORY Climate provider for Temperature Calibration: Weiss SB22 500 Climate Chamber. Operation Range: -50˚C / +50˚C 10%RH / 90%RH Fluke 7381 Liquid bath. Operation range: -80˚C /+110˚C Calibration Center - Ankara 2010

  18. TEMPERATURE & HUMIDITY LABORATORY Secondary devices for Temperature & Humidity Laboratory: Fluke PRT measuring unit, Zeno 3200 Data Logger and Fluke 8846A 6 ½ digit multimeter are used to measure the data. Calibration Center - Ankara 2010

  19. TEMPERATURE & HUMIDITY LABORATORY Calibration at The Triple Point of Water (TPW) and Ice Point Calibration : Calibration Center - Ankara 2010

  20. PRESSURE CALIBRATION LABORATORY • Pressure sensors typesto be calibrated: • Frequency output barometers, • Digital output barometers, • Barometerswith digital display, • Mechanical barometers. Calibration Center - Ankara 2010

  21. PRESSURE CALIBRATION LABORATORY Scope of Accreditation Calibration Center - Ankara 2010

  22. PRESSURE CALIBRATION LABORATORY Reference device for Pressure Calibration: Parosceintific 785 digiquartz barometer Calibration Center - Ankara 2010

  23. PRESSURE CALIBRATION LABORATORY Climate providers for Pressure Calibration: Fluke PPC4-ui A100K Calibrator. Dry air are used (Azoth gas). Pressure Chamber Calibration Center - Ankara 2010

  24. PRESSURE CALIBRATION LABORATORY Secondary devices of Pressure Laboratory: Druck DPI 145 Measurement device for frequency output barometer. Measurement device for digital output barometer. Calibration Center - Ankara 2010

  25. PRESSURE CALIBRATION LABORATORY Secondary devices of the Pressure Laboratory: Paroscientific 765 digiquartz barometer. Ruska 6200 barometer. Calibration Center - Ankara 2010

  26. WIND SPEED CALIBRATION LABORATORY • WindSpeed sensor typesto be calibrated (Cup, hotwire, ultrasonic, pitotand propelleretc.): • Electronics wind speed sensors, • Mechanical wind speed sensors. Calibration Center - Ankara 2010

  27. WIND SPEED CALIBRATION LABORATORY Scope of Accreditation Calibration Center - Ankara 2010

  28. WIND SPEED CALIBRATION LABORATORY Reference devicesfor Wind Speed Calibration: Furness Control FCO510 Micromanometer and Pitot tube. Thies 2D 4.3810.00.340 Ultrasonic wind speed sensor Calibration Center - Ankara 2010

  29. WIND SPEED CALIBRATION LABORATORY Medium provider for Wind Speed Calibration: Wind Speed Tunnel is the Göttinger type, closed loop and open test area ( 0 – 40 m/s). Calibration Center - Ankara 2010

  30. WIND SPEED CALIBRATION LABORATORY Devices of the Wind Speed Calibration Laboratory: Hotwire Testo 435 Wind Speed Sensor. Druck PTX7517-1 Pressure (600–1100mbar) PR 5333A Temperature (0-40ºC) sensors. Calibration Center - Ankara 2010

  31. WIND SPEED CALIBRATION LABORATORY Devices of the Wind Speed Calibration Laboratory: Vaisala WMS302M and WAA151, Thies Clima 4.3303.22.007, Qualimetrics ECN4738, Siap handel type, Vector İnst. A100M/HE-1 and Lastem DN002 Wind Speed sensors are used. Calibration Center - Ankara 2010

  32. RADIATION CALIBRATION LABORATORY Pyranometers are calibrated at the Radiation Calibration Laboratory under the constant irradiance and new sensitivity coefficients for the calibrated sensors are calculated. Calibration Center - Ankara 2010

  33. RADIATION CALIBRATION LABORATORY Scope of the Calibration Calibration Center - Ankara 2010

  34. RADIATION CALIBRATION LABORATORY Reference device for Pyranometer Calibration: Kipp&zonen CMP22 Pyranometer Calibration Center - Ankara 2010

  35. RADIATION CALIBRATION LABORATORY Medium provider for Pyranometers Calibration: Kipp&Zonen Calibration Facility for Pyranometers. Indoor Pyranometer Calibration System under constant irradiance light (0 - 700W/m²). Calibration Center - Ankara 2010

  36. RADIATION CALIBRATION LABORATORY Devices of the Pyranometers Calibration Laboratory: Kipp&Zonen CMP21 Pyranometer Keithley 2000 6 ½ digit multimeter Calibration Center - Ankara 2010

  37. PRECIPITATION CALIBRATION LABORATORY • Precipitation (Rain Gauge) sensors typesto be calibrated: • Tipping bucket rain gauges, • Weighing type rain gauges. Calibration Center - Ankara 2010

  38. PRECIPITATION CALIBRATION LABORATORY Scope of calibration Calibration Center - Ankara 2010

  39. PRECIPITATION CALIBRATION LABORATORY Reference devices for Rain Gauges Calibration: Mettler Toledo XP4002S Precision Electronic Balances 800 gr. / 4100 gr. Baykon F1 Class Weigths 1 gr / 2 kg Calibration Center - Ankara 2010

  40. PRECIPITATION CALIBRATION LABORATORY Medium provider for Rain Gauges Calibration: Precipitation simulation is prepared by using the Watson Marlow 520Du peristaltic pump which calculates the necessary amount of water for the desired amount of Rain İntensity. Calibration Center - Ankara 2010

  41. PRECIPITATION CALIBRATION LABORATORY Secondary devices of the Precipitation Laboratory: Zeno 3200 Data Logger Caps and tubeswith different type and dimensions. Calibration Center - Ankara 2010