ESPECTROSCOPIA DE ABSORCION ATOMICA

ESPECTROSCOPIA DE ABSORCION ATOMICA. LOS PRINCIPIOS BASICOS DE ABSORCION ATOMICA PUEDEN EXPRESARSE EN TRES GRANDES CONCEPTOS :. TODOS LOS ATOMOS PUEDEN ABSORBER LUZ LA LONGITUD DE ONDA A LA CUAL LA LUZ ES ABSORBIDA ES ESPECIFICA PARA CADA ELEMENTO EN PARTICULAR

nara
Télécharger la présentation

ESPECTROSCOPIA DE ABSORCION ATOMICA

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ESPECTROSCOPIA DE ABSORCION ATOMICA LOS PRINCIPIOS BASICOS DE ABSORCION ATOMICA PUEDEN EXPRESARSE EN TRES GRANDES CONCEPTOS : • TODOS LOS ATOMOS PUEDEN ABSORBER LUZ • LA LONGITUD DE ONDA A LA CUAL LA LUZ ES • ABSORBIDA ES ESPECIFICA PARA CADA ELEMENTO • EN PARTICULAR • LA CANTIDAD DE LUZ ABSORBIDA ES PROPORCIONAL • A LA CONCENTRACION DE ATOMOS ABSORBENTES

  2. ESPECTROSCOPIA DE ABSORCION ATOMICA REQUERIMIENTOS DE UN SISTEMA PARA ABSORCION ATOMICA - FUENTE DE EMISION DE LUZ CARACTERISTICA - UN SISTEMA DE ATOMIZACION PARA CREAR UNA POBLACION DE ATOMOS - UN MONOCROMADOR PARA SEPARAR LUZ DE UNA LONGITUD DE ONDA CARACTERISTICA - UN SISTEMA OPTICO PARA DIRIGIR LA LUZ DESDE LA FUENTE A TRAVES DE LA POBLACION DE ATOMOS Y HACIA EL MONOCROMADOR - UN DETECTOR SENSIBLE A LA LUZ - SISTEMA ELECTRONICO EL CUAL MIIDE LA RESPUESTA DEL DETECTOR

  3. ESPECTROSCOPIA DE ABSORCION ATOMICA

  4. ESPECTROSCOPIA DE ABSORCION ATOMICA REQUERIMIENTOS ANALITICOS -CONVERTIR LA MUESTRA EN SOLUCION - UNA MUESTRA CON AUSENCIA DE ANALITO ( BLANCO ) - SOLUCIONES DE CALIBRACION - CONSTRUIR UNA CURVA DE CALIBRACION - ATOMIZACION DE LA MUESTRA

  5. Teoría

  6. Teoría

  7. Teoría

  8. Teoría

  9. Teoría

  10. Teoría

  11. Teoría

  12. Teoría

  13. Teoría

  14. Teoría

  15. Teoría

  16. Teoría

  17. Teoría

  18. Teoría LA RELACION ENTRE ABSORCION DE LUZ y CONCENTRACION DE ANALITO ES DEFINIDA POR UNA LEY FUNDAMENTAL DE ABSORCION DE LUZ : LEY DE LAMBERT¨S : LA FRACCION DE LUZ ABSORBIDA POR UN MEDIO TRANSPARENTE, ES INDEPENDIENTE DE LA INTENSIDAD DE LA LUZ INCIDENTE, y CADA CAPA DE ESPESOR INFINITESIMAL, ABSORBE UNA FRACCION IGUAL DE LUZ QUE PASA A TRAVES DE ESTA. LEY DE BEERS : LA ABSORCION DE LUZ ES PROPORCIONAL AL NUMERO DE ATOMOS ABSORBENTES EN LA MUESTRA.

  19. Teoría

  20. Teoría LA ABSORBANCIA ES UNA MEDIDA DE LA CANTIDAD DE LUZ ABSORBIDA POR LOS ATOMOS BAJO CONDICIONES DADAS, Y LA LEY DE LAMBERT BEER NOS PERMITE RELACIONAR LA ABSORBANCIA MEDIDA CON LA CONCENTRACION DEL ANALITO EN LA MUESTRA

  21. Teoría

  22. A • abc Ley De Beer-lambert Real

  23. SpectrAA Absorción Atómica Hardware

  24. Objetivos • Componentes de un sistema de absorción atómica • Funciones e interacciones de AAS

  25. Io It Resonante No-resonante Gas de relleno Resonante Componentes detector sensible a la luz monocromador sistema electrónico de lectura fuente de luz atomizador (llama, horno o hidruros)

  26. Lamp Element Code Contacts Lamp Code Dynode Pyrex Envelop Quartz End Window mp Cathode La Cu trAA Mica Shield ec Sp Graded Seal Base Anode Getter Electrical Contacts Alignment Pin Lámpara de Cátodo Hueco Diseño

  27. Anode Electrical Discharge Ne++ e- Photon Specific to Excited Atom e- Atom Atom Hollow Cathode Excitation Emission En Eo En Eo Relaxation Sputtering HCL Operación

  28. Io It Sobreposición de Líneas de Emisión con Líneas de Absorción Emisión de la Lámpara Baja Temperatura Baja Presión Absorción Atomica Alta Temperatura Alta Presión Ambas ocurren a exactamente la misma longitud de onda

  29. Separación espectral de una longitud de onda Líneas resonantes Líneas de gas de relleno Función del Monocromador Líneas no-resonantes Intensidad Longitud de onda

  30. Monocromador Slit de Salida Spherical Mirror Slit Adjustment Wheel Angle of the Grating Determines the Wavelength Focused on the Exit Slit Grating Spherical Mirror Slit de Entradat Changing  Knob Changes Grating Angle

  31. Función del Tubo fotomultiplicador PM Tube Energía lumínica (h Energía eléctrica

  32. Photomultiplier Tube Operation Anode Insulator Dynode’s (9-13) e- Photocathode e- e- e- e- e- e- e- e- e- e- e- e- Light Energy Quartz Window *100 Million Amplification of Signal

  33. Effect of EHT(Photomultiplier Voltage) Noise EHT 400 600 800 200

  34. Optica lámpara de Deuterio 50% 50% 50% RBC HCL 100% 50% 100% D2

  35. Anodo Apertura Ventana de cuarzo Cátodo Termoiónico Operación lámpara de Deuterio • Usada para medir absorción • no-atomica • Util desde 190 - 425nm • Gas de relleno es Deuterio (D2) • Una descarga de corriente excita el gas de D2 • Alta emisión de luz a través de la apertura de descarga

  36. Intensidad de una fuente de Deuterio vs Longitud de onda Rango normal 190-300 nm baja corriente HCL mA’s 300-425 nm No D2 425-900 nm