1. QE65000 Highlights Slits
Included in price
Gratings
H1-H14
HC1 cost extra
Filter
Only OFLV-200-1100 available in 3 weeks cost extra
No L2 lens Detector
Low noise S7031-1006
Back thinned - 90% QE
2-D structured array
TEC Cooled
Spectrometer
USB 2.0 interface
16 bit ADC
Low noise electronics
Other features
2. QE65000 Resolution
3. S7031 Detector Pixels
1024x58 (active) 24um square
1044x64 (total elements)
24.5 x 1.39mm Active area
No dark pixels
Sensitivity
2.2uV/e
Well depth
300Ke individual
~1.5Me summing
Low noise
Dark Signal ~200e/pixel/s @ 0C
Readout noise 16e
4. QE65000 Spec’s USB 2.0 Interface
Transfer scan <1ms
Integration time
7ms to 15min
A/D: 16 bit 500KHz
TEC
40C below ambient
-5 to –15C
0.1C resolution & stability
Triggering:
Only software
No External Hardware Trigger or External Synch
Strobe Signals (fully programmable)
Single Strobe
Continuous Strobe
GPIO: 10
5. QE65000 Spec’s Noise
Readout noise: 1.2 counts rms (36 e)
Dark signal:
2.5 counts rms @ int times <100ms
25 counts rms @ 5 min
Dynamic Range 25000:1
S/N ~ 1200:1 (well depth) Dark Signal
Double/halves for every ~6C
~30 counts/pixel/sec @ -15C
Integration time to half well is 1000s or ~16min
6. QE65000 Dark Characteristics
7. QE65000 Sensitivity Up to 90% QE
Sony is at best 40% QE and ~4% in UV
Sensitivity
30 e/count
0.03 counts/e
21 photons/count
8. QE65000 Sensitivity
9. QE65000 Linearity Non-Linearity Very good ~1%
Corrected Non-Linearity ~0.25%
10. Software
11. QE65000 Application Strengths Long Integration Time
For stationary signals/samples
Use QE65000 >20 sec
Use USB2000 <20 sec
Constrained applications
Use QE65000 if they can’t average more to increase S/N
If the sample is moving and they only have 1 shot
12. QE65000 Etalon Effect Etalon present 800-1000nm
Caused by passivation ontop of detector
Normalizable if Detector Temp constant