1 / 38

Electromagnetic Blood Flow Meter Dr. Erdem Topsakal, Advisor Brian McCalebb Taffa Porter Ky

Electromagnetic Blood Flow Meter Dr. Erdem Topsakal, Advisor Brian McCalebb Taffa Porter Kyle Eubanks Nashlie Sephus. Outline. Problem Statement Solution Introduction/Historical Information Technical Constraints Practical Constraints

paul2
Télécharger la présentation

Electromagnetic Blood Flow Meter Dr. Erdem Topsakal, Advisor Brian McCalebb Taffa Porter Ky

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Electromagnetic Blood Flow Meter Dr. Erdem Topsakal, Advisor Brian McCalebb Taffa Porter Kyle Eubanks Nashlie Sephus

  2. Outline • Problem Statement • Solution • Introduction/Historical Information • Technical Constraints • Practical Constraints • Design Approaches/Tradeoffs - Testing Apparatus - Calibration - Probe - Electronics • Timeline • References

  3. Problem • The electromagnetic blood flow meters were originally used by the University of Mississippi Medical Center. • Obtaining replacements is no longer possible. • Other commercially available flow meters proved inadequate.

  4. Solution • Reproduce and improve the original probe • Reproduce similar results to the original meter • Minimum Cost • Dependable • Easy to use • Obtain accurate results

  5. Electromagnetic Blood Flow Meter • What Are Flow Meters? • Electromagnetic blood flow meters measure blood flow in blood vessels • Consists of a probe connected to a flow sensor box

  6. Electromagnetic Blood Flow Meter • Why They Are Used? • Offers quantitative data during surgical operations • Provides a functional assessment of newly joined vessels, grafts and organs • Used in prosthesis in conjunction with cardiovascular surgical procedures • Most accurate results through both acute and chronic implants

  7. Electromagnetic Blood Flow Meter How Do They Work? • Faraday's Law of Magnetic Induction • Liquid acts as a conductor • Voltage is induced directly related to the average flow velocity • “The faster the flow rate, the higher the voltage” • Voltage is measured by sensing electrodes mounted in the meter tube • Voltage is then sent to the flow sensor box

  8. Outline • Problem Statement • Solution • Introduction/Historical Information • Technical Constraints • Practical Constraints • Design Approaches/Tradeoffs - Testing Apparatus - Calibration - Probe - Electronics • Timeline • References

  9. Technical Constraints

  10. Measurement Accuracy • Affected by: - Stray magnetic fields detected by electrodes - Non-uniform magnetic field - Turbulent fluid flow - Non-homogenous fluid • Accurate to within ±1 cm/s from 0.1-1 m/s - 10% maximum error at 0.1 m/s - 1% maximum error at 1 m/s

  11. Maximum Fluid Velocity • Fluid velocities in an aorta - 89±9.5 cm/s during heart contraction - 36±6.0 cm/s between contractions • Fluid velocity for flow meter - 1 m/s maximum • Importance of low maximum fluid velocity - Maximizes accuracy for low fluid velocities by allowing more precision in A/D conversion

  12. Conductivity • Calibrated conductivity range • Conductivity of blood - 70 Siemens per centimeter - Varies by up to 20% based on flow rate • Acceptable conductivity range of flow meter - 60–80 Siemens per centimeter

  13. Size • Probe Size • 22 mm inner diameter • Reasoning - Requested by sponsor - Diameter of aorta ranges from 21-35 mm - Larger aortas taper down to smaller diameters

  14. Practical Constraints

  15. Sustainability • Implanted for 2-3 months • Protection of wire leads, magnetic core, & wire coil • Maintenance of electrodes

  16. Ethical • Designed for cow’s aorta only • Not approved or tested for human use

  17. Outline • Problem Statement • Solution • Introduction/Historical Information • Technical Constraints • Practical Constraints • Design Approaches/Tradeoffs - Testing Apparatus - Calibration - Probe - Electronics • Timeline • References

  18. Testing Apparatus

  19. Testing Apparatus

  20. Testing Apparatus

  21. Testing Apparatus • Via Aqua 1800 - $25.00 - 480 GPH, variable flow rate - 3/4” connections - Saltwater safe • Acrylic tubing - $3.32 / 6ft. - 7/8” OD, 3/4” ID, 1/16” thickness - Insulating material

  22. Testing Apparatus • Dialysis tubing - $5.25 / 10ft - Will be used in future testing - 22 mm diameter - 1 mil thickness - Closely replicates the conductivity of an aorta - May use multiple layers to adjust the conductivity or increase the water pressure it can withstand

  23. Calibration • Calibration and error can be minimized by selecting the proper voltage source waveform • DC - Voltage drift from polarization of electrodes • Sinusoidal AC - Reduces voltage drift by changing polarity - Induces emf in electrodes from magnetic flux variation • Square-wave AC - Reduces voltage drift by changing polarity - Reduces emf in electrodes from magnetic flux variation

  24. Calibration • Why does the output voltage of the probe not directly correlate to the fluid velocity? • Ideally, they are directly proportional • However, sources of error include: - Stray magnetic fields picked up by electrodes - Resistive and capacitive current leakage - Voltage loss through tubing - Voltage drift - Noise

  25. Calibration • Fundamentals of Calibration • Under no flow conditions - Voltage observed is the combination of all sources of error since no voltage from fluid flow • Zero the output waveform - Cancels out all unwanted voltage contributions - Output now has a zero baseline voltage • Output voltage is now directly proportional to the fluid velocity

  26. Calibration • Procedure - The most correct solution is to subtract the entire unwanted waveform - Requires unnecessary computations - Sampling occurs at the same place every cycle - Under no flow, calculate the average voltage at the desired sampling location for every cycle over some period - Subtracting this voltage for every sample gives the voltage contribution due to fluid velocity alone

  27. Outline • Problem Statement • Solution • Introduction/Historical Information • Technical Constraints • Practical Constraints • Design Approaches/Tradeoffs - Testing Apparatus - Calibration - Probe - Electronics • Timeline • References

  28. Probe Key components • Magnetic core • Magnet wire • Electrodes

  29. Probe • Magnetic cores • Permeability - Describes how much a material is affected by magnetic fields and how strong a field it can generate • High permeability results in: - A stronger magnetic field

  30. Probe • Magnetic cores

  31. Probe • Magnet wire Key factors - Thin - Well-insulated - Close to range of 34 gauge (commonly used in industrial electromagnetic blood flow meters) - Availability - Low cost

  32. Probe • Electrodes Desirable qualities • Low resistance - More sensitive • High capacitance - Reduces source impedance and signal attenuation

  33. Electrodes - Capacitance Platinized Platinum Capacitance (µF) Frequency (Hz)

  34. Electrodes - Resistance Resistance (kΩ) Platinized Platinum Frequency (Hz)

  35. Electrodes material Electrodes

  36. Electronics Flow sensor box

  37. Design Approach Timeline

  38. References [1] Shercliff, J.A., The Theory of Electromagentic Flow-Measurement. Cambridge: University Press, 1962,pp. 2-4, 125. [2] Flowmeter Directory. Flowmeter Directory. 2007. http://www.flowmeterdirectory.com/flowmeter_electromagnetic.html [3] Tavoularis, S., Measurement in Fluid Mechanics. Cambridge: University Press, 2005, pp. 217. [4] Omega Engineering. Omega Engineering. 2006. http://www.omega.com/techref/flowcontrol.html [5] EesiFlo. Eesiflo. 2007. http://www.eesiflo.com/applications.html [6] Braun, U., and vetJosef, F., “Duplex ultrasonography of the common carotid artery and external jugular vein of cows,” American Journal of Veterinary Research, vol. 66, no. 6, pp.962-965, 2005. [7] Khan, S. R., and Islam, M. N., “Studies on the prospect of bioprostheses by cow aortic valve for human use,” Bangladesh Med Res Counc Bull, vol. 17, no. 2, pp. 75-80, 1991. [8] Wyatt, D. G., and Phil, D., "Problems in the Measurement of Blood Flow by Magnetic Induction," Phys. Med. Biol., vol. 5, pp. 369-399, 1961.

More Related