Stripixel Sensor Pre-production at Hamamatsu. Junji Tojo RIKEN VTX Meeting, Dec. 9 th , 2004. Pre-Production at HPK. Design 3 sensors/wafer on 6” wafer 625um and 500um thickness. Two designs : “STD” & “NEW” design. Process breakdown “STD” design w/ 625 um thickness : 6 wafers

trapper
Télécharger la présentation

Stripixel Sensor Pre-production at Hamamatsu

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Stripixel Sensor Pre-productionat Hamamatsu Junji Tojo RIKEN VTX Meeting, Dec. 9th, 2004

  2. Pre-Production at HPK • Design • 3 sensors/wafer on 6” wafer • 625um and 500um thickness. • Two designs : “STD” & “NEW” design. • Process breakdown • “STD” design w/ 625 um thickness : 6 wafers • “STD” design w/ 500 um thickness : 6 wafers • “NEW” design w/ 625 um thickness : 12 wafers • “NEW” design w/ 500 um thickness : 11 wafers • HPK process was finished. 35 wafers were processed in total. • Originally, 20 good wafers in contract. • HPK processed 35 wafers (105 sensors), since the wafer yield was unknown. • We asked HPK to deliver all 35 wafers to us (under negotiation). • Wafers are under testing at HPK.

  3. MPD sensor MPD 6 inch wafer Photograph of Stripixel Sensor Wafer(“NEW” Type Design) MPD : Monitor Photo-Diode

  4. Testing at HPK • Testing item • CV measurements with Monitor Photo-Diodes (MPDs) • Full depletion voltage (VFD) measurement. • There are 6 MPDs per wafer. • IV/CV measurements • Current/Capacitance characteristics as a function of Vbias. • Only a few channels were measured as a reference. • I/C measurements at V≈VFD+100 V • Measurement for all the channels • Evaluation of the strip yields (important for mass-production)

  5. CV Measurement with MPDs • 6 MPDs per wafer : UL, UM, UR, DL, DM, DR

  6. Wafer number CV Measurement with MPDs • Typical CV curve for “NEW” Design MPDs • Figure shows the results for “DL”-MPDs.

  7. CV Measurement with MPDs • VFD estimated for “NEW” design wafers • ~155-175 V for 500 um wafers, ~100-120 V for 625 um

  8. Wafer number CV Measurement with MPDs • Typical CV curve for “STD” Design MPDs • Figure shows the results for “DL”-MPDs.

  9. CV Measurement with MPDs • VFD estimated for “STD” wafers • ~145-170 V for 500 um wafers, ~95-115 V for 625 um VFD results are close to an estimated value from wafer resistivity .

  10. I / C GND GND IV/CV Measurement • Measured only for “NEW” design wafers. • Only one channel for all the sensors. • The 6th pad from the left in X1 pads • Guard ring and a few adjacent pads were grounded. • IV/CV in 5-500 V.

  11. C (pF) I (nA) IV/CV Measurement • Results for wafer #1 : “NEW”, 625 um Sensor Left, Middle, Right • Preliminary result • C ~ 26, 21, 23 pF at V=VFD • I ~ 123, 49, 90 nA at V=VFD (VFD ~ 110 V ) • More results will come.

  12. I/C Measurement Wafer #1 “NEW”, 625 um • Leakage current measurements at Vbias = 230V for all the 625um wafers. • Measurements of C for 625 um wafers and I/C for 500 um wafers are going on. • More results will come.

  13. Expected Schedule • HPK sensor • “NEW” design sensor has the highest priority. • Delivery expected to be early next year. • Dicing & more testing at BNL in January. • StripFee prototype • Completion is expected to be ready in February (See Vince’s slides). • Detector tests • Detector integration (gluing & wire-bonding) at RIKEN in February. • Laser tests at ORNL (or RIKEN) in March.

  14. Summary • CV measurement with MPDs • VFD =145-175 V for 500um wafers, 95-120 V for 625 um wafers • Consistent with wafer resistivity • IV/CV measurements • Measured for only for “NEW” wafers (One channel per sensor) • Preliminary result was shown. More will come. • I/C measurements • Measured for all the channels • Measurements of current for 625 um wafers were done. • C for 625 um wafers, I/C for 500 um wafers are going on. • Wafers are expected to be delivered around the early next year.