SPACE TELESCOPE SCIENCE INSTITUTE. Correcting STIS Point-Source CCD Spectra for CTE loss. Operated for NASA by AURA. Paul Goudfrooij * Space Telescope Science Institute. Methods to measure CTE of STIS CCD (Visual) effects of CTE loss

turner
Télécharger la présentation

Correcting STIS Point-Source CCD Spectra for CTE loss

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. SPACE TELESCOPE SCIENCE INSTITUTE Correcting STIS Point-Source CCD Spectra for CTE loss Operated for NASA by AURA Paul Goudfrooij* Space Telescope Science Institute • Methods to measure CTE of STIS CCD • (Visual) effects of CTE loss • CTE correction formula for point-source spectroscopy • Plans for the near future * With contributions of Ralph Bohlin, Jesús Maíz-Apellániz, & Randy Kimble

  2. A Brief History of STIS CTE Measurements CTE: “Fraction of charge transferred per pixel transfer”(< 1 due to traps in the silicon). Typically, one measures CTI  1 – CTE. • Significant parallel CTI discovered on STIS in 1998 (Gilliland, Goudfrooij & Kimble,1999, PASP, 111, 1009). • Apparent non-linearity, most significant for faint sources. • Review of effects of radiation damage on STIS CCD: Kimble, Goudfrooij & Gilliland, 2000, Proc. SPIE, vol. 4013, p. 532. • Empirical determination of functional form for imaging CTE: Goudfrooij & Kimble, 2002, in Calibration Workshop Proceedings • First empirical determination of functional form for spectroscopic CTE: Bohlin & Goudfrooij, STIS ISR 2003-03 • Alternative determination using modelinvolving published parameters of ‘known’ traps (Bristow, STIS-CE ISRs, 2002-2004)

  3. Amp D Nominal Readout Direction Axis2 (Y) Nominal Clocking Direction 1 2 (fluxD / fluxB) Y CTI ~ Parallel (virtual) overscan Amp A Amp B Amp C Axis1 (X) Serial overscan Serial overscan STIS CCD Architecture; Measurement Method STIS CCD: • 4 Readout Amps (1 / corner) • Nominal Amp: D (lowest RN) • Bi-directional Clocking yields CTI  1 – CTE: Measured using “Sparse Field Tests”

  4. “Internal” Sparse Field Test “Sparse Field” Test • Sparse fields to ensure that sources do not overlap, in which case (e.g.) PSF wings would fill traps for sources along the readout direction • Typical situation for point source spectroscopy • Lamp images along narrow, cross-dispersion slits, projected at 5 positions along columns (or rows) • Representative of “worst-case” point source spectroscopy (essentially no background to fill traps) • Test conducted before launch and throughout STIS life

  5. Visual effects of CTE loss: Dependence on Y position “External” Sparse Field Test: Outer Field in NGC 6752 (Oct 2001) Exp. time = 20 s

  6. Visual effects of CTE loss: Effects of Intensity and Sky “External” Sparse Field Test: Outer Field in NGC 6752 (Oct 2001) Exp. time = 20 s Exp. time = 100 s

  7. “Internal” Sparse Field Test Visual effects of CTE loss: Centroid shifts • Charge trailing and centroid shift measurable; most significant at low signal. • Impacts ‘shape’ measurements (surface photometry), especially for faint objects.

  8. Spectroscopy Imaging CCD Column Number CCD Row Number Functional Dependence on Signal and Background Levels • To be done separately for imaging and spectroscopy: • For given “extracted” signal level, spectroscopy has higher “per column” signal and hence: • Lower CTE loss • Different dependence on background level

  9. Internal Sparse Field Test: CTIAnalysis • Clear signatures of CTE effect: • Fractional loss decreasing with increasing signal • Absolute loss increasing with increasing signal • Measurements identical every epoch (and one measurement before launch) • Hence used to derive time constant of CTI degradation B amp / D amp ratio  Rows from B amp 

  10. Insights from Monitoring of Spectrophotometric Standards • Wide-slit spectra of flux standard AGK+81D266 taken every few months • Division of (early, middle, late) data by overall average reveals CTE effect (even for such high-S/N spectra) • CTI dependence on background B and gross signal G modeled (in spring 2003) as CTI  G–a exp (–b(B/G)c)

  11. 0.21 Insights from Monitoring of Spectrophotometric Standards Other important constraints on functional form of CTI: – Observed flux ratios between spectra of flux standards: • MAMA/G230L vs. • CCD/G230LB or CCD/G430L – Functional form producing best fit to the data as of Spring 2003 (cf. ISR 2003-03 and implemented in CALSTIS pipeline): Example: LDS 749B, 90 – 1800 e– B G CTI = 0.0355 G–0.75 (1 + 0.243 (t –2000.6))  exp (–2.97 )

  12. Final Close-out Effort • 2003 solution overestimated CTI somewhat for G750L spectra redward of about 8500 Å • Likely due to presence of extended PSF (“red halo”), which acts as effective “extra background”, filling traps • Working on implementation as follows: CTI  G–a exp (–b (cB+H)/G)d) [where H is the ‘halo’ signal in the PSF beyond (above) the standard 7-pixel extraction box for point source spectra] • H values to be calculated by pipeline, using available encircled energy tables as function of wavelength (the PCT reference files)

  13. Improvement to CTI corrections (P. Goudfrooij & R. Bohlin) • Inclusion of ‘halo’ component yields significant improvement beyond 8000 Å

  14. Future Plans on Empirical CTE corrections • Finalize correction formula for spectroscopic data • Write ISR (Nov 2006) • Implement in CALSTIS pipeline (Spring 2006) • Write IRAF (post-observation) script to correct imaging photometry of compact sources for CTI (Spring 2006) • Analysis of extended source CTI measurements (Spectroscopy & Imaging; Spring 2006)

  15. bckgr counts External Sparse Field Test: CTIAnalysis Clear dependence on background level (“sky”) • Slope systematically flatter with increasing flux • “Sky” presumably fills traps in bottoms of potential wells, mostly affecting transfer of small charge packets. • Suggests CTI  exp –