1 / 23

Safety Meets Sustainability

Safety Meets Sustainability. Rob Klein, MS, CIH Environmental Health & Safety Yale University New Haven, CT Making Universities Sustainable Copenhagen, Denmark 23 October 2014. Partnerships at Yale. Sustainability Facilities / Utilities / Systems Engineering

zeus-davis
Télécharger la présentation

Safety Meets Sustainability

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Safety Meets Sustainability Rob Klein, MS, CIH Environmental Health & Safety Yale University New Haven, CT Making Universities Sustainable Copenhagen, Denmark 23 October 2014

  2. Partnerships at Yale • Sustainability • Facilities / Utilities / Systems Engineering • Purchasing and Procurement • Yale Animal Research Center • Lab Managers & Researchers • Environmental Health & Safety

  3. Modern Laboratories • Protect research materials and people • Highly engineering for odor control, comfort, and safety • High flux of supplies, materials, and hazardous wastes • Trend towards disposal products • Primarily 100% exhaust air • Energy-intensive • All labs are not =

  4. 1 Includes surgical path lab services, autopsy, and gross anatomy

  5. Laboratory Footprint at Yale !!!

  6. Energy Consumption in Yale Labs (US$ 32 M/year) Chilled water 41% Steam 31% Electricity 10% Vs. US$ 2M/year regulated waste disposal

  7. Historical Tension Areas • General lab room ventilation • Fume hood flow rates • Recycling chemical containers • Excess protective equipment use in animal research • Freezer and contents management

  8. Obstacles to Change • “We’ve always done it this way” • Requirements from regulatory agencies and accrediting organizations • Over-designed systems • Changes in research – reagents, equipment • Indirect costs externalize involvement • Funding rules that limit “sharing” and encourage over-buying (“use it or lose it”)

  9. General Laboratory Ventilation • Background: • Control emissions, reduce odors, and clear spills • Tension: • 100% exhaust of conditioned lab air, historically at rates of 4-12 or more ACH; even higher in animal facilities • Safety concerns: • Older epidemiological data suggest higher incidence of disease among chemists and other lab workers • Air flows too low will result in exposures

  10. … in a Perfectly Mixed Room C = C oe-n

  11. Benchtop Work (x 6) “fugitive emissions” Small Spills

  12. Original Supply Air Diffusers New Radial Style Supply Air Diffusers

  13. Spills – Duct MeasurementsOriginal 6 Supply Air Diffusers

  14. Spills – Duct MeasurementsOriginal But Only 3 Supply Air Diffusers

  15. Spills – Duct Measurements3 New Style Supply Air Diffusers

  16. Supply Air Diffuser EffectsSpills (Duct Measurements) • Observations: • Higher [peak]’s • Faster clearance

  17. CFD Modeling - Residual Contamination(benchtop emission model, time = 70 min, 6 ACH) 6 original supply air diffusers 3 new supply air diffusers

  18. Empirical Conclusions • Higher ACH → lower concentrations, faster clearance • Greatest improvement from ~ 4 - 6 ACH • But diminishing returns ≥ 8 ACH • Reducing number and style of SA diffusers provides functional equivalent of 2 – 4 additional ACH vs. original room HVAC design • Must consider original HVAC system design before lowering ventilation rates • Be aware of the limitations of demand control systems

  19. Very Serious Cost Savings • @ US$ 8 / CFM exhaust / year (average): • 1 ACH reduction ~ $1,400 / yr test lab alone • $116,000 / yr this bldg alone • $1.95 M / yr entire campus • With proper room HVAC design, we now accept 6 ACH, less in unoccupied labs Shameless self-promotion: 2 articles in J Chemical Health & Safety – Klein, King, and Kosior

  20. Some Other Successes

  21. Conclusions and Advice • Data-driven • Dialogue, communication, and teamwork • Open mind and willingness to consider change

  22. Questions?