Radiological Equipment - Advanced

Radiological Equipment - Advanced. Objectives. For each radiation detection instrument: Perform advanced operating procedures Become familiar with software Identify advanced maintenance procedures Prerequisite training: Radiation awareness Basic operation and maintenance of each instrument.

Pat_Xavi
Télécharger la présentation

Radiological Equipment - Advanced

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Radiological Equipment - Advanced

  2. Objectives • For each radiation detection instrument: • Perform advanced operating procedures • Become familiar with software • Identify advanced maintenance procedures • Prerequisite training: • Radiation awareness • Basic operation and maintenance of each instrument

  3. IsoTrak DoseGuard

  4. Basic Operation Review • Prepare for operation • Check dose and dose rate • Check alarms

  5. Advanced Operation Overview • Recommended procedures not to be done by operator • Reset default reading • Turn off/on chirp • Reset dose reading to zero • Check/change dose or rate alarm set points • Deactivate dose or rate alarm • Test battery during operation • Use reader and software

  6. Function Settings • Toggle button to feature, hold to reset • Feature options • 1x = non-default display (dose rate) • 2x = OFF • 3x = chirp off/on • 4x = reset dose to zero • 5x = dose alarm display/reset • 6x = rate alarm display/reset • 7x = battery test

  7. Reset Default Reading • Normal default is total dose reading • Can switch with current dose rate reading • Procedure • Press button 1x • Display shows dose rate (non-default) • Hold button until long beep; release • Display shows dose rate as new default

  8. Chirp Feature • “Chirp” alerts user of increased dose rate • Disabled chirp does not mute alarms • Display shows horn if chirp is enabled • Procedure to disable/enable • Press button 3x • Display reads “CHr” • Hold button until long beep, release

  9. Reset Dose to Zero • Verify dose is zeroed before use • Procedure • Press button 4x • Display reads “CLr” • Hold button until long beep • Display shows current dose flashing • Will revert to default if no further action • While flashing, hold button until long beep • Dose resets to zero, returns to default

  10. Check Dose Alarm Set Point • Verify before use • Procedure • Press button 5x • Display reads “dAL” • Hold button until long beep, release • Display shows dose alarm setting, returns to default

  11. Reset Dose Alarm Set Point • Reset to current guidance • Procedure • Check alarm: Press button 5x, display “dAL”, hold • Display shows dose alarm setting for several sec. • During display, toggle button through list of choices • Select zero to disable dose alarm • Hold button until long beep; release

  12. Check Dose Rate Alarm Set Point • Verify before use • Procedure • Press button 6x • Display reads “drA” • Hold button until long beep, release • Display shows rate alarm setting, returns to default

  13. Reset Dose Rate Alarm Set Point • Reset to current guidance • Procedure • Check alarm: Press button 6x, display “drA”, hold • Display shows rate alarm setting for several sec. • During display, toggle button through list of choices • Select zero to disable rate alarm • Hold button until long beep; release

  14. Test Battery During Operation • Use for troubleshooting battery • Procedure: • Press button 7x • Display reads “diA” • Hold button until long beep; release • Alarm sounds for 5 sec. • If battery symbol appears after test, replace battery

  15. Using the Reader • Install software to PC • Use PC to • Prevent user from changing settings via button • Customize settings • Print data

  16. Start Window

  17. Main Window

  18. Quick Configuration Window

  19. Print Configuration Data

  20. Introduction to Maintenance • Operator level • Perform simple, “common sense” checks without tools • Hospital (organization) level • Verify failures • Perform advanced programming or diagnostics • Replace authorized components • Factory or supporting maintenance • Calibrate, repair, rebuild, replace internal components

  21. Advanced Maintenance Overview • System diagnostics • No corrective maintenance • Calibration • Warranty information

  22. System Diagnostics • No user adjustments possible • Note error code and return to factory • Error codes: • Er1 = calibration error • Er2 = fault in nonvolatile RAM • Er3 = fault in detector • Er4 = fault in RAM

  23. Calibration • Frequency: Recommended annually • Based on local protocols • Procedure for sending to factory • Track due date—no surprises • Send instrument to factory • Track shipment until returned • Cost estimate (2006): $40

  24. Warranty Information • 12 months free of defects in material or workmanship

  25. Ultra Radiac

  26. Basic Operation Review • Prepare for operation • Check dose and dose rate • Turn on back light • Check stay time • Data logging • Alarms

  27. Advanced Operation Overview • Recommended procedures not to be done by operator • Reset dose to zero • Use sleep mode • Use source finder mode • Check/change alarm indicators • Check/change dose or rate alarm set points • View data log • Clear data log • Use cable and software

  28. Reset Dose to Zero • Verify dose is zeroed before use • Log data before zeroing • Procedure • Press Dose key, display shows dose reading • If dose not 000, continue • Hold Dose + Clear/Test keys until “000” is displayed

  29. Use Sleep Mode • Extends battery life >1000 hours • Wakes up every 5 minutes • Alarm sounds if preset levels exceeded • Set sleep mode • When installing batteries • Hold Rate + On/Off key • Resume normal operation • Hold On/Off key

  30. Use Source Finder Mode • Purpose: Use as an alternate survey meter • Activate: • Hold Rate key until flashing “1” • Use as survey meter • Audible clicker activated • Deactivate: Hold Rate key until flashing “0”

  31. Check Alarm Set Points • Dose rate alarm set points: High and Low • Press Alarm + Rate keys • Press Rate key to select H or L • Press Clear/Test key to see alarm set point (H or L) • Vibrator alarm active: display also shows “- - -” • Dose alarm set points: High and Low • Same procedure using Alarm + Dose keys

  32. Change Alarm Indicators • Enable/disable visual, audio, or vibration • Procedure (rate alarm): • Check alarm: Rate + Alarm, Rate (H/L), Clear/Test • Press Alarm until desired combination appears • Wait 10 sec. for return to normal display • Dose alarm procedure same, use Dose key

  33. Change Alarm Set Points • Procedure (rate alarm): • Check alarm: Rate + Alarm, Rate (H/L), Clear/Test • Press Clear/Test again to start • Press Rate until desired decimal and units appear, press Clear/Test to select • Repeat for each digit • Same procedure for dose alarm • Use dose key

  34. View Logged Data • View data from instrument • Power Off: Hold Rate + On/Off keys 1 sec. • Displays current log number • Scroll up (Rate) or down (Dose) • Number, rate, dose, next number … • Return to normal: Power Off and On • PC access (advanced operations) • Download, view, or clear data log

  35. Clear Data Log • Recommend download to PC first • Procedure • View data log (power Off, hold Rate + On/Off) • Hold Rate + Alarm, press Clear/Test (all 3) • Display “0” indicates all data cleared • Return to normal: Power Off and On

  36. Using the PC Software • Use the PC to • Print and download logged data • Change instrument parameters • Observe readings in real time

  37. Initializing Setup • Install software to PC • Connect cable between PC and meter • Start software • Set password 12345 • Main screen appears

  38. Main Window

  39. Text File Export • File menu on Main Screen • Export to text or Excel

  40. Interrogate Meter

  41. Advanced Settings

  42. Manage Templates

  43. Monitoring Mode

  44. Advanced Maintenance Overview • Troubleshooting • System diagnostics • No corrective maintenance • Calibration • Warranty information

  45. Troubleshooting

  46. System Diagnostics • Test key function during operational test • Initiate test: Clear/Test, 888  000, press keys • Key codes: • Dose = 002 • Alarm = 010 • Rate = 001 • Light = 020 • Clear/Test = 004 • On/Off = OFF

  47. Calibration • Frequency: Recommended 1-2 yrs • Based on local protocols • Procedure for sending to factory • Track due date—no surprises • Schedule to keep some instruments on hand • Call for tracking # • Send instrument to factory • 2-3 week turnaround • Cost estimate (2006): $125 • FDNY uses local calibration option

  48. Warranty Information • Published in User’s Manual • 1 year free of defects from materials or workmanship • 90 days software operates as specified

  49. Radiation Survey Meter

  50. Controls and Indicators Calibration switch well open to see adjustments