Imid žing magnetnom rezonancijom

Imid žing magnetnom rezonancijom. Malo istorije …. Otkriće fenomena NMR –Isaac Rabi 1938. u molekularnim snopovima (Nobelova nagrada 1944). NMR u tečnostima i čvrstim supstancijama – Bloch & Purcell 1945 (Nobelova nagrada 1952.)

abdalla
Télécharger la présentation

Imid žing magnetnom rezonancijom

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Imidžing magnetnom rezonancijom

  2. Maloistorije… Otkriće fenomena NMR –Isaac Rabi 1938. u molekularnim snopovima (Nobelova nagrada 1944). NMR u tečnostima i čvrstim supstancijama – Bloch & Purcell 1945 (Nobelova nagrada 1952.) Richard Damadian 1971.Tkivo tumora ima drugačije NMR parametre od zdravog tkiva. Paul Lauterbur 1973. otkriće MRI

  3. Magnetni moment i magnetizacija s m Magnetni moment jezgra je sumamagnetnihmomenataprotonaineutrona, vodeći računa o pravilu sparivanja Magnetni moment nukleona m=(e/2m)s =gs Ukupni nuklearni magnetni moment u uzorku u odsustvu magnetnog polja jedak je nuli

  4. Magnetni moment jezgra S(H) = ½ S( 2H) = 1 S( 4He) = 0 • Jezgra koja imaju paran broj protona i neutrona imaju • ukupan spin 0 • ...sa neparnim brojem protona i neparnim brojem neutrona • celobrojni spin • ...ostala polubrojni spin Sabiraju se samospinoviistihčestica

  5. NMR osobine nekih jezgara

  6. Spinovi u magnetnompolju Е2 Mz N2 Е2 Е1 Е1 N1

  7. PrecesijaiLarmorovajednačina w = (e/2m)B= gB w= 2pn= 2pgB

  8. Uvođenje ekscitacije i magnetna rezonancija B1 =B01 sin(wt) Elektromagnetni talas frekvencije jednake Larmorovoj B01 je amplituda vektora magnetnog polja • je ugao za koji polje B1obara magnetizaciju ka xy ravni

  9. RF puls gate q=90o RF polje se primenjuje u pulsnom režimu

  10. Ponašanje magnetizacije posle pulsa od 900 Mx vreme

  11. Detekcija raspada slobodne indukcije (FID) Mx My FID je NMR signal

  12. T2 vreme relaksacije • vreme = 0 Mxy je proporcionalno broju • spinova (spinska gustina, protonska gustina) • Vremenska konstanta eksponencijalnog • opadanja magnetizacije se naziva spin-spin • vreme relaksacije T2 • U stvarnosti magnetizacija opada brže zbog • nehomogenosti magnetnog polja u sistemu • Vremensko opadanje magnetizacije je u tom slučaju karakteriše konstanta T2*

  13. Spin-spin relaksacija

  14. Spin-rešetka vreme relaksacije (T1)

  15. T1 i T2 vrednosti za različita tkiva

  16. FID i spin eho FID EHO

  17. + 90o 180o +

  18. Od čega zavisi NMR signal Np broj spinova (protona), spinska gustina TR, TE parametri akvizicije signala T1 –spin-rešetka vreme relaksacije T2 –spin-spin vreme relaksacije

  19. Dobijanje MR slike - gradijent magnetnog polja B B B + FT FT

  20. Gradijentmagnetnogpolja (2) B NMR zeugmatografija CT

  21. Gradijent magnetnog polja (2) B B w1 + 0 w1 Gs B=Bo+ xGx w= g(Bo+ xGx) RF

  22. Gradijent izbora preseka – debljina preseka • Veći gradijent, tanji presek

  23. Gq Gs Gq Gn Gn Gs Gn Gq Gs Odgovarajućom orijentacijom gradijenata je direktno dobiti bilo koji aksijalni , koronalni ili sagitalni presek

  24. f n Gradijent faznog kodiranja & Gradijent frekvencijskog kodiranja omogućavaju lokalizaciju signala u ravni preseka

  25. Dobijanje MR slike –gradijentikodiranjafrekvencijei faze Gphs n2 Gfrq n1

  26. 4th av. 3rd av. 2nd av. 1st av. -1st av. -2nd av. -3rd av. -4th av. -N3 -N2 -N1 N0 N1 N2 N3

  27. K-prostor ky Gp Gf kx

  28. Slika K-prostora Fazno kodiranje Frekvenciono kodiranje

  29. FT1 FT1

  30. Digitalizacija ADC FT

  31. Digitalizacija- nekorektno uzorkovanje

  32. Širina oblasti frekventnog kodiranja izražena u jedinicama frekvencije naziva se frekventni opseg akvizicije (bandwidth) Broj stupnjeva faznog kodiranja x broj segmenata u frekventnom kodiranju/čitanju signala = MATRIX Broj stupnjeva faznog kodiranja direktno utiče na vreme za koje će se dobiti MRI slika Veličina gradijenata PC i FC određuje koji deo objekta će biti sniman – t.j. određuje field of view

  33. Pulsne sekvencije i mehanizmi relaksacije

  34. Pulsne sekvencije, osnovni pojmovi Podsećajunanotnizapis Određujuredosled, polaritet, međusobniodnositrajanje RF pulsevaigradijenatakoji se koristepriformiranju MRI slike. Obavezan je jedanradiofrekventnipulsod 90oi tri gradijenta. U zavisnostiodtipasekvencijemožebitiprisutnojedaniliviše RF pulseva.

  35. Osnovna MRI sekvencija ky kx

  36. Spin eho sekvencija – zlatni standard 180o 90o 90o TE/2 TE/2 TR

  37. + 90o 180o +

  38. Spin eho u K-prostoru 180o kx

  39. Kontrasti u spin ehu

  40. Kontrast u spin ehu (2)

  41. T1W i T2W snimanje

  42. T1i T2 – mast ivoda

  43. Tipične vrednosti parametara kod SE

  44. Ekonomičnost –smultana akvizicija u više preseka Kod klasičnog spin eha 1 akvizicija = 1 linija K-prostora Za jednu akviziciju kod spin eha se utroši manje od 1/20 TR. “Slobodno” vreme se može iskoristiti za akviziciju u drugim presecima TR1 TR2 TR3

  45. Double spin echo Prvi kontrast – PD, drugi T2w

  46. Još više ekonomičnosti: turbo spin eho

  47. Turbo spin eho u K-prostoru TR1 TR2 TR3 vreme akvizicije je obrnuto proporcionalno broju ehoa

  48. Kontrast kod turbo spin eha

  49. TSE primer

  50. Inversion recovery