Treatment of In-Stent Restenosis by Paclitaxel Coated PTCA Balloons

PACCOCATH ISR. Treatment of In-Stent Restenosis by Paclitaxel Coated PTCA Balloons. Presented at The American Heart Association Scientific Session 2006 Presented by Dr. Bruno Scheller. PACCOCATH ISR: Background.

afi
Télécharger la présentation

Treatment of In-Stent Restenosis by Paclitaxel Coated PTCA Balloons

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. PACCOCATH ISR Treatment of In-Stent Restenosis by Paclitaxel Coated PTCA Balloons Presented at The American Heart Association Scientific Session 2006 Presented by Dr. Bruno Scheller

  2. PACCOCATH ISR: Background • Treatment of coronary in-stent restenosis is hampered by high incidence of recurrent in-stent restenosis. • Multiple studies have shown the benefit of drug-eluting stents in preventing restenosis in de novo lesions, but only one randomized trial, ISAR-DESIRE, has been conducted for patients with pre-existing lesions. • While the ISAR-DESIRE trial showed a reduction in restenosis for drug-eluting stents compared with bare metal stents, the stent-in-stent approach for treating restenosis may not be optimal or feasible for all patients. • The goal of this trial was to assess the efficacy and safety of a paclitaxel eluting balloon catheter compared with a non-eluting balloon catheter among patients with coronary in-stent restenosis. Presented at AHA 2006

  3. PACCOCATH ISR : Study Design 52 Patients with single in-stent restenosis in a coronary artery with a diameter stenosis of > 70%, < 25mm length, and vessel diameter of 2.5 mm to 3.5 mm; stable or unstable angina or a positive functional study. Randomized. Double-Blind. Parallel. 29% female, mean age 64 years, mean follow-up: 12 months Paclitaxel-Eluting Balloon Catheter (3 µg paclitaxel/mm2) n=26 Non-Eluting Balloon Catheter (same balloon design without drug) n=26 • Primary Endpoint: Angiographic late lumen loss at 6 months • Secondary Endpoint: Rates of restenosis; major adverse cardiac events (MACE) Presented at AHA 2006

  4. PACCOCATH ISR : Primary Endpoint In-segment late Lumen Loss (mm) • In-segment late lumen loss was smaller in the paclitaxel group than the bare balloon group (0.03+/- 0.48 vs. 0.74 +/- 0.86, p=0.002). • Minimum lumen diameter at 6-months was larger in the paclitaxel group than the bare balloon group (2.22 mm vs 1.57 mm, p=0.005). P=0.002 0.740 ± 0.86 mm 0.03 ± 0.48 Presented at AHA 2006

  5. PACCOCATH ISR: Secondary Endpoint Binary Restenosis (%) p=0.002 • Binary restenosis occurred less frequently in the paclitaxel-eluting group than the bare balloon group (5% vs. 43%, p=0.002). Presented at AHA 2006

  6. PACCOCATH ISR: Secondary Endpoint Rate of Major Adverse Cardiac Events (%) p=0.01 • At 12 months, the rate of major adverse cardiac events was 31% in the bare balloon group and 4% in the coated-balloon group (p=0.01). • This difference was primarily due to the need for target lesion revascularization in six patients in the bare balloon group (p=0.02). Presented at AHA 2006

  7. PACCOCATH ISR: Limitations • The scale of the trial does not justify clinical application or regulatory approval of the drug-coated balloons. • Although the core laboratory was not aware of study-group status, the coated balloons have a faintly white color, a difference that might have unblinded the clinical investigators. • Larger clinical trials should be conducted in order to determine whether the results of this trial can be replicated. • This trial does not evaluate whether paclitaxel eluting balloons would be efficacious in the treatment of de novo lesions as opposed to in-stent restenosis. Presented at AHA 2006

  8. PACCOCATH ISR: Summary • In this pilot study, treatment with a paclitaxel-eluting balloon catheter was associated with lower late lumen loss at 6 month angiography compared with a bare balloon catheter among patients with coronary in-stent restenosis. • The study also showed that treatment of coronary in-stent restenosis with paclitaxel-coated balloon catheters significantly lowered the incidence of adverse events. • The findings suggest that the inhibition of restenosis by local drug delivery may not require the implantation of stents and the prolonged release of a drug. Presented at AHA 2006