An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. PISA Orlando Deffer, MD

  2. PISA and its hydrodynamic principle. • Calculation – EROA= 6.28rsqxVa/PKreg • Measurement of PISA by color flow mapping requires adjustment of the aliasing velocity such that a well hemisphere is shown.

  3. Limitations- More accurate for central jets and for regurgitation with a circular orifice Difficulty in judging the precise location of the orifice and the flow convergance shape. Recent modifications of the described PISA method use the distance between two aliasing contours

  4. PISA Much less experience exists for assessment of AR compared to MR ECHO is performed from the apical, para-apical views, or the upper right-sternal border, with images zoomed on the valvular and supravalvular region. Due to interposition of valve tissue it’s less feasible than MR

  5. Pitfalls- Difficult in obtaining high quality images of the flow convergance region. Timing of measurement of the flow convergance radius, which should be in early diastole, closest to the peak regurgitant velocity. Ascending aortic aneurysms, which deform the valvular plane, may lead to underestimation of AR by this method. Thresholds for severe AR are an EROA>0.3cmsq or RV>60ml.