Hysteroscopic Sterilisation ESSURE TM
100 likes | 379 Vues
Hysteroscopic Sterilisation ESSURE TM. Sarah Wallage Gynae / SRH Consultant Aberdeen 2013. What is Essure ?. Vaginoscopy, 20 mins on table. Permanent IVF for subsequent pregnancy No effect on periods can have subsequent endometrial ablation
Hysteroscopic Sterilisation ESSURE TM
E N D
Presentation Transcript
Hysteroscopic Sterilisation ESSURE TM Sarah Wallage Gynae / SRH Consultant Aberdeen 2013
Permanent IVF for subsequent pregnancy • No effect on periods can have subsequent endometrial ablation • Hysteroscopy not laparoscopy • Not require GA • Takes 3 months to work • Confirmation test X ray / USS 15% need hysterosalpingogram ( HSG)
Who can have EssureTM • Not suitable after previous tubal surgery previous endometrial ablation nickel allergy ??? immunesuppressed 6/12 + HSG • Can offer if not suitable for lap sterilisation eg BMI >30 previous midline scar/ abdominoplasty not fit for GA
Essure TM • 86-96% bilateral placement – endometrium/ tubal spasm distortion • <2 % perforation/ migration - need removal • <2% expulsion • 1:1000 pregnancy rate after OK confirmation test ( 5yr data) • Outpatient procedure if LA • Back to normal within a day • Must use contraception until confirmation test
Aberdeen experience • Funding- £200 + HSG > lap steri hosp cost • No unplanned overnight stay • Clinic / Theatre ?? hysteroscope + kit 2 nurses 1 operator recovery – similar to IUD • Cycle stage- need thin endometrium provera prep
Information- staff training / clerking proforma • Preclinic pt info / GP awareness • Demand? No increase in steri referrals 2012 • Success - only 37/48 placed- ostia not clear/ tubal block • Follow up - USS X ray often difficult to interpret almost all need HSG logistics/ staff / acceptability
Summary • Another option if reversible methods are unacceptable • Similar to vasectomy – no GA, recovery time and delay till effective • Suitable for many women ineligible for lap steri • Possible in specialised clinic setting • Significant Failure rate for placing devices • Confirmation test – most need HSG