1 / 16

Ariaan van Walsum trauma surgeon Medisch Spectrum Twente Enschede The Netherlands &

The treatment of femoral neck fractures and the results of the Gannet keynote session Hip # 2 lecture # O244. Ariaan van Walsum trauma surgeon Medisch Spectrum Twente Enschede The Netherlands & W. Roerdink , J. Vroemen , H. Janzing , T. Winkelhorst.

anamartin
Télécharger la présentation

Ariaan van Walsum trauma surgeon Medisch Spectrum Twente Enschede The Netherlands &

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The treatment of femoral neck fractures and the results of the Gannetkeynote session Hip # 2lecture # O244 Ariaan van Walsum trauma surgeon Medisch Spectrum Twente Enschede The Netherlands & W. Roerdink, J. Vroemen, H. Janzing, T. Winkelhorst

  2. Conflictof Interest Disclosure Form NAME: ADP van Walsum AFFILIATION:MedischSpectrum Twente, Enschede, The Netherlands DISCLOSURE I havethefollowing potential conflict(s) of Interest toreport:

  3. Treatment of femoral neck fractures 35% REVISION RATE IN DISPLACED # 11% IN UNDISPLACED #

  4. Treatment of femoral neck fractures We got a problem!

  5. Biology BIOLOGY IS THE MOTHER OF ALL FIXATION G.R. Girdlestone

  6. Biology Achilles heel 1 Bone healing No cambium layer No periosteal callus Primary bone healing Requires: Anatomical reduction Absolute stability

  7. Biology Achilles heel 2 Viability Remaining vascularisation Reduction Fixation Revascularisation: Fixation stability Implant volume

  8. Why do implants fail?multiple parallel screws or pins • No angular stability • No rotational stability

  9. Why do implants fail?sliding hip screw devices • No rotational stability • During insertion • After insertion • +/- anti-rotation screw • High implant volume

  10. Can we make a better implant?

  11. Dynamic Locking Blade Plate“GANNET” CHARACTERISTICS: • Rotational stability • Angular stability • Dynamic compression • Minimal invasive

  12. GANNETexample Trauma female 57 year

  13. GANNETexample 3 year

  14. GANNET Failure Rate 3,9% for undisplaced fractures • N = 149 • Mean age 69 yrs 13,5% for displaced fractures • N = 192 • Mean age 61 yrs 60% for displaced unreduced fractures • N = 5

  15. CONCLUSIONfemoral neck fractures • AWARENESS! • Indications for Gannet: • Undisplaced fractures • all ages • Displaced fractures • young patients (< 65) • young elderly patients (65 – 75) • individualized on the biology • after anatomical reduction • high risk elderly patients (> 75)

  16. GANNETthe bird… BG: Бял рибояд CZ: Terej bílý DE: Basstölpel DK: Sule FI: Suula EE: Sool ES: Alcatraz FO: Súla FR: Fou de Bassan HU: Szula IS: Súla IT: Sula bassana NL: Jan-van-gent NO: Havsule PL: Głuptaków PT: Ganso-patola RO: Corbul de mare RU: Северная олуша SE: Havssula UK: Gannet THANK YOU!