1 / 17

CARIOCA Dead-time & inefficiency - UPDATE

CARIOCA Dead-time & inefficiency - UPDATE. M2 station has been reanalysed from the Davide measurements at 20 MHz BC of counting rates on the bi-gap physical channel counters at 4 luminosities ~ (0.4 - 0.6 - 0.8 - 1) x 10 33

audra
Télécharger la présentation

CARIOCA Dead-time & inefficiency - UPDATE

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. CARIOCA Dead-time &inefficiency - UPDATE M2 station hasbeenreanalysed from the Davide measurementsat 20 MHz BC of countingrates on the bi-gap physicalchannelcounters at 4 luminosities ~ (0.4 - 0.6 - 0.8 - 1) x 1033 PROCEDURE • Measureinefficiency of particlecounting • Extract the inefficiencyfraction due to CARIOCA dead-time δ • Evaluate (by MC) the δ valuegeneratingsuchinefficiency • From thisδ,evaluate the inefficiencywewillhaveat 40 MHz Allthis per region/readoutseparately G. Martellotti, G. Penso, D. Pinci

  2. Inefficiency of particlecounting • The inefficiency of the FE channelasparticlecounterismeasured by the differencebetween the measured rate R* and the rate of hittingparticlesRpart. • Rpartisnotknownbutitscales with luminosityL • 2 measurements are made at L1 and L2 > L1 • The ratioρ = (R2*/R1*) (L1/ L2) as a function of R2*- R1* • measures with goodapproximation the inefficiency of a channelwhen the channelcounting rate is R* = R2*- R1 * DEFINITIONS: Ro = BC rate = 20MHz (in a 0.7 fraction of time - 1262 BC/orbit) Rpart = rate of particleshittingeach Pad (notknown) R*= rate of Pad counting (Pad not in dead-time) Note that the rates RD* measured per secondby Davide have to be rescaled to R*=RD*/0.7 whencompared to Ro=20 MHz

  3. Inefficiency of particle counting Countingefficiency R* = R2* -R1* (GHz) The countinginefficiecyis due to 2 effects: # Non nullprobability of >1 particlehitting the samepad in the same BC (granularityeffect)  rate of padhit R < Rpartthisisnot a detector inefficiency # When the padis hit (rate R) ithas a probabilityto be in the CARIOCA dead-time δ  R*=R(1-δR*)  thisis a detector inefficiency Ifwewant to evaluatethe detector inefficiency (due to CARIOCA dead-time) from the counting rate, wehave to ‘’subtract’’ the granularityeffectcalculating R starting from Rpart

  4. GRANULARITY EFFECT can be calculated in terms of m and w k = number of interactions in the BC m = average number of interactions per BC Poissonian of k Pk = exp(-m) mk/ k! n = number of particles on the Pad w= average pad occupancy for 1 interaction when occurring k interactions in the BC <n>= k w Poissonian of n Pn = exp(-kw) (kw)n/n! Probability of pad hit in the case of k interactions = 1-exp(-kw) Overall probability of Pad-on (1 – Po)= Sk=1… exp(-m) mk/ k! (1-exp(-kw)) = 1 - exp{-m[1-exp(-w)]} A channel hit in average by mw particles/BC will have a particle rate Rpart = mw Ro and a lower rate of Pad-on R = (1–Po) Ro

  5. Ro = 20 MHz, Rpart = μωRo ω = Rpart/Roμ. • Wedon’tknowω butit can be quitewellapproximated by ω ≈ R*/Roμ • With thisωwecalculateR = Ro (1-Po) and write the dead-time formula R*/R = (1 - δR*) Giventwomeasurementsatdifferentluminositywe can calculate the ratio ρ = (R2*/R1*) (R1/ R2) = (1 - δR2*)/(1 - δR1*) where R1 and R2 are calculated with the sameω and the twodifferentμ1, μ2 Neglectingquadratictermsρ≈ 1 – δ(R2* - R1*) • Plottingρ versus (R2* - R1*) we can directlymeasureδ • Eventually, usingthisδvalue, we can recalculateω and make an iterationto improve the δvalue

  6. inefficiency due to CARIOCA dead-time R1* measuredat L1 = 0.6 x 1033 R2* measuredat L2 = 0.4 x 1033 • Ifwe express the rate in GeV, the slopeisexpressed in nano seconds • effective dead-time δeffdue to the CARIOCA dead-time folded with the time BC sequence and the detector time response (bigap time response and dead-time variance) L1-L2 efficiency R* = R2* -R1* (GHz) R* = R (1- δeffR*) The inefficiencyat R* isδeffR*

  7. # Channelsnothaving a meaningfullmeasurements for all the luminosities are eliminated # A fewmeasurementstoo far from the meanvalue are eliminated # The averagein bins of R2* – R1* can be performed  Linear Fit Apparently a change in luminosity of few per mill SLOPE = effective dead-time δeff = 48.5 ns efficiency GHz R* = R2* -R1* R* = R2* -R1*

  8. At a first look the slopesobtained for the otherluminositycombinationsappear to be consistent (a more detailedcomparisonis in progress) Luminosity = (4-6, 4-8, 4-10, 6-8, 6-10, 8-10) x 10 32 δeff = (48.5, 50.1, 47.9, 48.6, 49.6, 46.9 )ns Averagevalueδeff ≈ 48±1.5 ns Structures are visible – to be interpretedas due to significantlydifferentδvalues for differentcases(regions – wires – cathodes). Wemayalsohaveslightchangesof luminosityduring the measurements of data samplesused Nextstepisto analyseseparatelythe different ‘’regions’’

  9. L1-L2 Region R1 • Short leverarmwhenanalysingonlyoneregion • Analysethe combinationL1-L4 • Neverthelesswe can saythat • The slopeδeffissignificantlydifferent in differentregions. Itishigher for cathodesthan for wires • # Thisis due to the differentcapacitancebutalso to the hit quality (differentaveragechargedeposited) • # slightlydifferentextrapolations to zero. Possibly a change in luminosity of few per mill in between the measurements for wires and cathodes δeff 64ns δeff 65ns efficiency δeff 44ns δeff 51ns

  10. L1-L2 Region R2 Evenshorterleverarmfor R2 Certainlytoo short for R3 R4 to measure a meaningfulslope  AnalyseL1-L4 δeff53ns δeff53ns δeff R1 cathodes64-65 ns R1 wires44-51ns R2 cathodes53 ns R2 wires46-50ns δeff46ns δeff50ns

  11. Toy Monte Carlo Given # a value of CARIOCA dead time δcar (meanvalueand variance) # the time response of the bi-gap chamberσt # the BC rate = 20 MHz or 40 MHz • Inefficiency due to δiscalculated (expressed in terms of effective dead time δeff)

  12. ASSUMED : bi-gap gaussianresponseσt = 5ns δcar = 70 ± 11 ns (gaussian) BCR = 20 MHz BC structure PreviousBCscontributing to dead time Current BC

  13. For each <δcar> assumed (with itsvariance) MC  inefficiency Simple gaussianshapeshavebeenassumed for the time fluctuations. The maincontribution to the smoothingcomes from the δcar varianceassumed to increase with the δcar value. From the Rieglermeasurements I wouldsaythat the realbehaviouris in between the two last hypoteses (nearer to the last one). BCR = 20 MHz σt = 5ns σδcar= δcar /6 σt = 0 σδcar = 0 σt = 4 ns σδcar= δcar/10 δeff (ns) inefficiency DT eletctronics (ns) δcar (ns)

  14. BCR = 20 MHz For each‘’inefficiency’’ δeffmeasured in a region, δcar isevaluated Ifδeff ~ 50 ns the δcar isnotverywelldetermined. σt = 5ns σδcar = δcar/6 σt = 4nsσδcar = δcar/10 R1 cathodes R2 cathodes δeff (ns) Averageallregions δcar (ns) δcar (ns)

  15. For BCR = 40 MHzwith anyreasonableassumption on time fluctuations (in particular for the δcar variance) wewillhave a sufficientsmoothing inefficiencywill be ~ proportional to CARIOCA dead-time - δeff ~ δcar -12.5 ns σt = 4ns, σδcar = δcar /10 σt = 5ns, σδcar = δcar /6 δeff (ns) inefficiency δcar (ns)

  16. ‘’PROVISIONAL’’ Summarytable Station M2 δeff (20 MHz)δcarioca δeff (40MHz) Average on allregions 48ns 67-73ns 54-61ns • R1 cathodes 64-65ns 90-95ns 77-83ns • R1 wires44-51ns ? 60-80ns 47-68ns • R2 cathodes 53ns 76-85ns 63-73ns • R2 wires46-50ns ? • R3 ? • R4 ? Inefficiencyat a rate R (measured in GHz) isobtained from the dead-time formula R*= R(1 -δeff R*) A reliableerror estimate isneeded Furtheranalysisisneeded of the samples with largerleverarm: L1-L4

  17. Summary # Inefficiency of physicalchannels due to CARIOCA dead time ismeasured from countingrates in terms of an effectiveδeff # preliminaryresults indicate that : 1) The averageδeffis of the order of 50ns 2)δeffissignificantlydifferent for differentregions/readouts. Itishigher in R1 (and for cathodesishigherthan for wires) – Assuming an averageδcarseems to be a crude approximation (optimistic for the innerregion) 3) From the δeff, the (larger) CARIOCA dead time (δcar) isevaluatedwith a toy MC makingsimpleassumptions on itsvariance and on the bi-gap time response. 4) Whengoingat 40 MHz BCR, wewillhave a (larger) inefficiencywellapproximated by a δeff≈δcar – 12.5 ns  The inefficiency of R1 cathodesisvery high. At the high ratesforeseen for future, itseemsthatinefficiencywill be too high to work with the END of wires and cathodes # With the muon FE counters, carefullmeasurementswouldallow to monitor relative luminosities with very high precision