1 / 18

What Dw does not tell you.

New QC Procedure for Eichrom Resins: Why We’re Changing Joel Williamson, Jill Bernstein, Dan McAlister and Phil Horwitz Eichrom Technologies, Inc.

Télécharger la présentation

What Dw does not tell you.

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. New QC Procedure for Eichrom Resins: Why We’re ChangingJoel Williamson, Jill Bernstein,Dan McAlister and Phil HorwitzEichrom Technologies, Inc.

  2. Since the formation of Eichrom, an essential feature of our quality control system was the measurement of Dw for each batch of new resin manufactured.

  3. Dw was measured under conditions in which the analyte is strongly retained by the resin and weakly retained by the resin. • What do these types of measurements tell you? • Did you use the correct extractant? • Did you use the correct quantity of extractant?

  4. What Dw does not tell you. • How completely will an analyte be retained by the column? In other words, how much breakthrough will occur? • When exactly will an analyte elute from the column? • How efficiently will an analyte elute from the column? (% recovery) • How much cross contamination will occur between two or more analytes?

  5. New QC Protocols • Extractants (Solvent extraction test) -helps to determine purity of extractant 2) Every batch of resin (Dw Test) -ensures correct extractant used 3) Resin with new lot of at least one starting material (New column QC) a) 1 chromatogram b) 4 vacuum box tests -ensures good performance of resin in most commonly used application

  6. Chromatogram Simplified version of common Eichrom method performed Aliquots taken for gamma, LSC or ICP-AES analysis cpm or ppm plotted vs Bed Volumes

  7. Chromatogram • Important Parameters • Recovery of analytes (>95%) • Cross-contamination (<1-3%) • Reproducibility (4-5 replicates)

  8. Vacuum Box Test Essentially the same as the chromatogram test, however, fewer fractions are collected Provides reproducibility data

  9. General Time Line 1) Determine separation schemes to test each Eichrom resin. 2) Perform liquid-liquid extraction (extractants) and chromatogram and vacuum box tests on quality control retains (10-12 resin lots) to determine reasonable recovery/separation metrics for future lots. 3) Implement new QC procedure and provide improved certificates of analysis for each lot of Eichrom resin. UTEVA Ln (a+s) Pb Resin TRU TEVA DGA, n DGA, br Tritium Nickle MnO2 Sr Resin Beryllium Anion Cation 1) 2) 3) Progress as of 4-22-05 Goal: Complete 4 main resins by summer 2005

  10. Proposed QC Protocols TRU: Load: Am and Pu from 3 M HNO3 Rinse: 3 M HNO3 + 0.1 M NaNO2 Strip Am with 4 M HCl Strip Pu with 0.03 M ammonium oxalate TEVA: Load: Th and Pu from 3 M HNO3 + 0.1 M NaNO2 Rinse: 3 M HNO3 + 0.1 M NaNO2 Strip Th with 6 M HCl Strip Pu with 0.02 M HCl + 0.02 M HF

  11. Proposed QC Protocols UTEVA: Load: Th and U from 4 M HNO3 Rinse: 4 M HNO3 Strip Th with 6 M HCl Strip U with 0.1 HCl Sr: Load: Sr and Ba from 8 M HNO3 Rinse: 8 M HNO3 Strip Sr with 0.05 M HNO3

  12. Results (TRU)

  13. Results (TRU) >95% yields for Pu and Am Good separation of Pu and Am <1% Pu in Am <1.5% Am in Pu Very reproducible different lots same lot Only significant variability observed was the amount of Pu remaining on column

  14. Results (TEVA)

  15. Results (TEVA) >95% yields for Pu and Th Good separation of Pu and Th <1% Pu in Th <2% Th in Pu Very reproducible different lots same lot Only significant variability observed was the amount of Pu remaining on column

  16. Conclusions • Eichrom is implementing a new QC program. • The new QC program will include liquid-liquid extraction tests (extractants), Dw tests (every resin lot) and column elution tests (resin lots with new raw materials). • So far, TRU and TEVA from 2004 have been tested to generate data from which metrics to evaluate new resin lots will be determined. • UTEVA and Sr Resin tests are next, followed by DGA and the remaining Eichrom products.

More Related