An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. 신제품 출시 - EliA MPOS

  2. Why a new EliA MPOS test? • 검사성능 개선 • 기존의 EliA MPO에서 민감도 향상 • 타사 제품보다 월등한 성능 • EliA PR3S와 함께 성능개선 • 두 가지 검사 모두 anchor technique을 이용하여 ANCA 검사의 성능향상 • 새로운 international reference에 따라 표준화, International Units (IU/ml)단위의 결과 제공

  3. The new sensitive anti-MPO test Product name Package size Article number • EliA MPOS Well 48 14-5537-01 new • “S”는 ”sensitivity”를 의미 • High sensitivity는 anchor technique를 이용한 indirect coating 방법으로 antigen이 더 잘 보이도록 해서 얻은 결과

  4. EliA MPOS- Calibration • CDC MPO-ANCA Human Reference Serum #15와 일치하는 calibration curve를 사용하여 International Units (IU/ml)으로 결과 산출 • 따라서, 기존의 EliA MPO 검사와 다른 참고치를 적용 • negative < 3.5 IU/ml • equivocal 3.5 – 5.0 IU/ml • positive > 5.0 IU/ml

  5. EliA MPOS- Blood Donors and Normal Range 400 blood donors were measured on Phadia 250. A substudy using 70 blood donors showed good comparison of Phadia 100. Test Unit No of Mean 95% 99% samples value percentile percentile EliA MPOS IU/ml 400 0.6 0.9 1.5 Frequency IU/ml All blood donors far below cut-off. Good discrimination.

  6. EliA MPOS- Serum / Plasma 검체 간의 상관관계 Serum과 모든 types의 plasma 검체 사용가능

  7. EliA MPOS - 임상적 검사성능, 질환의 변별력 평가 Samples used for performance validation n • Microscopic polyangiitis (MPA) 80 • Churg-Strauss syndrome (CSS) 60 • Necrotizing crescentic glomerulonephritis (NCGN) 12 • Disease controls 150 • Non-ANCA associated vasculitis (NAAV) 22 • Ulcerative colitis 30 • Crohn’s disease 30 • Systemic lupus erythematodes (SLE) 15 • Systemic Scleroderma 15 • Infections 38 • Total 302

  8. EliA MPOS -임상적 검사성능 Good sensitivity in MPA. Only one false positive (EBV infection). Microsc. Polyangiitis

  9. High Specificity • 단 하나의 disease control만 양성 • 이 검체는 EBV infection환자 • 이 검체에 대해 IFA ANCA검사는 높은 양성의 p-ANCA pattern을 나타냈고, 모든 타사 검사법도 MPO-antibodies를 검출

  10. MPO antibodies in Churg-Strauss Syndrome • 논문에 따르면, CSS 환자의 약 38%가 IFA 검사에서 양성.이 양성환자의 약 ¾이 MPO antibodies에 의함(30%) (Sinico RA et al, 2005, Arthritis Rheum 52:2926-35) • 자사 평가에 의하면 60명의 CSS 환자 혈청 중 23명에서 ANCA pattern이 나타남 (38.3%). 하지만 단 10명만이 IFA 검사결과 전형적인 pANCA pattern을 보임 (16.7%). 11명의 혈청에서 MPO antibodies가 검출됨 (18%). • EliAMPOS보다 민감도가 좋은 타사 제품은 없었기 때문에, 낮은 민감도는 EliA 때문이 아니라 sample panel 때문에발생. • 평가에 사용된 Churg-Strauss 검체에서 anti-MPO antibodies의 prevalence가 낮아 2번의 assay 결과를 가지고 계산 (with and without Churg-Strauss sera).

  11. EliA MPOS - 임상적 검사성능, 타사 제품과의 비교평가 • 주요 타사 제품과의 비교: • EliA MPO • Inova QuantaLite MPO IgG ELISA • Orgentec Anti-MPO (p-ANCA) • Euroimmun Anti-MPO-ELISA (IgG) • Eurodiagnostica Wieslab Capture-MPO-ANCA (only test with IU/ml) • Nippro • The Binding Site • IFA와의 비교

  12. 3. EliA MPOS - 최고의 민감도, 특이도!타사 제품과의 비교평가

  13. 3. EliA MPOS - 최고의 민감도, 특이도!Comparison to Competition with preset specificity The performance of tests can be evaluated better if you look at the sensitivity at a certain set specificity. Using ROC analysis we calculated the sensitivity at a specificity of 99%. This evaluation shows that EliA MPOS performs best and this is not dependent on the recommended cut-off.

  14. EliA MPOS – 임상적 검사성능, 타사 제품과의 비교평가Churg-Strauss Sera 제외

  15. EliA MPOS – 임상적 검사성능, 타사 제품과의 비교평가Churg-Strauss Sera 제외 The performance of tests can be evaluated better if you look at the sensitivity at a certain set specificity. Using ROC analysis we calculated the sensitivity at a specificity of 99%. This evaluation shows that EliA MPOS performs best and this is not dependent on the recommended cut-off.

  16. EliA MPOS – EliA MPO와 ROC분석 비교 EliA MPOSperforms clearly better than the “old” EliA MPO. EliA MPOS EliA MPO

  17. EliA MPOS – ROC Analysis compared to Competition:Wieslab MPO Capture and Orgentec Anti-MPO ROC Analysis reveals that EliA MPOS performs better than the MPO tests from Wieslab (Eurodiagnostica) and Orgentec. EliA MPOSvs. Orgentec Anti-MPO EliA MPOSvs. Orgentec MPO Capture EliA MPOSvs. Wieslab MPO Capture

  18. EliA MPOS – ROC Analysis compared to Competition:Inova and Euroimmun Anti-MPO Assay ROC Analysis reveals that EliA MPOS performs better than the Euroimmun and the Inova Anti-MPO tests. EliA MPOSvs. Invoa EliA MPOSvs. Euroimmun MPO

  19. EliA MPOS – Comparison with IFA on granulocytes** An ANCA pattern was found in 63.2 % of ANCA-associated vasculitis (AAV) and in 29.3 % of controls. The classical MPO-associated pANCA pattern was found in 35.5 % of AAV and 2.7 % of controls. *x-ANCA: in most cases positive in ethanol-fixed granulocytes but negative in formalin-fixation. **Measurements are based on the MPOs serum panel (see slide 8) plus 30 additional sera from patients with Wegener’s granulomatosis in % Almost 30% of non-AAV are positive in IFA. pANCA is less sensitive and less specific than EliA MPOS. The clinical usefulness of IFA as screening test is questionable.

  20. EliA MPOS - 임상적 검사성능, 요약 • EliA MPOS검사는 기존의 EliA MPO 보다 민감도 향상. • EliA MPOS검사는 타사 검사법보다 검사성능 우월. • EliA MPOS는 Wieslab의 Capture-MPO-ANCA (capture assay)보다 약간 더 우수하고,이 두 가지 검사만이 IU/ml (standardized to the CDC standard)을 사용할 수 있기 때문에 최고의 검사법으로 인정. • IFA는 AAV (ANCA associated vasculitis)에서 민감도가 높지만 특이도는 아주 낮음. Disease control의¼ 이상에서 ANCA 양성 (granulocytes)의 결과를 보임. • pANCA는 EliA MPOS에 비해 민감도와 특이도가 모두 낮음. • 따라서, IFA의 임상적 유용성은 의문의 여지가 있음.

  21. EliA GBM for detection of anti-GBM in ANCA-positive samples • ANCA 양성 환자는 GBM antibodies를 반드시 확인해야 한다! • ANCA 양성 검체의 5-14% 정도는 anti-GBM도 양성. Anti-GBM 음성인 vasculitis 환자보다 두 검사에서 모두 양성인 환자가 질환에 대한 예후가 더 심각.(Levy et al, 2004, see PoM 01/05) • 빠르게 악화하는 glomerulonephritis로 의심되는 환자 중 약 2%가 ANCA와 anti-GBM 모두 양성. 예후는 ANCA만 양성인 경우가 anti-GBM만 양성인 경우보다 좋다.(Jayne DRW et al, Kidney Int 1990; 37:965-70)

  22. EliA MPOS and EliA PR3SCutting edge automated ANCA testing!