Micro PIV

Micro PIV. An optical diagnostic technique for microfluidics (e.g. MEMS, biological tissues, inkjet printer head) Requirements: Measure instantaneously 10 3 - 10 4 vectors Spatial resolution of 1 - 10 m m Wide velocity range: 50 m m/s - 400 m/s Accurate to within 3% full scale

Télécharger la présentation

Micro PIV

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Micro PIV • An optical diagnostic technique for microfluidics (e.g. MEMS, biological tissues, inkjet printer head) Requirements: • Measure instantaneously 103 - 104 vectors • Spatial resolution of 1 - 10 mm • Wide velocity range: 50 mm/s - 400 m/s • Accurate to within 3% full scale References • Meinhart, Wereley and Santiago (1999) • Santiago et al. (1998) • Private communication

  2. Video Microscopy • Mature technology in bio-medical fields The smallest resolvable size dp = l/NA , NA (Numerical Aperture)= n sinq For comparison, recall diffraction limit for camera: ddiff = 2.44l/(D/f)=2.44l(f#) • Microscopy + PIV Resolve particles of sub-microns Measurement of particle displacement Image field: 30~300mm n q dp

  3. Field of View: 30 ~ 300 mm Vector Spacing: 1 ~ 10 mm Interrogation Cell: 2 ~ 20 mm (50 % overlap) min. 10 pairs of particles for correlation “Plane” Thickness dz: Depth of Field of microscope ~ 1mm 30 ~ 300 mm 1 ~ 10 mm 2 ~ 20 mm Laser sheet thickness ~ 1 mm Micro PIV vs. PIV Shrink 1000 times

  4. Micro PIV Small-- Follow flow Do not clog the device Do not alter fluid property But not too small-- Suppress Brownian motion Generate enough light signal Dp = 0.3 ~ 0.7 mm Regular PIV Small enough to track flow, need to be detectable by the camera Dp = 3 ~ 30 mm Tracer Particles

  5. Challenges by Sub-micron Particles • 1. Optical Resolution: need Dp = 300 – 700 nm (Nd:YAG: l ~ 500 nm) Visible light400 nm-----750 nm If NA <1, cannot resolve dp less than l sin q <1 n: index of refraction between specimen & objective • 2. Low Light Signal

  6. Solutions • Oil immersion lens (n  1.5) to get NA >1 NA =1.4for 60x - 100x objectives • Fluorescence (epi-illumination, reflection) dp < l & stronger signal • Differential Interference Contrast (DIC) microscopy Shearing interference to highlight refraction change

  7. Mercury arc lamp Exposure t ~ 2 ms Pulse delay Dt ~ 100 ms (Also depend on camera transfer) Velocity up to 50 mm/s Pulsed laser (Dual Nd:YAG laser) t ~ 5 ns Dt ~ 500 ns up to 1 m/s Light Source and Camera Digital CCD Camera (1030 x 1300 x 12 bit cooled interlined transfer can record back-to-back images within 500 ns)

  8. Data Processing • Correlation • Significant Noise: • Out-of-plane motion • Brownian motion • Ensemble-averaging correlation technique (average 20 instantaneous correlations) • Limited to steady or periodic flows

  9. Example 1 • Santiago et al. (1998)

  10. Result • Santiago et al. (1998)

  11. Meinhart, Wereley and Santiago (1999) Example 2

  12. Meinhart, Wereley and Santiago (1999) Result Ensemble-averaged velocity-vector field measured in a 30 mm deep, 300 mm wide, 25 mm channel. The spatial resolution is 13.6 mm x 4.4 mm away from the wall, and 13.6 mm x 0.9 mm near the wall. A 50% overlap between interrogation spots yields a velocity vector spacing of 450 nm in the wall- normal direction near the wall

  13. Inkjet Printer Head • Field of view 50 ~ 500 mm • Need objective lens working distance >1mm (Cover Glass) Smaller NA Larger particle size (~ 0.6) (~ 0.7 mm) • Unsteady flow in the cycle of droplet ejection: need instantaneous or phase-averaged measurement

  14. Basic Limitation of Micro PIV • DOF (~ 1mm) limits to strictly 2D flow • Not only 2D vector map, • Out-of-plane motion can cause measurement to fail • Hence must select a plane with only 2D motion PIV Plane