An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. DIATHERMY PT 156: Physical Agents 1 AilaNica J. Bandong, PTRP Instructor Department of Physical Therapy UP-College of Allied Medical Professions

  2. Objectives At the end of the lecture, the students should be able to: • Define diathermy • Identify the parts of the SWD/MWD machine • Describe how heat is produced in SWD/MWD • Enumerate the therapeutic effects of diathermy • Differentiate the types of SWD • Enumerate the indications, contraindications, and precautions/guidelines for the use of the modality • Use clinical decision making skill in choosing appropriate modality • Apply evidence regarding use of SWD/MWD

  3. What is DIATHERMY? • The use of non-ionizing electromagnetic energy from the radio-frequency spectrum as therapeutic agent

  4. Types of Diathermy • Long wave - longest wavelength 300 – 30 m - most penetrating - no longer utilized due to high potential of causing burns and interference with radio transmissions • Shortwave • Microwave

  5. SHORTWAVE DIATHERMY

  6. Superficial and deep heating modality • Frequency - 27.12 MHz • Wavelength - 11 m • Method of HeatTransfer - Conversion • Manner of Delivery - continuous - pulsed

  7. Continuous vs Pulsed SWD

  8. PULSED SWD • Pulse Repetition Rate (PRR) - 15 to 800 Hz • Pulse Duration (PD) - 25 to 400 microseconds • Peak Pulse Power (PPP) - 100 to 1000 watts • Duration - 20 minutes (5-15 acute; 10-20 chronic)

  9. Calculating Mean Power in PSWD Cycle Duration = 1000 / PRR % cycle SWD delivered = (PD x 100) / Cycle Duration Mean Power delivered = PPP x % cycle SWD

  10. SHORTWAVE DIATHERMY UNIT

  11. HEAT PRODUCTION • Dependent on: SPECIFIC ABSORPTION RATE Tissue conductivity charged molecules dipolar molecules non-polar molecules Electrical field magnitude

  12. Charged Molecules • Ions and certain proteins • Molecules are accelerated along lines of electric force • Most efficient way of heat production + - - + + - + -

  13. Dipolar Molecules • Water and some proteins • Positive pole of the molecule aligns itself to the negative pole of the electric field (vice versa) • Moderately efficient heat production

  14. Non-Polar Molecules • Fat cells • Electron cloud is distorted but negligible heat is produced • Least efficient heat production

  15. ! Clinical Implication ! • Blood, having high ionic content, is a good conductor  vascular tissues as well • Metal and sweat are good conductors  if metal implants and sweat are present within the electric field, may cause burn

  16. SHORTWAVE DIATHERMYCondenser InductionField Field

  17. CONDENSER FIELD • Patient’s tissues are used as DIELECTRIC between the conducting electrodes • Oscillation and rotation of the molecules of the tissues produces heat • Either flexible metal plates (malleable) or rigid metal discs can be used as electrodes • Can be applied in 3 ways: contraplanar, coplanar, or longitudinal

  18. Ways of Application • Contraplanar - aka Transverse positioning - plates are on either side of the limb

  19. Ways of Application • Coplanar - plates parallel with the longitudinal section of the body part; same side

  20. Ways of Application • Longitudinal - plates are placed at each end of the limb

  21. In what way should SWD be applied? • No conclusive evidence as to the technique of application that will produce the most effect on the heated tissue (Kitchen and Bazin, 1996)

  22. ! GUIDELINES ! Electrodes should be: • Equal in size • Slightly larger than the area treated • Equidistant and at right angles to the skin surface

  23. INDUCTION FIELD • Patient is in the electromagnetic field or the electric circuit  produce strong magnetic field  induce electrical currents within the body (EDDY currents) • Utilizes either an insulated cable or an inductive coil applicator

  24. Arrangement of Inductive Coil • Monode: coil arranged in one plane • Hinged Diplode: permits electrode to be positioned at various angles around the three sides of the body part, or in one plane

  25. Which is more effective: condenser or induction field? • Some studies argue that inductive diathermy produces greater increase in temperature of deeper tissues compared to condenser/capacitive technique • Any deep effects following capacitive technique requires considerable increase in superficial tissue temperature

  26. THERAPEUTIC EFFECTS(continuous and pulsed) • Increase blood flow • Assist in resolution of inflammation • Increase extensibility of deep collagen tissue • Decrease joint stiffness • Relieve deep muscle pain and spasm

  27. INDICATIONS • Soft tissue healing - conflicting evidence as regards effectiveness of SWD - controlled animal studies revealed insignificant results as well as trials involving human subjects (Kitchen and Bazin, 1996); to date, no studies in the treatment setting was conducted

  28. INDICATIONS • Recent ankle injuries - inconclusive results following three double-blind protocols (Kitchen and Bazin)

  29. INDICATIONS • Pain Syndromes - Pulsed SWD may provide better pain relief in some musculoskeletal conditions (neck and back) than SWD A.Nerve Regeneration - studies were done on cats and rats - PSWD induced regeneration of axons, acceleration and recovery of nerve conduction B. Osteoarthritis - no established effect C. Post-operative - insignificant (abdominal surgery

  30. MICROWAVEDIATHERMY

  31. Superficial and deep heating • Frequency: 300 MHz to 300 GHz • Wavelength: 1m to 1mm • Therapeutic Parameters: A. 122.5 mm – 2456 MHz B. 327 mm – 915 MHz C. 690 mm – 433.9 MHz • Dosage: acute 5 to 15 minutes chronic 10 to 20 minutes

  32. MICROWAVE DIATHERMY APPARATUS

  33. How is microwave produced? • Direct current (DC) is shunted to the cathode in the magnetron valve • Release of electrons from the cathode to the multi-cavity anode valve • Electrons oscillate at predetermined frequency • High frequency alternating current is transmitted along a coaxial cable • Coaxial cable transmits energy to a director

  34. Fate of Microwave • Absorbed - energy is taken up by the material • Transmitted - pass through the material without being absorbed • Refracted - direction of propagation is altered • Reflected - turned back from the surface

  35. THERAPEUTIC EFFECTS • Increased blood flow or circulation to the area • Increased tissue temperature • Increased metabolism • Facilitate relaxation • Increased pain threshold • Decreased blood viscosity

  36. INDICATIONS • Soft tissue injury • Mobilization • Pain relief

  37. CONTRAINDICATIONS • Pacemakers • Metal implants • Impaired sensation • Pregnancy • Hemorrhage • Ischemic Tissue • Testicles and eyes • Malignant CA • Active TB • Fever • Thrombosis • X-ray exposure • Uncooperative patient • Areas of poor circulation

  38. PRECAUTIONS • Operator should observe caution when handling the machine: same contraindications apply

  39. References Gorgon, E. J. (2004). Lecture notes on high frequency currents: Shortwave and microwave diathermy. University of the Philippines- College of Allied Medical Professions. Hayes, K. W. (1993). Manual for physical agents (4th Ed). Connecticut: Appleton and Lange. Hecox, B., Mehreteab, T. A., and Weisberg, J. (1994). Physical agents: A comprehensive text for physical therapists. Connecticut: Appleton and Lange. Kitchen, S. and Bazin, S. (1996). Clayton’s electrotherapy (10th ed). Philadelphia: W.B. Saunders Company. Low, R. Reed, A. (1995). Electrotherapy explained: Principles and practice (2nd Ed). Oxford: Butterworth-Heinemann Ltd. Michlovitz, S. L. (1996). Thermal agents in rehabilitation (3rd Ed). Philadelphia: F. A. Davis Company.