Polycrystalline CVD Diamonds for the Beam Calorimeter of the ILC

Polycrystalline CVD Diamonds for the Beam Calorimeter of the ILC. C.Grah ILC ECFA 2006 Valencia, 9 th November 2006. Contents. Very forward region of the detectors for the International Linear Collider BeamCal – requirements Testbeam at CERN Linearity of the response of CVD diamonds

Télécharger la présentation

Polycrystalline CVD Diamonds for the Beam Calorimeter of the ILC

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Polycrystalline CVD Diamonds for the Beam Calorimeter of the ILC C.Grah ILC ECFA 2006 Valencia, 9th November 2006

  2. Contents • Very forward region of the detectors for the International Linear Collider • BeamCal – requirements • Testbeam at CERN • Linearity of the response of CVD diamonds • Testbeam at the TU Darmstadt • Radiation hardness of CVD diamonds • Conclusions C.Grah: CVD Diamonds

  3. Very Forward Region of the ILC Detectors • The purpose of the instrumentation of the very forward region is: • Hermeticity: increase the coverage to polar angles > 5mrad • Optimize Luminosity BeamCal Interaction point • EM calorimeter with sandwich structure: • 30 layers of 1 X0 • 3.5mm W and 0.3mm sensor • Angular coverage from 5mrad to 28 mrad • Moliére radius RM ≈ 1cm • Segmentation between 0.5 and 0.8 x RM LDC C.Grah: CVD Diamonds

  4. The Challenges for BeamCal e+ e- e+ e- γ Interaction γ e- e- e.g. Breit-Wheeler process Creation of beamstrahlung at the ILC • e+e- pairs from beamstrahlung are deflected into the BeamCal • 15000 e+e- per BX => 10 – 20 TeV total energy dep. • ~ 10 MGy per year strongly dependent on the beam and magnetic field configuration => radiation hard sensors • Detect the signature of single high energetic particles on top of the background. => high dynamic range/linearity ≈ 1 MGy/a ≈ 5 MGy/a C.Grah: CVD Diamonds

  5. Polycrystalline Chemical Vapour Deposited Diamonds (courtesy of IAF) • pCVD diamonds are an interesting material: • radiation hardness (e.g. LHC pixel detectors) • advantageous properties like: high mobility, low εR = 5.7, thermal conductivity • availability on wafer scale • Samples from two manufacturers are under investigation: • Element SixTM • Fraunhofer Institute for Applied Solid-State Physics – IAF • 1 x 1 cm2 • 200-900 μm thick (typical thickness 300μm) • Ti(/Pt)/Au metallization (courtesy of IAF) C.Grah: CVD Diamonds

  6. Linearity Test at CERN PS 17 s 10 ns Hadronic beam, 3 & 5 GeV Fast extraction mode ~104-107 particles / ~10 ns Setup Beam Scint.+ PMTs. Diamond gate signal ADC Response of diamond sensor to beam particles (no preamplifier/attenuated) Photomultiplier signals C.Grah: CVD Diamonds

  7. Response vs. Particle Fluence Fraunhofer IAF Element Six E64 FAP2 30% deviation from a linear response for a particle fluence up to ~106 MIP/cm2 The deviation is at the level of the systematic error of the fluence calibration. C.Grah: CVD Diamonds

  8. High Dose Irradiation Superconducting DArmstadt LINear ACcelerator Technical University of Darmstadt • Irradiation up to several MGy using the injector line of the S-DALINAC: 10 ± 0.015 MeV and beam currents from 10 to 100 nA corresponding to about 59 to 590 kGy/h V.Drugakov 2X0 Energy spectrum of shower particles in BeamCal C.Grah: CVD Diamonds

  9. Preparations and Programme preamp box GEANT4 simulation of the geometry Beam setup absorber collimator Apply HV to the DUT Measure CCD ~20 min Irradiate the sample ~1 hour => R = NFC/NSensor = 0.98 <Edep>/particle = 5.63 MeV/cm CCD: Charge Collection Distance C.Grah: CVD Diamonds

  10. Realization: Beam Setup exit window of beam line collimator (IColl) Faraday cup (IFC, TFC) sensor box (IDia, TDia, HV) C.Grah: CVD Diamonds

  11. CCD Setup PA Sr90 ADC diamond delay Sr90 source Scint. discr PM1 & Gate discr PM2 Preamplifier Sensor box Trigger box Setups partly financed by EUDET. typical spectrum of an E6 sensor C.Grah: CVD Diamonds

  12. Results: CCD vs. Dose 100 nA (FAP5) CCD = Qmeas/Qinduced x thickness Bias voltage = 400V ≈> 1 V/μm Silicon starts to degrade at 30 kGy. High leakage currents. Not recoverable. After absorbing 7MGy: CVD diamonds still operational. 100 nA (E6_4p) C.Grah: CVD Diamonds

  13. Behaviour after Irradiation No significant change of the current-voltage characteristics up to 1.5 MGy. Slight increase of currents for higher doses. C.Grah: CVD Diamonds

  14. CCD Behaviour after Irradiation ~ -80% ~ -30% after 7 MGy after 1.5 MGy C.Grah: CVD Diamonds

  15. After Irradiation: IAF Sample before/after ~ 7MGy ▪before irradiation ▪after irradiation strong „pumping“ behaviour. signal recovery after 20 Gy ▪FAP 5 irradiated, 1st measurement ▪FAP 5 irradiated, additional 20 Gy C.Grah: CVD Diamonds

  16. Summary • PCVD diamonds are an interesting material as sensors for the BeamCal. • The linearity of pCVD diamonds up to particle fluences of 106 particles/cm2/10ns is better than 30%. • High dose irradiation using 10MeV electrons shows: • all CVD diamonds stay functional even after absorbing up to 7MGy. • degradation of the signal at high doses and dose rates. • partial recovery of the signal after absorbing additional small doses (~20 Gy). • wide variation of the signal sizes as a function of the absorbed dose is an issue for the use as a sensor for BeamCal. • Started to investigate also other materials (GaAs, SiC) as alternative. • Successfull irradiation-testbeam at the S-DALINAC. We plan to repeat the irradiation with more and other types of sensors next year. http://www-zeuthen.desy.de/ILC/fcal/ C.Grah: CVD Diamonds