1 / 1

Introduction

Figure 4 - BTEX Separation. Figure 1 - Plunger Position of Diaphragm Valve. a.). b.). Picture taken from works by Pawliszyn. Figure 3 - SPME Device.

ciara
Télécharger la présentation

Introduction

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Figure 4 - BTEX Separation Figure 1 - Plunger Position of Diaphragm Valve a.) b.) Picture taken from works by Pawliszyn Figure 3 - SPME Device Monitoring the desorption of analytes from nonpolar SPME fibers using high speed gas chromatographyAuthors: Kimberly Jasch, Tony Borgerding*Department of Chemistry, University of St. Thomas, St. Paul, MN 55105 Introduction Methods Results • Why study this? • Volatile Organic Compounds (VOCs) • Can be found all over the world • Many are toxic • i.e. over exposure to benzene can contribute to leukemia • Very reactive in the atmosphere • i.e. methane contributes to global warming • How do we study this? • Gas chromatography is a widely accepted practice used to separate and identify compounds. • However, not an efficient method for continuous monitoring. • Other methods that could be used for monitoring are forms of spectroscopy (MS or IR). • However, these are not practical for identifying between similar molecules. • Typically, an analyte is injected into a long capillary column where a Flame Ionization Detector (FID) detects a signal which is sent to a recording device. • Longer columns means long separation times of 15 minutes or longer • Concentrations can change significantly in that time interval • Continuous monitoring the presence of VOCs is not possible • Solution? • High speed gas chromatography (HSGC) • fast separation • continuous monitoring possible • Goals • Make using the GC for monitoring a more efficient process • Use the developed HSGC process to monitor more complex systems • Gas Chromatography Apparatus • Hewlett Packard 5890A Chromatograph with a Flame Ionization Detector (FID) • Uses a 6-port valve secured to the ceiling of the oven • Samples made in 1-liter Tedlar sampling bag with septa • Gaseous samples injected with syringe or SPME and pushed through the valve with a flow of helium gas. • When injected, sample travels through column, separates, and is detected by FID • Diaphragm Valve • 6-port, 2-position diaphragm valve. • Requires only a small amount of pressure to switch positions • Figures 1 and 2 • Quick injection pulses possible • Injection Pulse • Narrower injection pulse allows for less sample to be placed into the loop • Better resolution • More reproducibility • Pulses of 6ms were possible from the diaphragm valve • Pulses of 5ms or less, consistency is lost • Sampling • benzene, toluene, ethylbenzene, and o-xylene • 8L of each syringed into 1-L bag • Solid-Phase Microextractor (SPME) • What is it? • Refer to Figure 3 • A sampling device for liquids or gases • Nonpolar fibers of PDMS or Carboxen were used • HSGC can monitor the desorption of the analytes from the fibers • Limitations: not many studies of desorption time performed • Separation of BTEX • Completed in as fast as 5.4 seconds • Refer to Figure 4 • Desorption Time and Injection Variation • Effect of Flow • Refer to Figure 5 • Peaks will come off faster • Note: differences in integrated results within each graph are independent of the flow rates • Retention times remain constant • PDMS tends to hold analyte longer than carboxen • Effect of Temperature • Amplitudes of peaks increase significantly • Temperatures of 70˚C and lower will have good resolution. • No changes in retention times of BTEX • Note: each peak in Figure 6 is separated into four separate peaks like those in Figure 4 • Carboxen vs. PDMS • On average, PDMS will release the analyte slower as the flow rates increase • Amplitudes from PDMS consistently greater than those given from Carboxen • Refer to Figure 7 for integrated results of the signals recorded a.) b.) Picture taken from Valco catalogue Figure 6 - Effect of peak amplitude and desorption time of a.) 70˚C and b.) 40˚C injector temperature Figure 5 - Various flows with a.) PDMS and b.) carboxen fibers. Picture taken from Valco catalogue a.) b.) Figure 2 - Direction of flow in diaphragm valve a.) load position, b.) inject position Future Exploration • Fast, reproducible gas chromatography will be good for monitoring: • Extractions • Environmental processes • Smaller oven unit • Portable HSGC to be brought outdoors • More investigation on desorption from SPME • Analytes beyond BTEX • Nonpolar and polar analytes Figure 7 - Comparing PDMS and carboxen fibers in amplitudesand desorption time. Runs performed at 5mL/min with an injector temperature of 30˚C.

More Related