1 / 72

CIC Photonics

CIC Photonics. IRGAS Training. Schedule - Day 1. FTIR analysis IR spectrum Michelson interferometer Fast Fourier transform and corrections Interferogram, single beam spectrum, absorption spectrum and transmission spectrum Beer’s law Instrument resolution

cooper
Télécharger la présentation

CIC Photonics

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. CIC Photonics IRGAS Training

  2. Schedule - Day 1 • FTIR analysis • IR spectrum • Michelson interferometer • Fast Fourier transform and corrections • Interferogram, single beam spectrum, absorption spectrum and transmission spectrum • Beer’s law • Instrument resolution • Quantification analysis-Classical Least Squares introduction • Hardware description • Bomem WorkIR • Instrument purge • Manual manifold • SPGAS software • IRGAS 100 • IRGAS configuration manager • Data retrieval

  3. Schedule - Day 2 • SPGAS software cont. • Qmax – quantification manager • IRGAS spectra reprocessing software • Hardware installation • System power • Pipe installation • Instrument purge • Software installation • IRGAS software • Bomem Ethernet drivers • System verification • System maintenance

  4. FTIR Analysis Tab 1

  5. Light Spectrum • Infrared is invisible light ranging from 1mm to 750nm in wavelength • Infrared light can be divided into three parts: • Far infrared -1mm to 10µm • Mid infrared - 10µm to 2.5µm • Near infrared - 2.5µm to 750 nm

  6. Infrared (IR) Spectrum • IRGAS System mid infrared range • 2.5µm – 25µm in wavelength • 4000 cm-1 – 400 cm-1 in wavenumbers • Wavelength (λ) • Wavelength = (1/wavenumber)*10,000 • Wavenumber (cm-1) • Wavenumber = (1/λ)*10,000

  7. IR Molecules • Not every molecule absorbs infrared light • Monoatomic • He, Ar, Ne, etc… • Homoatomic diatomic • N2, O2, H2, etc… • N N • Molecules that do absorb infrared light • Water is a good example O H H

  8. Michelson Interferometer 2 1 3 4 • Step 1: Beam leaves IR source and hits beamsplitter where it is sent straight through and at a 90° angle • Step 2: The 90° angle beam hits a fixed mirror and is sent back to the beamsplitter • Step 3: The beam that went straight through hits a movable mirror and is sent back to beamsplitter • Step 4: The two beams recombine, go through the gas cell and travel to the detector

  9. Michelson Interferometer AB AC • When AB=AC the phase of the frequencies look the same • When AB=AC+1/4λ, then the phase of the frequencies are opposite in regards to maximums and minimums

  10. Michelson Interferometer • When AB=AC and the two recombine you get stronger maximums and minimums • When AB=AC+1/4λ and the two recombine they cancel one another out and result in a flat line

  11. ABB Bomem Michelson Interferometer • It has two sets of mirrors that move by a pivoting motion • This design is called a wishbone configuration • This configuration is more robust • It can be placed in any orientation • This configuration only has to be smooth at one point vs. the traditional interferogram that has to be smooth along a rail

  12. ABB Bomem Michelson Interferometer

  13. ABB Bomem Michelson Interferometer Wishbone Beamsplitter IR Source Laser

  14. Fast Fourier Transform ZPD • The highest peak intensity is attained when AB=AC • The maximum of the highest intensity peak is called the zero path difference (ZPD) point • After the interferogram has been created by the instrument, the Fourier transform is applied to it, which then results in a single beam spectrum

  15. Single Beam Spectrum

  16. Single Beam Spectrum

  17. Single Beam Spectrum

  18. Single Beam Spectrum

  19. Transmission • The ratio between the sample and the background spectrum

  20. Transmittance Spectrum

  21. Chemometrics • Based on the transmission spectrum chemometrics can be applied • Chemometrics: The application of statistical and mathematical methods for the design or optimization of chemical experiments and for the efficient extraction of information from chemical data • Two types of chemometrics: • Qualitative (identification) • Quantitative (quantity)

  22. Absorbance

  23. Beer’s Law • Says that concentration is directly proportional to absorbance (linearity) • Beer’s law equation is A= abC • Where A = absorbance a = absorptivity of the molecule b = pathlength that the light travels C = concentration

  24. Instrument Resolution • The more points per peak the higher the resolution • The higher the resolution the more noise, but the better peak separation • Common resolution used for an ABB Bomem instrument is 2 cm-1 • Ranges between 1 cm-1 to 128 cm-1

  25. Why Do We Need a Gas Cell • Intensity of a peak is directly related to the # of moles in a sample • In the same area: • Solid will be very packed • Liquid will be less packed • Gas will be even less packed

  26. Long Path Gas Cell Objective mirrors ¼” VCR fittings Field mirror Window retainers

  27. 4Runner 6 Meter Gas Cell

  28. Gas Cell Mirrors Top view of field mirror

  29. IR Beam Path

  30. Classical Least Squares (CLS) • The base equation is As = Ac*K + e • Where As = sample absorption Ac = calibrated absorption K = concentration e = noise • Using the above equation find K that minimizes e • To minimize e we use the CLS method • In this situation there are more equations then variables

  31. Classical Least Squares (CLS) • To simplify matters we assume that e = 0 • The equation then becomes: As = Ac*K • Matrix form:

  32. Classical Least Squares (CLS) • In order to solve for K (Ac-1*As = K), Ac needs to be an inverse matrix • Therefore: AcT*As = (AcTAc)*K (1x1) = (1x1)*K

  33. Problems with Initial CLS Approach • Baseline becomes unstable throughout the day • It can shift, slope, or curve • These changes can be compensated for in the calibrated absorption matrix

  34. Classical Least Squares (CLS) • Accounting for these baselines changes the equation • The eqn. becomes: As = AcK1+ K2+ K3+ K4 • The calibrated absorption matrix can be increased to accommodate the number of species being tested

  35. Weighted Multi-band CLS • A more complex version of the standard CLS • The spectrum is separated into bands • Each band is then calculated • After all the bands are calculated they are added in a weighted averaged fashion • The ones with the highest error and lowest signal are counted for less then the ones with the lowest error and highest signal

  36. Hardware Description Tab 2, 3, & 4

  37. ABB Bomem WorkIR

  38. Manual Manifold Purge Gas In Process Gas In Check Valve Maintenance Valve Process Gas Out Permeation Box Purifier Gas Cell Nitrogen Outflow Check Valve Pressure/Flow Transfer Optics Gas Flow Optical Path Pressure/Flow Spectrometer Nitrogen Outflow

  39. Manifold Parameters • Flow Restrictor • A flow of 30 psi in will give a flow of 5 slpm to the instrument • Purifier • Gives dry N2 to below 2 ppb • Has a lifetime of more than a year if it is used 24/7

  40. Vibrations • There are a number of designs of suspension systems to counteract vibrations • These designs help to keep the data from being affected by a simple bump of the instrument bench

  41. Typical Gas Cells • Have a flow similar to turbulent flow and have a longer residence time

  42. Laminar Flow Gas Cells • The flow is like a waterfall • Therefore there will be less turbulence • Heated laminar flow gas cells • The gas is in contact with the walls letting it reach a temperature similar to the gas cell prior to entering the cell

  43. Gas Cell Flow Diagrams

  44. f/5 Beam Geometry • The higher the f / #, the smaller the objective mirrors, the more light that is lost, the smaller the throughput

  45. SPGAS Software Tab 6, 7, 8, 9, & 10

  46. Specialty Gas Analysis Software (SPGAS) • IRGAS 100 system • Collection & quantification • Qmax • Quantification manager • IRGAS Configuration Manager • Configures parameters • Quantification Reprocessing Tool • Recalculating spectra

  47. IRGAS 100 System • After opening the software the first window is the monitor screen • On the right side there is the available species • On the bottom is the legend of the species that are being shown in the top window • The top shows the concentration of all the species over time

  48. IRGAS 100 Monitor Screen Concentration of species over time Available Species

  49. IRGAS 100 System • Collecting a background • After opening the program, pressing the start button will automatically send the spectrometer to collect a background and then begin collecting a sample • Seeing the sample • Clicking on the desired species tab at the top will show you that species in real time • That screen shows the fast concentration tracker (FCT) and the averaged sample

  50. IRGAS Configuration Manager

More Related