An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. IGRTat UCSFEPI & CBCT by Michèle Aubin MSEE

  2. IGRT Steps Prescribe Plan Set up Review • Physicians • Create and approve a prescription • Place an order for daily or weekly imaging (EPI and/or CBCT) • Dosimetrists • Prepare the reference data in Site Setup for EPI and CBCT • Send the Plan, DRRs, CT and ISO to Mosaiq • Print out the CT with ISO • Therapists • Initialize and schedule imaging • Associate a proto- col to MVCB • Align the patient • Associate the CBCT • Archive & clean the database • Physicians • Review each images and shifts in Mosaiq before next treatment

  3. Place an Order in Mosaiq Prescribe

  4. Remind the dosimetristWork done in Pinnacle and Mosaiq Plan • Type of Imaging (2D or 3D) • If 3D, inform of which contours to send for alignment • Print out the CT and ISO

  5. Assist the therapistSyngo Station Set up • Verify that Isocenter on the patient’s CT match the printout • Guide the patient’s alignment on dry run

  6. Acquire and Align RTT Syngo Station Set up

  7. Prostate Patient

  8. H&N Patient

  9. Align the PatientSyngo Station Set up • Prostate + Markers: always performed by RTT • Head&Neck: < 2 mm  no move 2 mm ≤ shift ≤ 5 mm  Move by RTT shift > 5 mm : call attending

  10. Align the PatientSyngo Station Set up Every moves (Lat, Long, Vert) are typed in Mosaiq by RTT

  11. Align the PatientRTT Syngo Station Set up • Prostate 4 MU to 6 MU • H&N 3 or 4 MU • Lung 4 to 6 MU

  12. Verify the Images and ShiftsMosaiq Review • Can verify and approve images in any Mosaiq station. • In order to charge and be legal, all IGRT have to be approved before next treatment

  13. Thank You!