An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. 150°X 200m 150° X 300m New KeraringSI-6for Keratoconus treatmentASCRS,BOSTON 2010Francisco SAnchez Leon MDInstituto Novavision Cd. Méxicowww.novavision.compacornea@yahoo.comTel +52555 3438882No financial interest to mediphacos Co.

  2. Pourpose • EvaluatethecornealremodelingusingthenewKeraring SI 6 fortheKeratocousTreatment.

  3. Corneoplastic Techniques • Mechanical methods • Intracorneal segments • Incisional procedures • Biochemical methods • Cross-linking UV light • Conductive Keratoplasty (CK) • Mix methods • Excimer laser ablation • Inlays (Acufocus)

  4. Optimized prismatic design • Scalene triangular shape SI-6 compared to Isosceles triangle in SI-5. • Improves effect and optimizes total internal reflection, reducing glare and halos

  5. 800µ base width instead of 600µ • Wider base provides 25% greater implant volume and compensates for reduction of effect in the larger optical zone

  6. When should Keraring SI-6 implantation be considered? • Mesopic pupil diameter ≥ 5mm • BCVA is ≥ 0.6 • Moderete Sphere ≤ 5 D • Cylinder value is ≤ 5 D • Segment thickness should not exceed 60% of the minimum pachymetry at the 5.5 optical ring zone • When PRK is planned as a second procedure. • When phakic IOL implantation is planned as a second procedure

  7. Different Arc Lengths for each Refraction and Topography appearance

  8. New Keraring SI-6IS-6 IntraLase 60 Hz parameters • Depth in Cornea 80% at 5.5 mm • Incision Axis at steepest meridian • Entry Cut Length 1 mm • Inner Diameter 6.0 mm • Outer Diameter 6.8 mm • Ring Energy 1.5 • Entry Cut Energy 1.3

  9. 150°X 200m 150° X 300m New Keraring SI-6150°Asymmetric Thickness Implantation -3.50/-4.00X135°=20/60 -0.50/-1.25 X 30°=20/30

  10. 120°X 250m New Keraring SI-6120°symmetric implantation +1.50 / -7.50 X 20°= 20/60 Plano/-1.25 X 45°= 20/30

  11. Single Inferior Segment for PMDCorneoplastyAchivement Preop Postop

  12. New SI-6 KeraringEarly Conclusions • Keraring SI-6 achieves Corneoplasty purposes. • Keraring SI-6 is not intended to replace the IS-5, but rather intends to offer the surgeon additional tools to customize the corneal remodeling according to each individual patient’s needs