1 / 30

Resident Physics Lectures

Ultrasound Transducer. Speakertransmits sound pulses. Microphonereceives echoes. Acts as both speaker

dunne
Télécharger la présentation

Resident Physics Lectures

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


    1. Resident Physics Lectures 01: Ultrasound Basics Principles

    2. Ultrasound Transducer Acts as both speaker & microphone Emits very short sound pulse Listens a very long time for returning echoes Can only do one at a time

    3. Piezoelectric Principle Voltage generated when certain materials are deformed by pressure Reverse also true! Some materials change dimensions when voltage applied dimensional change causes pressure change when voltage polarity reversed, so is dimensional change

    4. US Transducer Operation alternating voltage (AC) applied to piezoelectric element Causes alternating dimensional changes alternating pressure changes pressure propagates as sound wave

    5. Ultrasound Basics What does your scanner know about the sound echoes it hears?

    6. What does your scanner know about echoed sound? How loud is the echo?

    7. What does your scanner know about echoed sound? What was the time delay between sound broadcast and the echo?

    8. What else does your scanner know about sound echoes? Direction sound was emitted

    9. What else does your scanner know about echoed sound? The sounds pitch or frequency

    10. What Does Your Scanner Assume about Echoes (or how the scanner can lie to you) Sound travels at 1540 m/s everywhere in body average speed of sound in soft tissue Sound travels in straight lines in direction transmitted Sound attenuated equally by everything in body (0.5 dB/cm/MHz, soft tissue average)

    11. Luckily These Are Close Enough to Truth To Give Us Images Sound travels at 1540 m/s everywhere in body average speed of sound in soft tissue Sound travels in straight lines in direction transmitted Sound attenuated equally by everything in body (0.5 dB/cm/MHz, soft tissue average)

    12. Ultrasound Display B-scan (Brightness Mode) Image series of gray shade dots For each dot, scanner must calculate position Gray shade

    13. Images from Echos

    14. Dot Placement on Image Dot position ideally indicates source of echo scanner has no way of knowing exact location Infers location from echo

    15. Dot Placement on Image Scanner aims sound when transmitting echo assumed to originate from direction of scanners sound transmission aint necessarily so

    16. Positioning Dot Dot positioned along assumed line Position on assumed line calculated based upon speed of sound time delay between sound transmission & echo

    17. Distance of Echo from Transducer Time delay accurately measured by scanner distance = time delay X speed of sound

    18. What is the Speed of Sound? scanner assumes speed of sound is that of soft tissue 1.54 mm/msec 1540 m/sec 13 usec required for echo object 1 cm from transducer (2 cm round trip)

    19. So the scanner assumes the wrong speed? Sometimes

    20. Gray Shade of Echo Ultrasound is gray shade modality Gray shade should indicate echogeneity of object

    21. How does scanner know what gray shade to assign an echo? Based upon intensity (volume, loudness) of echo

    22. Gray Shade Loud echo = bright dot Soft echo = dim dot

    23. Complication Deep echoes are softer (lower volume) than surface echoes.

    24. Gray Shade of Echo Correction needed to compensate for sound attenuation with distance Otherwise dots close to transducer would be brighter

    25. Depth Correction

    26. Echos Gray Shade Gray Shade determined by Measured echo strength accurate Calculated attenuation

    27. Attenuation Correction scanner assumes entire body has attenuation of soft tissue actual attenuation varies widely in body

    28. Ultrasound Display One sound pulse produces one image scan line one series of gray shade dots in a line Multiple pulses two dimensional image obtained by moving direction in which sound transmitted

    29. Moving the Sound Beam electronically phased or pulsed transducer arrays

    30. Scan Patterns Linear beam translated moved sideways produces rectangular image sector beam pivoted produces pie-shaped image

More Related