An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. PET with Resistive Plate Chambers (RPC-PET) P.Fonte Contributors DFUC: Custódio Loureiro, Paulo Gordo ESTSP: Filomena Clemêncio GSI: Michael Traxler ICNAS: Ângela Cruz, Antero Abrunhosa, Francisco Oliveira, José Sereno LIP: Afonso Bernardino, Alberto Blanco, Américo Pereira, Carlos Silva, Joaquim Oliveira, Luís Lopes, Nuno Carolino, Orlando Cunha, Paulo Martins, Rui Alves LIP/DFUC: Paulo Crespo, Rui Ferreira Marques LIP/IPC: Miguel Couceiro, Paulo Fonte, U.Frankfurt: Jan Michel U.J.Kracow: Marek Palka, Grzegorz Korcyl, Marcin Kajetanowicz

  2. Outlook • 9’ repetition of the slides of Jornadas 2016 • 1’ upgrade 2016/17

  3. Positron Emission Tomography (PET) • Annihilation photons, detected in temporal coincidence (a few ns) by a set of detectors surrounding the patient, are recorded defining a Line Of Response (LOR) • Image is reconstructed from millions of acquired LORs Coincidence processing Transaxial plane Sagittal plane Positron/electron annihilation Image reconstruction

  4. S tacked RPCs e- e+ - e .......... - e - e The basic idea for RPC-based TOF-PET The converter-plate principle Use the electrode plates as a g converter, taking advantage of the natural layered construction of the RPCs. [Blanco 2002] Time resolution for 511 keV photons: (our routine lab-test tool) 90 ps s for 1 photon 300 psFWHM for the photon pair Glass Gas A previous work on PET with gaseous detectors(21 lead plates + 20 MWPCs = 7% efficiency) “The Rutherford Appleton Laboratory´s Mark I Multiwire Proportional Counter Positron Camera” J.E. Bateman et al. NIM 225 (1984) 209-231

  5. Started a long time ago… Excellent position resolution  small animal PET Affordable in large areas • full-body field of view human PET (not covered in this talk) • Increased sensitivity at physics-limited resolution

  6. N Plates Lead Glass Lead-Glass Detection Probability (511 keV Photons) Plate Thickness (m) Bateman et al. N Plates Optimum Thickness (µm) Comparison with GEANT - efficiency Optimum efficiency is balanced by beam absorption (thicker plates) and extraction probability (thinner plates) Optimum thickness depends on the number of plates and on the material. Our measurement: for few gaps ~0.2%/gap @ 511 keV [Blanco 2005]

  7. Small animal PET - a first prototype Charge-sensitive electronics allowing interstrip position interpolation Aimed at verifying the concept and show the viability of a sub-millimetric spatial resolution. 32 strips 16 plates ....... ....... 16 stacked RPCs Depth of interaction (now enjoying a second life as an exhibition item) Z 2D measurement of the photon interaction point X Transaxial

  8. Proceeding IEEE MIC (2004) M2-177 Image spatial resolution (gaussian fitting) Maximum likelihood-expectation maximization with resolution modeling (ML-EM) ~ 305 m FWHM Filtered Back Projection FBP ~ 465 m FWHM 0.46 0.31 0.47 0.48 0.45 0.30 0.32 0.29 [A.Blanco, IEEE MIC 2004, PhD thesys] Homogeneous spatial resolution over the entire field of view

  9. Full scanner for mice – project & calculations Expected quantum efficiency and resolution X We are here Almost 4 coverage

  10. Detector & readout HV distribution layer (signal-transparent) kapton insulation 5 gaps 0.35 mm 6 x 0.38mm glass ~5 mm thick Full head

  11. Preliminary resolution tests in simplified geometry but realistic readout [P.Martins JINST 2014 + PhD thesys] Two detectors with XY localization Detector a Step Motor Acceptance ±56º Detector b Needle source, 0.2 mm int. Planar (disk) source

  12. Preliminary resolution tests in simplified geometry but realistic readout [P.Martins JINST 2014 + PhD thesys]

  13. Expected performance vs. other systems [P.Martins JINST 2014 + PhD thesys] Expected performance exceeds the “target”

  14. Construction at LIP (2013/14)

  15. X’ Resolution in final geometry (2 heads only, needle source) PRELIMINARY Calibration not done in full.

  16. Installed at ICNAS (UC) in July 2014

  17. First client (31/7/2014) Poor animal resting after 18FDG injection and in the scanner tunnel (The head is central and the heart is at the edge of the field-of-view) • Maximum intensity projection. • The two hot spots in the head are likely the Harderian glands and the heart walls seem resolved

  18. Some interesting images of mice Live heart transaxial sections with 18FDG 1 mm2 Harderian glands and left striatum with 11C-raclopride

  19. Some interesting images of mice 11C Raclopride (ratinho I)

  20. Some interesting images of mice 64Cu ASTM (ratinho G) ( with strong involvement of P.Martins)

  21. Some interesting images of mice 64Cu ASTM (recent) Co-registration with MRI (Francisco Oliveira)

  22. GUI client to data processing + visualization

  23. GUI client to (custom) OSEM image reconstruction server

  24. Slow control

  25. Daily quality assurance/sensitivity check

  26. Enlarged for rats Rats A & B

  27. trigger probe  22Na charge time (trigger) charge charge-randoms time (trigger) Efficiency measurement Simulation Best chance at this configuration = 6.7% Less because of dead material (1glass+1PCB electrode/rpc) Observed: Charge = 5.12 % Trigger = 3.60 % GEANT simulation needed + better FEE

  28. Work in 2016/17

  29. Biology results start to appear Congress of the Portuguese Society of Biochemistry, 2016 Several other studies in the pipeline

  30. Upgrade 2016/17 – new DAQ + FEE (v2) HV mess will be also solved… Higher gain trigger DAQ + trigger Comms + Power management Assembled and deployed (hw+sw) by the new member of the RPC group João Saraiva Similar to the systems used for the MASTER and TOMUVOL telescopes

  31. Upgrade 2016/17 – new DAQ + FEE (v2) Custom 48 ch charge amp boards with improved isolation 90% of the analog electronics Ample room for more… Noise performance intrinsic: ~0.1 fC (600 e-) In-system: ~1fC (0.5 with expensive FETs)

  32. Upgrade 2016/17 – new DAQ + FEE (v2) TRB3 platform developed by the TRB collaboration (trb.gsi.de) Configurable trigger system (C.Loureiro + F.Clemencio) FE ADCaddon (48 ch @ 40MSPS) System: 2 TRBs, 5 ADCs, 5 qFEE boards, 240 ch

  33. charge time (trigger) charge charge-randoms time (trigger) Upgrade 2016/17 – improvement on sensitivity Efficiency measurement in v1 ~x2 @ 8000V To be expected ~(5.1/3.6)^2=2.0  satisfactory electronics V1 V2

  34. Upgrade 2016/17 – new rat-sized tunnel O velho era impresso e vertia como uma peneira

  35. Upgrade 2016/17 – new operator interface + multiuser system (60% done) Current one: shell script + matlab… Browser-server architecture Front-end: html, JS, AJAX, PHP Back-end: matlab Data protection for multiple users Backups etc

  36. Summary • More than 200 mice (and 2 rats) examined so far with 11C, 18F, 64Cu for biological research • Operated independently by a team at ICNAS • MRI co-registration via a transfer marker • Experience: detectors flawless; some problems with electronics • At TRL 7 (Technology Readiness Level 7: tested in realistic environment) 2016/17 • First biology results appeared • Sucessfull electronics and mechanics upgrade • Started user-friendly operator interface + multiuser system • PTDC submitted (M.Couceiro) Next 2 years (manpower allowing) • Submit C2020 project in human brain/organs PET • New rat-sized heads = full body for mice • Thorough calibration • Nice standard images (NaF, Raclopride, tumor) • Tests by NEMA NU2 standard • Finish operator interface + multiuser system • Completion of slow control •  TRL 8 (market readiness)