1 / 98

Rad Protection in Fluoro & INTRO RHB regs

Rad Protection in Fluoro & INTRO RHB regs. RT 244 - 2008 Week 1,2 Wed. Regulatory Requirements. 1. Regarding the operation of fluoroscopy units 2. Regarding personnel protection 3. Regarding patient protection. Fluoroscopic Positioning Previewing.

Télécharger la présentation

Rad Protection in Fluoro & INTRO RHB regs

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Rad Protection in Fluoro & INTRO RHB regs RT 244 - 2008 Week 1,2 Wed

  2. Regulatory Requirements 1. Regarding the operation of fluoroscopy units 2. Regarding personnel protection 3. Regarding patient protection

  3. Fluoroscopic Positioning Previewing • Radiographers are trained in positioning • Unnecessary radiation exposure to patient is unethical • Fluoroscopic equipment should not be used to preview patient’s position

  4. Patient Protection • Tabletop exposure rate • Maximum 10 R/min • Typically 1 – 3 R/min • Some books ave is 4 R/min **

  5. Patient Protection • Minimum source-to-skin distance • 12” for mobile equipment • 15” for stationary systems • Audible alarm at 5 mins. • Same rules for collimation

  6. Patient Protection • Typical exposure rates • Cinefluorography • 7.2 R/min • Cassettes • 30 mR/exposure • 105 mm film • 10 mR/exposure

  7. Protection of Radiographer and Radiologist • Lead apron • 0.25 mm Pb/eq • Highest energy scatter • 90o angle to the incident beam • Same level as radiographer /radiologist’s gonads

  8. Protection of Radiographer and Radiologist • Single step away from the table decreases exposure exponentially • Bucky slot cover • Lead rubber drape • Radiologist as shielding

  9. ISOEXPOSURE CURVES

  10. Protection of Others • Radiographer’s responsibility to inform others in the room to wear lead apron • Do not initiate fluoroscopy until all persons have complied

  11. PUBLIC EXPOSURE • 10 % OF OCCUPATIONAL • NON MEDICAL EXPOSURE • .5 RAD OR 500 MRAD • UNDER AGE 18 AND STUDENT • .1 rem 1 mSv

  12. COLLIMATION(for X-ray TUBE Collimator) • The PATIENT’S SKIN SURFACE • SHOULD NOT BE CLOSER THAN • ___________ CM BELOW THE COLLIMATOR? • ____________ INCHES? 15 cm / 6.5 inches

  13. The Patient & Scatter

  14. Radiation Protection (recap) • Tube in never closer to the patient than 15” in stationary tubes and 12” with a C arm • As II moves away from the patient the tube is being brought closer • Bucky tray is connected to a lead shield called the Bucky slot cover. It must be 0.25 mm Pb • There should be a protective apron of at least 0.25 mm Pb that hangs down from the II • Every machine is required to have an audible timer that signals 5 minutes of fluoroscopy time • Exposure switch must be a “dead man” type

  15. Regulations about the operation • Fluoroscopic tubes operate at currents that range from0.5 to 5 mA with 3 the most common • AEC rate controls: equipment built after 1974 with AEC shall not expose in excess of 10 R/min; equipment after 1974 without AEC shall not expose in excess of 5 R/min

  16. Other regulations • Must have a dead man switch • Must have audible 5 min. exposure timer • Must have an interlock to prevent exposure without II in place • Tube potential must be tested (monitored)weekly • Brightness/contrast must be tested annually • Beam alignment and resolution must be tested monthly • Leakage cannot exceed 100mR/hr/meter

  17. Fluoroscopy exposure rate • For radiation protection purposes the fluroscopic table top exposure rate must not exceed 10 mR/min. • The table top intensity should not exceed 2.2 R/min for each mA of current at 80 kVp

  18. Patient Protection • A 2 minute UGI results in an exposure of approximately 5 R!! • After 5 minutes of fluoro time the exposure is 10-30 R • Use of pulsed fluoro is best (means no matter how long you are on pedal there is only a short burst of radiation) • ESE must not be more than 5 rads/min

  19. Rad Protection • Always keep the II as close to the patient as possible to decrease dose • Highest patient exposure happens from the photoelectric effect (absorption) • Boost control increases tube current and tube potential above normal limits • Must have continuous audible warning • Must have continuous manual activation

  20. ESE FOR FLUORO • TLD PLACED AT SKIN ENTRACE POINT • 1 – 5 R/MINUTE AVE IS 4 R/MIN • INTERGRAL DOSE – • 100 ERGS OF TISSUE = 1 RAD EXPOSURE • OR 1 GM RAD = 100 ERGS

  21. SSD – TUBE TO SKIN DISTANCE • FIXED UNITS • 18” PREFERRED • 15 “ MINIMUM MOBILE UNITS ( C-ARMS) • 12’ MINIMUM

  22. PATIENT PROTECTION • LIMIT SIZE OF BEAM • BEAM ON TIME • DISTANCE OF SOURCE TO SKIN • PBL • FILTRATION (2.5 mm Al eq) @ 70 • SHEILDING • SCREEN/FILM COMBO

  23. GONAD SHIELDING • MUST BE . 5 MM OF LEAD • MUST BE USED WHEN GONADS WILL LIE WITHING 5 CM OF THE COLLIMATED AREA (RHB) • KUB. Lumbar Spine Pelvis • male vs female shielding

  24. Gonad shielding & dose • ♀ receive 3x more dose than • ♂ for pelvic x-rays • 1 mm lead will reduce exposure (primary) by about 50% ♀ • by about 90 – 95 % ♂

  25. KEEP I.I. CLOSE TO PATIENT

  26. Over vs under the table fluoro tubes

  27. PATIENT EXPOSURE • REDUCE DISTANCE OF IMAGE INTESIFIER • INCREASE DISTANCE FROM THE TUBE

  28. Patient entrance skin exposure (ESE) is higher when the fluoroscopic x-ray tube is too close to the tabletop.

  29. PATIENT EXPOSURE • REDUCE SIZE OF COLLIMATED BEAM WHEN POSSIBLE

  30. Framing and patient dosesyll = Pg 31 • The use of the available film area to control the image as seen from the output phosphor. • Underframing • Exact Framing, (58 % lost film surface) • Overframing,(part of image is lost) • Total overframing

  31. EXPOSURE RATES FLUORO • MA IS 0.5 MA TO 5 MA PER MIN • AVE DOSE IS 4 R / MIN • IF MACHINE OUTPUT IS 2 R/MA/MIN = WHAT IS PT DOSE AT 1.5 MA FOR 5 MIN STUDY? • 15R

  32. EXPOSURE RATES FOR FLUORO • CURRENT STANDARD • 10 R/MIN (INTENSIFIED UNITS) • HLC: BOOST MODE 20 R/MIN • OLD (1974) NO ABC NON IMAGE INTES • 5 R/MIN

  33. DOSE REGULATIONS • BEFORE 1974 - AT TABLETOP • 5R/MIN (WITHOUT AEC) • 5R/MIN (WITHOUT AEC) – BOOST MODE • After 1974 with AEC • 10 R/MIN 20R/MIN BOOST

  34. RADIATION PROTECTIONThe Patient is the largest scattering object • Lower at a 90 DEGREE ANGLE from the patient + PRIMARY BEAM AT 1 METER DISTANCE - • 1/1000 OF INTENSITY PRIMARY XRAY or 0.1%

  35. BUCKY SLOT COVER .25 MM LEAD

  36. Bucky Slot Cover

  37. ISOEXPOSURE CURVES

  38. SCATTER FROM THE PATIENT TABLE TOP, COLLIMATOR, TUBE HOUSING, BUCKY STRAY RADIATION – LEAKAGE OR SCATTER RADIATION PERSONNEL PROTECTION

  39. TOWER CURTAIN .25 MM LEAD EQ

  40. Lead curtain & dose reduction

  41. Pulsed Fluoro • Some fluoroscopic equipment is designed for pulsed-mode operation. With the pulsed mode, it can be set to produce less than the conventional 25 or 30 images per second. This reduces the exposure rate. • Collimation of the X ray beam to the smallest practical size and keeping the distance between the patient and image receptor as short as possible contribute to good exposure management.

  42. PERSONNEL PROTECTION • STANDING BEHIND A PROTECTIVE PRIMARY (1/16TH pb) BARRIER: • PRIMARY RADIATION EXPOSURE – 99.87% REDUCED • PORTABLE BARRIER = 99 % REDUCTION

  43. PERSONNEL PROTECTION • PROTECTIVE APRONS – • 0.25 PB = 97% ↓ TO SCATTER • 0.5 PB = 99.9% ↓ TO SCATTER • THYROID SHEILDS (0.25 & 0.5) • GLOVES (0.25 & 0.5)

More Related