1 / 26

A Modern Approach to Controlling ICP Spray Chamber Temperature

A Modern Approach to Controlling ICP Spray Chamber Temperature. Jerry Dulude, Glass Expansion, jdulude@geicp.com . Limitations of Externally Controlled Spray Chambers. Requires antifreeze solution Requires bulky floor model chiller Cumbersome coolant lines from chiller to chamber

jaxon
Télécharger la présentation

A Modern Approach to Controlling ICP Spray Chamber Temperature

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. A Modern Approach to Controlling ICP Spray Chamber Temperature Jerry Dulude, Glass Expansion, jdulude@geicp.com

  2. Limitations of Externally Controlled Spray Chambers • Requires antifreeze solution • Requires bulky floor model chiller • Cumbersome coolant lines from chiller to chamber • Freezing of condensate on lines • Requires intricate jacketed chamber Glass Expansion www.geicp.com 1

  3. Limitations of other Peltier Controlled Spray Chambers • Lack of programmability • Insufficient cooling • Require external fluid supply • Limited temperature range • Limited compatibility Glass Expansion www.geicp.com 1

  4. IsoMist™ Programmable Temperature Spray Chamber Glass Expansion www.geicp.com 1

  5. Features of IsoMist • Powerful Peltier-effect chiller reaches -5C in 15 minutes • Programmable from -10 to 60C in 1 degree increments • Maintains temperature to within 0.1 degree • Compact design (7.5x4x4 inches) • 100% self-contained (no external lines) • Incorporates Bluetooth® technology for clean wireless control (USB available) Glass Expansion www.geicp.com 1

  6. PC Screen showing IsoMist Software Glass Expansion www.geicp.com 1

  7. Variety of Cyclonic Spray Chambers • Borosilicate glass: most aqueous and organic applications • Quartz: Trace boron determinations • PFA: HF and ultra trace applications Glass Expansion www.geicp.com 1

  8. Applications of the IsoMist • Analyze volatile solvents – Low temp. • Reduce oxide interferences – Low temp. • Lower detection limits – High temp. • Increase stability – Constant temp. Glass Expansion www.geicp.com 1

  9. Determination of Trace Metals in Naphtha • Interfere with the cracking process • Poisoning of the catalysts • Environmental release concerns • Origin and migration markers Glass Expansion www.geicp.com 1

  10. Spike Recoveries in Light NaphthaDirect (no dilution) Glass Expansion www.geicp.com 1

  11. Oxide Reduction in ICP-MSIsobaric oxide interferences Robert Thomas, Practical Guide to ICP-MS, Marcel Dekker, NY, 2004. Glass Expansion www.geicp.com 1

  12. Reducing Oxides for ICP-MS Data courtesy of David Jones, ALS Chemex, Brisbane, Australia Glass Expansion www.geicp.com 1

  13. Analysis of Limited Sample Volumes • Biological applications • Intracellular fluids • Neonatal samples • Forensic applications • Requires micro flow nebulization • Low DL’s desirable Glass Expansion www.geicp.com 1

  14. Effect of Temperature on Normalized LOD (20ul/min Uptake) PE Optima 2100DV Glass Expansion www.geicp.com 1

  15. Constant temperature benefits • Environmental monitoring • Tighter CCV’s • Fewer QC’s • Higher productivity • Metal assays • Tighter bracketing • Higher accuracy Glass Expansion www.geicp.com 1

  16. PE ICP & ICP-MS Varian ICP & ICP-MS Agilent ICP-MS ThermoFisher ICP & ICP-MS Horiba J-Y ICP Teledyne Leeman ICP GBC ICP Spectro ICP Borosilicate glass Quartz Inert PFA IsoMist Customization Models Spray Chamber Materials Glass Expansion www.geicp.com 1

  17. Chassis Configurations of IsoMist Glass Expansion www.geicp.com 1

  18. IsoMist on Optima 2100DV Glass Expansion www.geicp.com 1

  19. IsoMist on Optima 2100DV Glass Expansion www.geicp.com 1

  20. IsoMist on the Agilent 7500 ICP-MS Glass Expansion www.geicp.com 1

  21. IsoMist on the Varian Vista MPX Glass Expansion www.geicp.com 1

  22. IsoMist on the Varian Vista MPX Glass Expansion www.geicp.com 1

  23. IsoMist on the Thermo iCAP 6500DV Glass Expansion www.geicp.com 1

  24. IsoMist on the PE Elan 9000 Glass Expansion www.geicp.com 1

  25. IsoMist on the PE Elan 9000 Glass Expansion www.geicp.com 1

  26. Summary • Control of Spray chamber temperature enhances the capability of your ICP. • The IsoMist is easily installed. • The IsoMist is easily and precisely controlled. • Compatible with all major ICP-OES & ICP-MS models. Glass Expansion www.geicp.com 1

More Related