An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Design ReviewSR BPM Upgrade Tony Pietryla December 2, 2005

  2. Prototype System Block Diagram • Receiver Interface • Buffer Position and Intensity signals • Receive control signals for receiver • Two 14-bit ADCs • Stratix II FPGA

  3. Prototype Data Acquisition Chassis Stratix II FPGA 2 Channel ADC Coldfire IOC

  4. Prototype Receiver Interface Backplane Receiver Receiver Interface Receiver Housing/Heatsink

  5. Lab Test Setup • Tests performed with: • CW signal from generator, Dx & Dy terminated on receiver, or • 100Hz tone applied to ADC inputs

  6. Lab Measurement – Calibration Data Scope Measurement: 14.35mV Measured Output: 14.74mV Scope Measurement: 2.82mV Measured Output: 2.96mV SRS DS360 Function Generator 100Hz Input directly into the ADC

  7. Lab Noise Measurement – RMS Motion 1Hz – 1kHz • CW input using filter comparator and WJ receiver • Both delta inputs on receiver were terminated • 60Hz signal approximately 85nm rms.

  8. Lab Noise Measurement – RMS Motion 200kHz – 44MHz • CW input using filter comparator and WJ receiver • Both delta inputs on receiver were terminated

  9. Lab Measurement – Long Term Drift • CW input using filter comparator and WJ receiver • Both delta inputs on receiver were terminated • Result: 0.91mm pk-pk per 0.2 °C 4.10 mm pk-pk 0.9 °C pk-pk

  10. Sector 38 (A:P3) Test Setup • The four button signals are combined then split to eliminate beam motion prior to the filter comparator

  11. Beam Based Calibration • All resolution numbers and graphs were converted to mm using a calculated value of 14.28 mm/mV • Experiments were performed to determine the calibration of the prototype unit by changing a corrector and measuring the response • Measured response of S35A:P3, S37A:P3 and prototype connected to S38A:P3 buttons • Took average of S35A:P3 and S37A:P3 responses and applied to S38A:P3 • 14-bit ADC resolution = 122.07mV/count • Result: • Horizontal = 15.3 mm/mV • Vertical = 17.5 mm/mV

  12. Single Bunch Fill Pattern – 1Hz to 1kHz • 5mA stored beam • Back-to-back splitters • 5 samples per bunch

  13. 24 Bunch Fill Pattern – 1Hz to 1kHz • 102mA stored beam • Back-to-back splitters • 4 samples per bunch

  14. 324 Bunch Fill Pattern – 1Hz to 1kHz • 102mA stored beam • Back-to-back splitters • All samples used

  15. Long Term Drift – 24 Bunch Fill Pattern • Back-to-back splitters • 4 samples per bunch • Result: 0.93mm pk-pk per 0.2 °C 4.65mm pk-pk 1 °C pk-pk

  16. Tcl/Tk Application for Manipulating Control Bits 4 samples Data portion of waveform Data portion of waveform

  17. Original BPM Specification 1 24 Bunch or greater 2 Stored Beam, Single or Multiple Bunches 3 24 hours, ± 0.2 Celsius

  18. BPM Upgrade Specification and Compliance Table 1 24 Bunch or greater 2 Stored Beam, Single or Multiple Bunches 3 24 hours, ± 0.2 Celsius