1 / 25

Research Infrastructure at the Department of Chemistry

Research Infrastructure at the Department of Chemistry. Neil Brooks. Outline. Overview, basics and examples of core infrastructure Total X-ray Fluoresence Spectroscopy – TXRF IR/Raman spectroscopy Differential Scanning Calorimetry (DSC) Nuclear Magnetic Resonance (NMR)

kuniko
Télécharger la présentation

Research Infrastructure at the Department of Chemistry

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Research Infrastructure at the Department of Chemistry Neil Brooks

  2. Outline • Overview, basics and examples of core infrastructure • Total X-ray Fluoresence Spectroscopy – TXRF • IR/Raman spectroscopy • Differential Scanning Calorimetry (DSC) • Nuclear Magnetic Resonance (NMR) • Single Crystal X-ray Crystallography (SC-XRD) • Summary • Assorted other equipment

  3. TXRF – Total X-ray Fluoresence Spectroscopy

  4. TXRF: key parameters • Two Bruker S2 Picofox TXRF spectrometers in the group • Key information: accurate elemental compositions/ratios • Fast and easy • Low detection limit possible (e.g. ppb) • Multi-element sample analysis • Difficult to detect elements lighter than chlorine • Ejected X-rays from lighter elements low in energy (matrix effects)

  5. TXRF: technique

  6. TXRF: technique

  7. TXRF example: [Br] and [Cl] in ionic liquids

  8. IR/Raman Spectroscopy

  9. IR: technique • FT-IR spectrometer Bruker Vertex 70 with Raman module Bruker RAM II • Vibrational spectroscopy • Key information: identification of particular bond vibrations

  10. IR: technique • FT-IR spectrometer Bruker Vertex 70 with Raman module Bruker RAM II • Vibrational spectroscopy • Key information: identification of particular bond vibrations • Spectral range 4500 to 400 cm-1 • Transmittance mode (sample prep necessary) • Attenuated total reflection • Diamond or ZnSe crystal • Fast (ca. 1-2 minutes) • Far infrared (400 to 40 cm-1) also possible • Variable resolution (e.g. 1, 2, 4 cm-1) • Required sample: <5 mg

  11. Raman: technique • FT-IR spectrometer Bruker Vertex 70 with Raman module Bruker RAM II • Different selection rules to IR • Nd-YAG laser λ = 1064 nm • Variable laser power (1-500 mW) • Fluoresence can be problem • Spectral range 3600 to 40 cm-1 • Variable resolution (e.g. 1, 2, 4 cm-1) • Sample dependent collection time (typically 2 minutes – 2 hours) • No background collection • Liq. N2 cooled Ge CCD detector • Required sample: 50-500 mg • Surface Enhancement possible (SERS)

  12. Raman example: adsorption of 1H-benzotriazole on silver substrate in [Ag(MeCN)4]2[Ag(Tf2N)3] • text Pure Ag-LMS 1H-benzotriazole (0.1 M) 0.1M benzotriazole in Ag-LMS on Ag surface 0.1M benzotriazole in Ag-LMS (bulk) Pure benzotriazole Pure Ag-LMS

  13. DSC – Differential Scanning Calorimetry

  14. DSC: technique • Mettler-Toledo DSC822e module • Key information: phase transition temperatures/energetics • Required sample: 2-5 mg • Temperature range -60 to 250 °C • Variable heating rate (typically 10 °C/min) • Complimented with polarised optical microscopy (thermomicroscopy) on an Olympus BX-60 polarising microscope equipped with a Linkam THMS 600 hot stage. Temperature range: -196°C - 600°C.

  15. DSC example: eutectic behaviour of DMSO2/acetamide mixtures DSC traces of DMSO2/acetamide mixtures Tamman plot

  16. NMR – Nuclear Magnetic Resonance Spectroscopy

  17. NMR: technique • Energy states of spin active nuclei split from degeneracy inside a magnetic field • E/M radation applied to populate the higher energy state • Difference in energy of higher and lower states measured is a function of the nucleus and its environment • Key information: chemical information about nucleus environment • 300, 400 and 600 MHz spectrometers available • Nucleus must be NMR active • Most common nuclei: 1H, 13C, 31P, 19F • Can not have unpaired electrons • Must have sufficient natural abundance • Liquid state: sample must be a liquid or disolved in a solvent • Required sample: 5-20 mg

  18. Single Crystal X-ray Crystallography (SC-XRD)

  19. SC-XRD: technique • Key information: absolute three-dimensional crystal structure • Allows the resolution of atom positions up to ±0.001 Å • Accurate bond and intermolecular distances • Scattering (diffraction) of X-rays from ordered array of molecules in a crystal leads to diffraction pattern (Bragg’s law: nλ=2dsinθ) • Diffraction pattern is directly related to the three-dimensional electron density pattern • Phase problem • Imperfect crystals

  20. SC-XRD: practicalities • Must have (good quality) single crystal! • Crystal size (each dimension) must be 0.05 to 0.5 mm • Larger crystals can be cut to size • Smaller crystals may be possible but will take longer • Crystal growth • Evaporation of saturated solution • Addition of antisolvent • Growth from melt by slow cooling • What can be determined? • Can be used for phase identification • Crystal structure already known (unit cell check) • New structure determination • Elemental composition (on specific lattice sites)

  21. SC-XRD example: structure of Ag-LMS • First synthesised a Cu-LMS of formula [Cu(MeCN)4][Tf2N] • Melting point 65 °C • New Ag-LMS of formula [Ag(MeCN)~2][Tf2N] • Melting point 18 °C • Slow cooling of Ag-LMS gave good quality crystals • Structure determined as [Ag(MeCN)4]2[Ag(Tf2N)3]

  22. SC-XRD: case study structure of Ag-LMS • If [Ag(MeCN)4]2[Ag(Tf2N)3] is heated at 50 °C for a period of time new crystals appear • Crystal structure analysis shows the new compound is [Ag(MeCN)Tf2N] • One-dimensional polymeric structure • Melting point 90 °C

  23. Overview

  24. Assorted other equipment • Mass spectrometry • Absorption spectroscopy (UV-VIS-IR) • Varian Cary 5000: UV-VIS-NIR spectrophotometer (175-3300 nm) • Luminescence spectrometers • Viscosimetry • Brookfield cone plate viscosimeter (LVDV-II+ Programmable Viscometer) with cone spindle CPE-40 • X-ray powder diffraction setup with rotating Mo-anode • SAXS and WAXS • CHN microanalysis • CE Instruments EA-1110 CHN elemental analyser

  25. Practicalities • Contact me by email: neil.brooks@chem.kuleuven.be to notify interest • Set up meeting to discuss how to proceed • Tom Vander Hoogerstraete (Chemistry) • M. Ganapathi (MTM) Acknowledgements