An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Progress in water vapour spectroscopy at RAL Dr Robert McPheat CAVIAR Annual Meeting Imperial College, 16/12/2008

  2. Measurements so far in the mid- and near-IR. These measurements are intended to target the weak water vapour continuum. Long path lengths and / or high vapour pressures. Measurements carried out at moderate (0.01cm-1) & high resolution (0.002cm-1) using the Bruker IFS125 HR. Initial measurements revealed a need for better line data too. MSF Measurements

  3. Long path lengths and / or high vapour pressures. Use Long pathlength absorption cell (LPAC) / Short pathlength absorption cell (SPAC). LPAC 32 to 640 metres in IR, over 1 km in visible Multi-pass reflective optics (White) Room temperature to 190K 0 to 1 bar pressure SPAC 1.2 to 20 metres Multi-pass reflective optics (White) Existing temperature range 200 to 350K 0 to 5 bar pressure LPAC & SPAC

  4. Measurement conditions Room temperature, ~ 290 K Water pressures up to 0.75 SVP (< 20 hPa) Resolution up to 0.002cm-1 but mainly at 0.01cm-1 Water/air mixtures at total pressures of 150, 500 & 1000 hPa Pathlengths of 32, 128 & 512m Made measurements in 3 spectral ranges covering mid/near-IR 650 to 2000 cm-1 MCT detector / KBr beamsplitter / Globar 1800 to 5500 cm-1 InSb detector / CaF2 beamsplitter / Globar 2000 to 11000 cm-1 InSb detector / CaF2 beamsplitter / Quartz tungsten halogen source LPAC Measurements

  5. Example LPAC Spectra 17hPa H2O + air to 1000hPa

  6. Water vapour partial pressure limited by vapour pressure at room temperature Increase temperature to observe weak lines & continuum Desired upper temperature for water vapour measurements 450K All parts of the cell & filling lines need to be at same temperature to avoid condensation in cold spots. Need 200ºC, 5 bar pressure gauge Other measurements using SPAC require lower temperatures therefore SPAC redesigned for -200 to 200ºC temperature range Other requirements 5 bar maximum pressure Mirrors adjustable from outside Accurate temperature & humidity measurement Corrosion resistant SPAC redesign

  7. Insulating Vacuum Helium Electrical heaters Liquid Nitrogen Sample Gas & White Optics

  8. SPAC upgrade Optics Inner Cooling Jacket

  9. High temperature pressure gauge & filling line Temperature & pressure logging Temperature control Outer vacuum jacket Detector box Transfer optics vacuum tank

  10. SPAC upgrade modifications • Additional base heaters added • Problems with thermal expansion • Can be used for hot measurements without cooling jacket but modifications required

  11. SPAC upgrade problems • Heating capacity limited by rating of spectrometer room ring main. • Use several extension leads from elsewhere • Original window seals (Indium) melted. • Initially replaced with gold wire which failed on temperature cycling. • PTFE seals have been installed successfully. • Long delays getting a working pressure gauge suitable for 5 bar & 200 °C • Ready to begin measurements beginning of new year

  12. Water line parameter measurements • Although CAVIAR is targeted at improving the water vapour continuum, errors in line parameters contribute to the measured continuum • First measurements with upgraded Bruker IFS120 HR • Upgraded electronics / moving mirror to same specification as IFS125 • High resolution measurements 0.002 cm-1 • Long integration times (~ 12 hours) in mid-IR

  13. Water line parameter measurements • 5 cm cell to avoid saturation of strong lines • Required sample pressures <0.5 hPa • Absorption of water on cell walls causes difficulty in maintaining stable sample pressures • Large variation of sample pressure with room temperature • Filling line all stainless steel and as short as possible • Measurements completed 1000 to 2000 cm-1, 2000 cm-1 to 9000 cm-1 next.

  14. Example Spectra Transmission Wavenumber /cm-1