1 / 6

Heat-Flux Measurement - Issues

Heat-Flux Measurement - Issues. Convection heat transfer effects (Schmidt-Boelter & Gardon) Non-uniform temperature distribution on sensor (Gardon) Different calibration methods by manufacturers Lack of suitable traceability to National Standards. Heat-Flux Measurement - Status.

liluye
Télécharger la présentation

Heat-Flux Measurement - Issues

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Heat-Flux Measurement - Issues • Convection heat transfer effects (Schmidt-Boelter & Gardon) • Non-uniform temperature distribution on sensor (Gardon) • Different calibration methods by manufacturers • Lack of suitable traceability to National Standards

  2. Heat-Flux Measurement - Status • No major break-through in sensor manufacture • Calibration issues still persist (Sarkos, Hill & • Johnson 1995)

  3. Calibration Approach - NIST Blackbody radiation Detector Based (Flux) Source Based (Temperature) Flux measured Absorbs incident flux Flux traceability to National Standards Flux derived Effective emissivity Configuration factor Source temperature Establish equivalence Factors: Calibration range, source temperature, view angle

  4. Resolving Calibration Issues National Level • Common Interest Group – Government/Users/ Manf. • Evaluation of different calibration techniques • Standard development efforts – ASTM/ASME • Provide traceability to Primary Standards

  5. Resolving Calibration Issues International Level • Fire Laboratories Round-robin [NIST-BFRL/PL] • International Standards Organization • ISO Flux Meter calibration standard Development • Ellipsoidal radiometer evaluation [SP, Sweden]

  6. Major Outcome of Collaboration • Flux-scale calibration suitable when • reflected component is large part of incident flux, • source characteristics difficult to establish • Temperature & flux scales equivalence demonstrated • Low/moderate heat flux - convection effects important • High heat flux – source effective emissivity important