1 / 6

The NCNR Spin-Polarized Triple-Axis Spectrometer (SPINS)

The NCNR Spin-Polarized Triple-Axis Spectrometer (SPINS). The SPINS instrument is located at the heart of the cold neutron guide hall of the NIST Center for Neutron Research. SPINS. The NCNR Spin-Polarized Triple-Axis Spectrometer (SPINS). Incident Energy 14 ~ 2 meV ( 2.4 ~ 6.1 Å )

lkarl
Télécharger la présentation

The NCNR Spin-Polarized Triple-Axis Spectrometer (SPINS)

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The NCNR Spin-Polarized Triple-Axis Spectrometer (SPINS) The SPINS instrument is located at the heart of the cold neutron guide hall of the NIST Center for Neutron Research. SPINS

  2. The NCNR Spin-Polarized Triple-Axis Spectrometer (SPINS) • Incident Energy • 14 ~ 2 meV ( 2.4 ~ 6.1 Å ) • Energy Resolution • 0.02 ~ 1 meV • Special Equipments • Supermirror Transmission Polarizer • A stack of single crystal Si plates with Fe/Si supermirror coatings • Multi-crystal Analyzer • 11 Horizontally Focusing PG Blades • Position Sensitive Detector • 3He, 20 × 25 cm2 • Filters • PG or Be filters before sample • Be or BeO filters after sample The SPINS instrument with a super-conducting magnet.

  3. Sample Single Detector Neutron Source Analyzer Monochromator Improvement Multicrystal analyzerandposition-sensitive detector The SPINS Operation Modes I:Conventional Triple-Axis Spectroscopy (TAS) A single point at a time TAS is ideally suited for probing small regions of phase space Shortcoming: Low data collection rate

  4. Sample D2q Monochromator Single Detector L = distance from sample to HF analyzer wa = total width of HF analyzer D2q = wa sinqa/L ~ 9 degree for Ef=5 meV at SPINS Multicrystal Analyzer The SPINS Operation Modes II:Horizontally Focusing (HF) Analyzer Mode Relaxed Q-resolution Useful for studying systems with short-range correlations

  5. Sample Position-Sensitive Detector D2qi hw Flat Analyzer qai qa Q The SPINS Operation Modes III:Multiplexing Detection System for TAS ~1meV qai = qa + D2qi = qa - atan(x sinqa/(L+xcosqa)) kfi = ta/2sinqai Qi = ki - kfi Probes scattering events at different energy and momentum transfers simultaneously Survey (hw-Q) space by changing the incident energy and scattering angle

  6. PSD Low-pass Filter Sample w Be or BeO L Ei S(Q,w) dw The PSD measures Ei-Ecutoff The SPINS Operation Modes IV:Position-Sensitive Detector in Two-Axis Mode The filter passes only those neutrons with 0 < Ef < Ecutoff Large angular acceptance = w / L ~ 11o for SPINS