An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. High resolution scintillation detectors for DBD Jean-stephane Ricol CENBG – CNRS - France ILIAS July 1-5, 2007 Blaubeuren

  2. Outline • SuperNemo R&D : very promising results using liquid scintillator : • Motivations for using LS • Measurements made in CENBG – Bordeaux • Measurements made in INR - Kiev • Prospects and improvements : • Large size detector • High QE PMTs • Optimization of surface for light collection

  3. Motivation and goal

  4. Solid / Liquid scintillator • Same light yield • Different attenuation length Liquid : Latt > 10-25 m Expect high resolution even for large size detector ~ 20cm x 20 cm x 20 cm Solid : Latt ~ 3 m h = 2 cm FWHM = 7-8% h = 10 cm FWHM = 9-10% My talk is focused on LS results

  5. Measurements in Bordeaux Choice of LS : LAB • Linear AlkylBenzene • High light yield (50% ant.) • Good transparency (Latt > 10 m) • High flash point (130°C) • Very low toxicity (literature data) Few measurements with liquid Bicron BC 531 (light yield 59% ant., latt = 3.5 m) for comparison

  6. LAB emission spectra Made and measured in Dubna Need absorption spectrum measurement

  7. Teflon Delrin 10 cm LS Teflon 5 cm LS Medium-size container PMT = Photonis 3” XP5312b QE ~ 33% LS active height = 5 or 10 cm (from bottom to PMT photocathode) 7 cm PMT is in the LS = direct light transmission

  8. Simulation Medium-size container • Entry face in Kapton : • Very good breaking strength • with 25 µm thickness 60 times lower than breaking point • Very good chemical strength LS doesn’t damage it • After 6 months Kapton foil didn’t sweat any LS leak • Kapton is yellow and transparent -> need reflector : Al deposit

  9. Aluminium deposit on Kapton Kapton foil Reference quartz Al sample in a smelter heated by 30 A currant (Tevaporation = 800 °C) Evaporation projects Al in 4pi, it stops when quartz measures ~ 100 nm deposit We obtain a good mirror All is inside a vacuum vessel (10-6 atm)

  10. The test bench CENBG test bench Black box : LAB container + PMT (3” XP5312b Photonis) Electron source Spectrometer 90Sr source, 370 MBq E/E ~ 1% Range 0.4 – 2 MeV XY table DAQ electronics

  11. Results : light emission and propagation BC-531 : light yield 59% ant., Latt = 3.5 m (data sheet) Dubna LS = LAB : light yield ~50% ant., Latt > 10 m (SNO+ values) LAB LS is much better than BC531 => Attenuation is more important than light yield Must compare emission+attenuation (l) measurements

  12. Results : light collection LAB LS • Delrin vs Teflon : same values within the errors • Diffusing medium is better than reflecting medium on sides for this geometry (7 cm diameter, 5 cm high)

  13. FWHM @ 1 MeV (%) Beam position (cm) Results : position effect No effect on resolution, light is emitted and collected independently of the interaction position

  14. Results : size effect No effect as expected from large Latt • Comparison with plastic scintillators (same PMT as LS) • PS1 = 7.5 cm x 2 cm : 8.0% • PS2 = 7.5 cm x 10 cm : 9.8% Strong attenuation with solid  liquid is a much better candidate for large size detector

  15. Prospects in Bordeaux • First of all we must prove the same promising results can be reached with large size container + 8” PMT • New container in construction, we will first test Kapton resistance (simulation of mechanical strengths claim it should be ok) • O2 purge by N2 bubbling should increase the detected number of photons by 10-20%

  16. Measurements in Kiev Similar container 84×92 mm Different LS : Toluene + pTP 5 g/l + POPOP 0.1 g/l PETF film 0.5 mg/cm3 Add AlMyl teflon Pipe for O2 purging (pure argon) Similar QE PMT (Photonis 3” XP2412 QE ~ 33%) from A. Danevich et al

  17. 207Bi Results FWHM = 7.4% @ 976 keV from A. Danevich et al • Resolution better than with LAB (7.4 vs 8.2 %) : • LS light yield or Latt higher ? need spectra measurement • O2 purge ? 15% more light collection : 8.2  7.6 %

  18. Large container Mylar film 1.7 mg/cm2 23 cm Teflon container:volume 6.3 dm3 window area 3.5 dm2 5 ” PMT EMI 9390 Unfortunately the PMT is not of so good quality QE≈24% A.A.Adadurov et al.

  19. 207Bi BG-subtractedspectrum Teflon tape 5.6 mg/cm2 +Aluminized Mylar 0.6 mg/cm2 207Bi Pileup of signals FWHM = 11 % (for 33% QE : FWHM  9.6%) Results for large container Resolution is not good (8% expected) Bad PMT ? High counting rate (18 kHz) ? A.A.Adadurov et al

  20. Uniformity of light collection ●100(1)% ●100(1)% ●100(1)% ●100(1)% ●100(1)% Uniformity tested by moving the 207Bi source by 2cm step No difference was observed within 1% accuracy A.A.Adadurov et al

  21. Prospects in Kiev • PMT with QE≈30% • Measurements in a shield to suppress BG counting rate • Lower activity 207Bi source

  22. Light collection simulation Very promising results in both labs but still need to be improved : light simulation can help us to optimize photon collection and improve the resolution Geant4 : detector + particle interaction + light simulation Need detailed spectra : preliminary results simulated with PS data Simulation of PS block F20cmx20cm

  23. Plot from E. Chauveau Refl. Diff. Refl % Refl. Refl. Diff. Diff. Light collection simulation Effect of light collection on side : reflection vs diffusion There is an optimization of the surface !! We can win 0.3-0.5% on the resolution Must try several layers and check it experimentally 2 layers

  24. Super super PMTs New PMTs : super super bialkali Photocathode CENBG test Photonis PMTs and UCL test Hamamatsu PMTs QE : 32% -> 42 % => resolution : 8 (7) % -> 7 (6) % we are very optimistic for the 7% goal !!!

  25. Conclusions • Very good results obtained from 2 teams on medium size container : LS is very promising • First results on large size container showed unexpected bad resolution,incoming new measurements (in both Kiev and Bordeaux) • Very good uniformity of light emission/collection measured in both labs • Experimental and simulation work on the optimization of light collection on sides : geometry, diffusion/reflection … in order to improve resolution • High QE PMTs will do a large part of the work • We are thinking about the container design for SuperNemo