1 / 44

Mass Spectrometry

Mass Spectrometry. Ionization Techniques. Mass Spectrometer. All Instruments Have: Sample Inlet Ion Source Mass Analyzer Detector Data System. http://www.asms.org. Ionization Techniques. Gas-Phase Methods Electron Ionization (EI) Chemical Ionization (CI). Desorption Methods

maalik
Télécharger la présentation

Mass Spectrometry

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Mass Spectrometry Ionization Techniques

  2. Mass Spectrometer • All Instruments Have: • Sample Inlet • Ion Source • Mass Analyzer • Detector • Data System http://www.asms.org

  3. Ionization Techniques Gas-Phase Methods • Electron Ionization (EI) • Chemical Ionization (CI) • Desorption Methods • Matrix-Assisted Laser Desorption Ionization (MALDI) • Fast Atom Bombardment (FAB) • Secondary Ion MS (SIMS) • Spray Methods • Electrospray (ESI) • Atmospheric Pressure Chemical Ionization (APCI)

  4. Electron Ionization http://www.noble.org/PlantBio/MS/ion_tech_main.html

  5. Electron Ionization • Samples must be vaporized in the ion source • Typically 1 of 1000 molecules entering the source is ionized • 10-20 eV of energy is imparted to the molecule • ~10eV is enough to ionize most molecules • Up to 230 kcal/mol is left to cause fragmentation

  6. Electron Ionizationb-lactam: Effects of ionization energy

  7. Electron Ionization (low picomole) Advantages • Well-Established • Fragmentation Libraries • No Supression • Insoluble Samples • Interface to GC • Non-Polar Samples • Disadvantages • Parent Identification • Need Volatile Sample • Need Thermal Stability • No Interface to LC • Low Mass Compounds (<1000 amu) • Solids Probe Requires Skilled Operator

  8. Chemical Ionization • Reagent gas is introduced into the source at ~0.5 torr • Reagent gas is preferentially ionized. Ions react mostly with neutral reagent gas • Reactions occurring depend on the nature of the reagent gas • Ions in the reagent gas plasma react with the analyte

  9. Chemical Ionization http://www.noble.org/PlantBio/MS/ion_tech_main.html

  10. Chemical Ionization: Methane • Methane primarily forms CH4+• with CH2+• and CH3+ • CH4+• + CH4→ CH5+ + CH3 (m/z 17) • CH2+• + CH4→ C2H3+ + H2 + H• • C2H3+ + CH4→ C3H5+ + H2 (m/z 41) • CH3+ + CH4→ C2H5+ + H2 (m/z 29)

  11. Chemical Ionization: Methane

  12. Chemical Ionization: Methane • Ions other than saturated hydrocarbons react via proton transfer • CH5+ + M → MH+ + CH4 (or via C2H5+ or C3H5+) • For saturated hydrocarbons, hydride abstractions is common • CH5+ + RH → R+ + CH4 + H2 • For polar molecules, adducts can form • CH3+ + M → (M+CH3)+ • MH+, R+, and adducts are pseudomolecular ions.

  13. Chemical Ionization: Isobutane

  14. Chemical Ionization: Isobutane

  15. Chemical Ionization: Isobutane • Reacts through Proton Transfer • C4H9+ + M → MH+ + C4H8 • For saturated hydrocarbons, no reaction • For polar molecules, adducts can form • C4H9+ + M → (M+C4H9)+ • Lack of reaction with hydrocarbons can be used for selective detection of compounds in mixtures containing hydrocarbons • Less fragmentation is observed with isobutane. (molecular species is more reliably formed)

  16. EI vs. Methane vs. Isobutane EI Methane CI Isobutane CI

  17. Chemical Ionization: Negative Ions • Low energy electrons are present in the CI plasma • These can attach to molecules with high electron affinities • There are two principal pathways: N2O/CH4 reagent gas • AB + e-→ AB-• (associative resonance capture) • AB + e-→ A• + B- (dissociative resonance capture) • Deprotonation can also occur if a basic ion is formed in the reagent gas plasma

  18. Chemical Ionization (low picomole) Advantages • Molecular Ion • Interface to GC • Insoluble Samples • Disadvantages • No Fragment Library • Need Volatile Sample • Need Thermal Stability • Quantitation Difficult • Low Mass Compounds (<1000 amu) • Solids Probe Requires Skilled Operator

  19. Ionization Sources - II • EI and CI have limitations • Both require a volatile sample • Samples must be thermally stable • Neither lends itself to LC/MS analysis • Other techniques have been developed • FAB (Fast Atom Bombardment) • SIMS (Secondary Ion MS) • MALDI (Matrix Assisted Laser Desorption) • ESI (Electrospray)

  20. FAB • Sample is dissolved in a non-volatile liquid matrix • Glycerol and m-Nitrobenzyl alcohol are common matrices • A high energy (5kV) beam of neutral atoms (typically Ar or Xe) is focused onto the sample droplet • Dissolved Ions and Molecules are ejected into the gas phase for analysis

  21. FAB

  22. FAB • For Organic Molecules M+H and M+Na ions are typically observed • M+H ions typically fragment more than M+Na ions • Salts such as NaI can be added to the matrix to induce M+Na formation

  23. FAB (nanomole) Advantages • Stable Molecular Ion • High Mass Compounds (10,000 amu) • Thermally Labile Compounds (R.T.) • Disadvantages • No Fragment Library • Solubility in Matrix (MNBA, Glycerol) • Quantitation Difficult • Needs Highly Skilled Operator • Not amenable to automation • Relatively Low Sensitivity

  24. SIMS • Analysis of surfaces in situ • A high energy (15-40 keV) beam of primary ions (In+, Ga+) or clusters (SF6, Au3, C60) is focused onto the sample droplet • Surface and slightly sub-surface atoms or molecules are ejected and ionized • Clusters increase secondary ion yield dramatically

  25. SIMS Courtesy of Mike Kurczy, Winograd Group, Penn State University

  26. SIMSGa+ vs. C60+ J. Phys. Chem. B 2004, 108, 7831-7838 Videos

  27. SIMS Advantages • Surface analyses • Stable Molecular Ion • Mass Limit ~ 10000 amu • Small primary ion beam spots and beam rastering make imaging possible • Disadvantages • No Fragment Library • Quantitation Difficult • Needs Highly Skilled Operator • Not amenable to automation • Relatively Low Sensitivity

  28. MALDIMatrix Assisted Laser Desorption • Sample dissolved in a solid matrix • Typically mixed in solution • Small droplet applied to target and dried • A wide variety of matrices exist • Choose based on hydrophobic/hydrophilic character of sample • Also based on laser absorbance (usually UV) • An ionization agent is often added • Agent must bind to the sample • TFA and its Na+ Ag+ salts are common

  29. MALDI

  30. MALDI • Choice of matrix based on empirical evidence • http://polymers.msel.nist.gov/maldirecipes/maldi.html • Typically singly charged ions observed • Some matrix adducts/cluster ions • Difficult to analyze low MW compounds due to matrix background • Typically used for MW 500-500,000

  31. MALDI

  32. MALDI

  33. UV-MALDI Matrices

  34. MALDI (low femtomole) Advantages • Parent Ion • High Mass Compounds (>100,000 amu) • Thermally Labile Compounds (R.T.) • Easy to Operate • Easily Automated • Disadvantages • No Fragment Library • Wide variety of matrices • Quantitation Difficult • Matrix Background

  35. ESIElectrospray Ionization • Sample dissolved in a polar solvent • Solution flows into a strong electric field (3-6 kV potential) • Electric field induces a spray of highly charged droplets (charges at surface) • As droplets shrink, repulsion increases until they break into smaller droplets • In small enough droplets, surface charges can be desorbed into the gas phase.

  36. ESI

  37. ESI • Ions formed via charge-residue or ion-evaporation • Molecules form M+H+ or M-H- ions • Large molecules: 1 charge / 1000 amu • Small molecules: Usually singly charged • Molecules with no acid/base groups • Can form adduct ions with Na+ K+ NH4+ Cl- OAc-, etc. • Salts may be added or already present in sample.

  38. ESI • ESI ions formed at high pressure must be transferred into high vacuum • Differential pumping is needed to move ions through small openings while maintaining low pressures • Ions become super-cooled by expansion. Solvent can recondense • Two methods to reduce cluster formation • High temperature transfer tube • Heated counter-current flow of N2

  39. ESI

  40. ESI-Multiply Charged Ions • Large Molecules produce an envelope of charge states • Deconvolution must be done to determine the charge states if isotopic resolution is not possible • Typically, MS data systems use software to deconvolute automatically

  41. ESI-Multiply Charged Ions M=16953 Δm = 1 amu ; ∆(m/z) ≈ 0.055; z = 18 ; Δ(m/z) ≈ 0.10 Δm = 1 amu z = 10

  42. j(m2-mp) z1 = M = z1(m1-mp) (m2-m1) ESI-Multiply Charged Ions • Consider (M+zH)z+ • z1m1 = M + z1mp (m1 = measured m/z) • Consider a peak of m/z=m2 which is (j-1) charge states away from peak m1 • m2(z1-j) = M + (z1-j)mp

  43. 10(1621.3-1.0073) j(m2-mp) = 51.0 z1 = z1 = (m2-m1) (1621.3-1303.8) ESI-Multiply Charged Ions j=10 1303.8 1621.3 M = z1(m1-mp) M = 51.0(1303.8-1.0073) M = 66485

  44. ESI (low femtomole to zeptomole) Advantages • Parent Ion • High Mass Compounds (>100,000 amu) • Thermally Labile Compounds (<0º C) • Easy to Operate • Interface to HPLC • Zeptomole sensitivity with nanospray • Disadvantages • No Fragmentation • Need Polar Sample • Need Solubility in Polar Solvent (MeOH, ACN, H2O, Acetone are best) • Sensitive to Salts

More Related