An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. DR. KUMAR J DOCTORTitle: Intra Operative Online Pachymetry in LASIK with Wavefront-Guided Ablation Author: Dr. KUMAR J. DOCTOR  AIOS No: D03945  E-Poster Serial No.      : 230 DOCTOR EYE INSTITUTE MUMBAI NO FINANCIAL INTEREST

  2. Purpose: To investigate Intraoperative Optical Coherence Pachymetry during various steps of LASIK while treating patient with Wavefront Guided Ablation

  3. Online OCPOptical Coherence Pachymeter ESIRIS SCHWIND LASER mirror s(t) light source beam splitter eye detector OCP-signal from Michelson Interferometer

  4. It comes with fiberglass coupling and through Automated measurements of changes in corneal thickness, allows continuous and real time values with 10 measurements per second, without having to interrupt the treatment. The direct integration of Online Pachymetry into the ESIRIS The reproducibility of current System is 1 to 2 microns.

  5. 25 patients IN Aug. 09 Myopic Eyes Corneal Thickness Assessed During Various Steps CP Microkeratome 110 micron head – 50 eyes One Blade Used Per Eye Wave Front Guided Treatments done Method

  6. Wavefront-Guided Ablation Every 3 micron distortion on the Cornea creates 1 micron of Wavefront Error “Corneal Wavefront is the component of eye refraction owed to the Cornea (or OPTICAL PATH DIFFERENCE OPD) Aberrations: “Ocular” = “Corneal” + “Internal”(or “Total”) From WAFromTC (100%)(80%)

  7. Results • Mean Spherical Equivalent - 4.0 D,+/-3.0 D • Mean Stromal Depth 80+/-22 microns • Mean Flap Thickness of 110, +/- 9 microns • Residual Stromal 300 microns & above in all eyes. • All optical zones 6.00 mm

  8. Table 1 Central corneal thickness measurements in eyes PRE-OP laser in situ keratomileusis Table 2 Central corneal thickness measurements in eyes POST-OP laser in situ keratomileusis

  9. Advantages of Online Pachymetry. The greatest advantage of Online Pachymetry is indeed the improved quality assurance, objective monitoring of the complete LASIK procedure for the first time. The special feature of Online Pachy provides • Extensive Diagnosis – Each step of the surgery can be checked during operation • Contact Free measurement of Corneal Thickness online, hence no chances of infection • Also the individual ablation depth and possible dehydration effects of the cornea were monitored continuously • Improvement in Intraoperative Safety and Quality of LASIK

  10. Conclusion • Evaluation of residual corneal thickness is crucial to avoid the condition of excessive corneal thinning (ectasia) • Important safety feature to monitor the thickness of each cut • Assessment of flap and residual stromal thickness. • Accurate Measurement of ablation depth during laser treatment. • TO CHECK A WELL OPPOSED INTERFACE AFTER WASH