Guide to Operating and Maintaining Suction Machines in Medical Settings

A suction machine, essential for airway clearance in patients unable to do so themselves, requires proper operation and maintenance. Steps include plugging in, setting pressure levels, and regular cleaning. Common failures such as leaks and motor issues should be addressed promptly to ensure optimal performance.

ouachikh
Télécharger la présentation

Guide to Operating and Maintaining Suction Machines in Medical Settings

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Suction Machine

  2. Introduction • A suction machine, also known as an aspirator, is a type of medical device that is primarily used for removing obstructions like mucus, saliva, blood, or secretions from a patient airway. When an individual is unable to clear secretions due to a lack of consciousness or an ongoing medical procedure, suction machines help them breathe by maintaining a clear airway. • The types suction devices : • Manual suction devices. • Stationary suction machines • Portable suction machines Manual suction devices. Portable suction machines

  3. Operating the Suction Machine 1. Plug the suction machine into a grounded outlet. 2. Check that the tubing from the machine to the collection jar is on and snug. 3. Check that the lid to the collection jar is closed tightly. 4. Attach the extension tubing to the collection jar. 5. Turn the machine on and kink the extension tubing to block the flow of air.  If the pressure gauge did not move when kinking the tubing, recheck all of your connections. Look for leaks in the system. The lid may not be closed tightly; a tub may not be on properly or the tube in punctured. 6. Look at the pressure gauge. Using the control dial set the gauge pressure between 15" and 20" of Hg. (for an adult)  An infant or a child will use less pressure. 7.After the pressure is set, connect the oral suction device or suction catheter to the suction extension tubing.

  4. Diagram

  5. Maintenance Daily • Clean filters • Clean air vents • Disinfect jars, tubing, other components that come into contact with patient fluids between each use in solution of water, detergent, and disinfectant • Change bacteria filter if wet or discolored • Check collection bottle for cracks, chips, and other damage • Make sure there is a sufficient supply of bacterial filters • Check that float valve moves freely •

  6. 1-3 months • Clean or replace air intake filter • Clean brushes on motors as necessary • Inspect power cord and plug • Ensure vacuum works over full range of suction pressures if there is a control/knob • Verify that overflow valve (float valve) works properly when container is filled with water • Grounding resistance between chassis and ground pin should not exceed 0.5 ohms •

  7. Common Failures Most common: Clogs, Leaks, Motor Failure  Leaks:  - Tube leak Jar/lid leak  Tube blockage  Missing/damaged parts:  -Bad motor - Bad power supply - No collection bottle - No filter - Clogged bacterial filter

  8. Float valve closes air pathway with full collection jar Insufficient pressure/suction  - Ventilation grill obstructed - Suction control knob is set to a low setting Diaphragm needs cleaning or replacing - - Brushes need cleaning Motor needs lubrication 